SHOCOVID: Oxygen Saturation: Analytical Comparison for COVID

Sponsor
Assistance Publique - Hôpitaux de Paris (Other)
Overall Status
Withdrawn
CT.gov ID
NCT04874805
Collaborator
(none)
0
1
1

Study Details

Study Description

Brief Summary

In many patients, respiratory Sars-Cov2 infection causes arterial hypoxemia, which remains without signs of verbalized respiratory distress, up to a point. This phenomenon, called "happy" or "silent" hypoxemia, has a plural pathophysiological basis. Hypoxemia has been shown to be predictively associated with admission to the ICU. Therefore, the question of constant monitoring of oxygenation, practiced on a large scale, at home, in asymptomatic patients or contact cases, arises. A large number of portable pulse oximeter are currently freely available on the market; however, their clinical validation remains sometimes doubtful, or even absent from FDA standards.

The objective of this study is to evaluate the accuracy of SpO2 values provided by portable pulse oximeter in COVID patients, in comparison with the reference method. The study will be conducted on a population of adult patients with COVID, hospitalized in the ICU, for whom gasometry sampling is already scheduled in the usual management.

Condition or Disease Intervention/Treatment Phase
  • Other: Measurement of oxygen saturation
N/A

Detailed Description

The constant monitoring of saturation by the pulse oximeter, the result of three centuries of technological progress, has introduced a true revolution in medical management that now relies on this vital parameter. In many patients, Sars-Cov2 respiratory infection causes arterial hypoxemia, which remains without signs of verbalized respiratory distress, up to a certain point. This phenomenon, called "happy" or "silent" hypoxemia, has a plural pathophysiological basis. Hypoxemia has been shown to be predictively associated with admission to the ICU. Therefore, the question of constant monitoring of oxygenation, practiced on a large scale, at home, in asymptomatic patients or contact cases, arises.

Despite its many advantages (non-invasive, kinetic), the pulse oximeter can be misused in several pathophysiological situations. The reference method for measuring hemoglobin oxygen saturation remains arterial gasometry.

However, a large number of portable pulse oximeter are currently freely available on the market; however, their clinical validation is sometimes questionable or even absent from FDA standards.

The objective of this study is to evaluate the accuracy of SpO2 values provided by portable pulse oximeter in COVID patients, in comparison with the reference method (arterial gasometry). The study will be conducted on a population of adult patients with COVID, hospitalized in the ICU, for whom gasometry sampling is already scheduled in the usual management.

Daily at 6am, during the systematic monitoring, arterial gasometry will be taken (as part of routine care) by the nurse or medical staff on duty. Synchronous measurement of Sp02 given by three portable pulse oximeter (1 measurement point per device, therefore 3 measurements/patient/day for a minimum of 3 days).

Study Design

Study Type:
Interventional
Actual Enrollment :
0 participants
Allocation:
N/A
Intervention Model:
Single Group Assignment
Masking:
None (Open Label)
Primary Purpose:
Other
Official Title:
Oxygen Hemoglobin Saturation in COVID Patients: Comparison of Arterial Gasometry and Pulse Oximetry
Actual Study Start Date :
May 2, 2022
Actual Primary Completion Date :
May 2, 2022
Actual Study Completion Date :
May 2, 2022

Arms and Interventions

Arm Intervention/Treatment
Experimental: Major patient admitted to the ICU for COVID

Other: Measurement of oxygen saturation
Measurement of oxygen saturation by pulse oximetry once a day, at 6 a.m., for a minimum of 3 days, thanks to three pulse oximeters (1 measurement point per device/day), in parallel with a synchronous sampling, programmed as part of the usual care, of arterial gasometry.

Outcome Measures

Primary Outcome Measures

  1. To evaluate the accuracy of SpO2 value provided by portable pulse oximeter in COVID patients compared with the reference method. [once a day]

    Measurement of oxygen saturation by pulse oximetry once a day, at 6 a.m., for a minimum of 3 days, using three pulse oximeters (1 measurement point per device/day), in parallel with a synchronous, programmed arterial gasometry sampling Evaluation according to the standards of the SFBC (French Society of Clinical Biology) for the comparison of methods: minimum 40 points per apparatus, evenly distributed over the pathophysiological range of the saturations explored (9, 10): 60-100%.

Secondary Outcome Measures

  1. Identification of pathophysiological or analytical biases specific to COVID patients. [through study completion, an average of 7 months]

    Data analysis will allow to identify potential biases in the measurement of oxygen saturation in COVID patients

Eligibility Criteria

Criteria

Ages Eligible for Study:
18 Years and Older
Sexes Eligible for Study:
All
Accepts Healthy Volunteers:
No
Inclusion Criteria:
  • Patients hospitalized in the intensive care unit for COVID

  • Age ≥ 18 years

  • Patients or family/trusted person informed and not objecting to participating in the research

Exclusion Criteria:
  • Carbon monoxide poisoning

  • Homozygous sickle cell disease

  • Severe anemia (< 7 g/L) with associated hypoxemia

  • Methemoglobinemia

  • Sulfhemoglobinemia

  • Sepsis, vasoconstrictive drugs

  • Patients under guardianship / curators

  • Patient under state medical aid

Contacts and Locations

Locations

Site City State Country Postal Code
1 Service de Pneumologie, Médecine Intensive et Réanimation (SPMIR-R3S) Hôpital Pitié-Salpêtrière Paris France 75013

Sponsors and Collaborators

  • Assistance Publique - Hôpitaux de Paris

Investigators

None specified.

Study Documents (Full-Text)

None provided.

More Information

Publications

None provided.
Responsible Party:
Assistance Publique - Hôpitaux de Paris
ClinicalTrials.gov Identifier:
NCT04874805
Other Study ID Numbers:
  • APHP210142
First Posted:
May 6, 2021
Last Update Posted:
May 10, 2022
Last Verified:
May 1, 2022
Studies a U.S. FDA-regulated Drug Product:
No
Studies a U.S. FDA-regulated Device Product:
No
Additional relevant MeSH terms: