LinkCor: Evaluation of Safety and Efficacy of a Bioengineered Corneal Implant for Treatment of Keratoconus
Study Details
Study Description
Brief Summary
This is a prospective, Phase I open-label safety study to evaluate the safety, feasibility and initial efficacy of a bioengineered corneal stromal substitute after implantation into the stroma of adults with advanced keratoconus. The bioengineered substitute and implantation technique may provide a less invasive treatment option for advanced keratoconus than current surgical approaches, that additionally does not rely on the limited availability of human donor corneal tissue.
Condition or Disease | Intervention/Treatment | Phase |
---|---|---|
|
N/A |
Detailed Description
This is a Phase I, open-label, non-randomized, non-controlled ongoing clinical case series in two participating centers in Indian and Iranian hospitals, to evaluate the safety and feasibility of minimally-invasive surgery of advanced keratoconus using a bioengineered corneal stromal substitute on a compassionate basis to alleviate blindness and low vision, due to the severely limited availability of human corneal donor tissue at these centers relative to the medical need. Patients fulfilling inclusion criteria are recruited consecutively and implanted intra-stromally with the bioengineered substitute, as a less invasive surgery than full-thickness or anterior lamellar corneal transplantation. In the first phase of the study, a total of 20 patients were recruited and safety and efficacy outcomes were evaluated in those having a minimum of 12 months of postoperative follow-up. In the second phase of the study (currently ongoing), up to 40 patients are being recruited, with one-year postoperative outcomes of all included patients to be evaluated as the final study endpoint. Primary safety and efficacy outcomes will be evaluated.
Study Design
Arms and Interventions
Arm | Intervention/Treatment |
---|---|
Experimental: Bioengineered corneal substitute A cell-free, sterilized bioengineered corneal substitute made from medical grade collagen |
Device: LinkCor Bioengineered Corneal Implant
Intrastromal implantation of a bioengineered corneal equivalent device
|
Outcome Measures
Primary Outcome Measures
- Safety outcome measure [6 months]
The primary safety endpoint is defined such that a rejection leading to implant removal should occur in no more than 10% of implanted eyes. The implant rejection is defined as a clinical situation when the implanted eye has decreased central corneal clarity and decreased implant clarity, corneal vascularization inside of laser created pocket and inflammation.
- Corneal curvature measured by keratometry (efficacy measure) [6 months]
The maximum and minimum corneal curvature (steepness) of each eye will be measured by a keratometer and reported in diopter (D) unit and represented as "K-value". The efficacy outcome is that more than 60% of the eyes should have a decrease in K-value at 6 months postoperatively.
- Central corneal thickness measured by topography techniques (efficacy measure) [6 months]
Central corneal thickness (CCT) will be measured in microns and reported using a two-dimensional thickness map made by optical coherence tomography (OCT) and/or Pentacam topography. The efficacy outcome is that more than 60% of the eyes should have an increase in CCT at 6 months postoperatively.
- Visual acuity (efficacy measure) [6 months]
Vision improvement equivalent of minimum one line in Snellen chart in uncorrected distance visual acuity (UDVA) or corrected distance visual acuity (CDVA).
Eligibility Criteria
Criteria
Inclusion Criteria:
-
Advanced keratoconus (according to Amsler-Krumeich classification)
-
No corneal scar
-
Male or female aged ≥ 18 years, no upper age limit
-
Subjects indicated for a first corneal stromal transplantation (treatment naïve)
-
Corneal thickness (including epithelium) at least 300µm centrally, as measured by OCT.
-
Patients who gave their written signed and dated informed consent for participation
Exclusion Criteria:
-
Prior corneal surgery (e.g., refractive surgery, cataract, collagen cross-linking, endothelial keratoplasty, etc.)
-
Dry eye / tear film pathology
-
Active ocular infection
-
Glaucoma / ocular hypertension
-
Active corneal ulceration
-
Acute or chronic disease or illness that would increase the operation risk or confound the outcomes of the study (immune- compromised, connective tissue disease, clinically significant atopic disease, etc.)
-
Any other medical condition that in the judgment of the clinical investigator or corneal surgeon is not compatible with the study procedures
-
General history judged by the investigator to be incompatible with the study (e.g., life-threatening patient condition, other condition where postoperative follow-up may be difficult).
-
known diabetes or other neuro-degenerative disorder (as corneal nerves can be affected leading to impaired wound healing)
-
Inability of patient to understand the study procedures and thus inability to give informed consent.
-
Participation in another clinical study within the last 3 months
-
Already included once in this study (can only be included for one treated eye).
Contacts and Locations
Locations
Site | City | State | Country | Postal Code | |
---|---|---|---|---|---|
1 | All India Institute of Medical Sciences, Dr. R. P. Centre for Ophthalmic Sciences, AIIMS, | New Delhi | India | 110029 | |
2 | Farabi Eye Hospital, Tehran University of Medical Sciences | Tehran | Iran, Islamic Republic of | 13366 |
Sponsors and Collaborators
- LinkoCare Life Sciences AB
Investigators
- Principal Investigator: Namrata Sharma, MD, ALL INDIA INSTITUTE OF MEDICAL SCIENCES, Dr. R. P. Centre for Ophthalmic Sciences, AIIMS, New Delhi
- Principal Investigator: Mahmoud Jabbarvand Behrouz, MD, University of Tehran, Farabi Eye Hospital
Study Documents (Full-Text)
None provided.More Information
Publications
None provided.- CMG-PRT-001