Performance of a Single-piece Toric Acrylic Intraocular Lens

Sponsor
Prim. Prof. Dr. Oliver Findl, MBA (Other)
Overall Status
Completed
CT.gov ID
NCT02155959
Collaborator
(none)
30
21.1

Study Details

Study Description

Brief Summary

Abstract Purpose: To evaluate the visual performance and rotational stability of the Tecnis Toric one-piece IOL (AMO, USA) during the first 3 post-operative months Design: Prospective single center study Setting: VIROS - Vienna Institute for Research in Ocular Surgery, a Karl Landsteiner Institute, Hanusch Hospitel, Vienna, Austria Methods: In this prospective study patients with age-related cataract and corneal astigmatism of 1.0 to 3.0 D measured with the IOL-Master 500 (Carl Zeiss Meditec AG, Germany) were included. Pre-operatively, rotating Scheimpflug scans (Pentacam HR, Oculus, Germany) were performed and the cornea was marked in the sitting position at the slit lamp. Patients recieved a single-piece toric hydrophobic acrylic IOL (Tecnis Toric, AMO, USA). Immediately and 3 months after surgery retroillumination photographs were taken to assess the rotational stability of the IOL. Additionally, Autorefraction (Topcon, USA), subjective refraction, uncorrected and distance corrected visual acuity, keratometry, Scheimpflug and ocular wavefront (WASCA, Carl Zeiss Meditec AG, Germany) measurements were performed at the 3 months follow-up.

Condition or Disease Intervention/Treatment Phase
  • Device: toric intraocular lens

Detailed Description

Whereas first designs of toric intraocular lenses (toric IOLs) in the early 1990's showed an IOL rotation of more than 30° in one fifth of the patients, modern toric IOLs typically show a mean absolute rotation of 3° to 5°, which would result in a loss of about 10% to 15% of the astigmatism reducing effect of the toric IOL.

Rotational stability of a toric IOL depends on the interaction between the toric IOL and the posterior capsule, whereas misalignment of the toric IOL (defined in this study as the difference between the 3 months postoperatively measured axis of the toric IOL and the intended axis) depends on several factors additionally to rotational stability. Intraoperatively, misalignment may happen due to cyclotorsion of the eye in the lying position or due to peribulbar anaesthesia, and due to imprecision of the surgeon when positioning of the IOL relative to the intended meridian. Both these imprecisions can be dealt with by pre-operative marking of the eye in the sitting position and diligence by the surgeon. Postoperatively, the IOL may rotate because it is undersized for the capsule bag or due to the capsule shrinkage that takes place during fibrotic contraction of the bag in the postoperative period. Since most current IOLs are slightly oversized for the capsule bag, the former is observed rarely and would be more likely in long eyes which tend to have a larger capsule bag diameter. However, capsule bag shrinkage is thought to induce rotation in IOLs with open-loop haptics due to the asymmetry of the haptic design. Typical IOL haptic designs that improve rotational stability are either plate haptic IOLs or special Z-haptic shaped open-loop haptics that attempt to counteract the rotational effect of compression of the shrinking bag. However, both of these IOL designs are thought to have downsides. The former seems to have an increased risk of posterior capsule opacification due to a less effective lens epithelial barrier effect of the optic edge and may also show more rotation in the bag immediately after surgery due to the shorter haptic overall length. The latter IOL design, such as the Z-haptic IOL, is cumbersome to implant and may be prone to damage during implantation.

Aim of this study was to evaluate the rotational stability of a novel single-piece hydrophobic acrylic toric IOL with a C-haptic design.

Study Design

Study Type:
Observational
Actual Enrollment :
30 participants
Observational Model:
Cohort
Time Perspective:
Prospective
Official Title:
Rotational Stability of a Single-piece Toric Acrylic Intraocular Lens: a Pilot Study
Study Start Date :
Mar 1, 2011
Actual Primary Completion Date :
Dec 1, 2012
Actual Study Completion Date :
Dec 1, 2012

Arms and Interventions

Arm Intervention/Treatment
one group

one group receiving a toric intraocular lens during cataract surgery.

Device: toric intraocular lens
toric intraocular lens
Other Names:
  • Tecnis toric IOL (Abott Medical Optics, USA)
  • Outcome Measures

    Primary Outcome Measures

    1. Rotational stability of the investigated toric lens [3 months]

      Rotation of the investigated intraocular lens using a photograph technique (retroillumination images) immediately after surgery and 3 months later.

    Secondary Outcome Measures

    1. reduction in post-operative higher order aberrations [3 months]

      the investigated intraocular lens was proven to reduce some higher order aberrations (this is a refractive problem caused by the cornea). It was shown that the investigated intraocular lens was reducind higher order aberrations in our study population.

    Eligibility Criteria

    Criteria

    Ages Eligible for Study:
    21 Years and Older
    Sexes Eligible for Study:
    All
    Accepts Healthy Volunteers:
    No
    Inclusion Criteria:
    • corneal astigmatism between 1.0 D and 3.0 D

    • cataract

    Exclusion Criteria:
    • irregular astigmatism

    • forme fruste Keratoconus

    • corneal scars

    • phakodonesis

    • pseudoexfoliation syndrome

    • traumatic cataract

    Contacts and Locations

    Locations

    No locations specified.

    Sponsors and Collaborators

    • Prim. Prof. Dr. Oliver Findl, MBA

    Investigators

    • Principal Investigator: Oliver Findl, MD, MBA, Vienna Institute for Research in Ocular Surgery

    Study Documents (Full-Text)

    None provided.

    More Information

    Publications

    None provided.
    Responsible Party:
    Prim. Prof. Dr. Oliver Findl, MBA, Prim. Prof. Dr., MBA, Vienna Institute for Research in Ocular Surgery
    ClinicalTrials.gov Identifier:
    NCT02155959
    Other Study ID Numbers:
    • toric_rotation_acrylic
    First Posted:
    Jun 4, 2014
    Last Update Posted:
    Jun 4, 2014
    Last Verified:
    Jun 1, 2014
    Keywords provided by Prim. Prof. Dr. Oliver Findl, MBA, Prim. Prof. Dr., MBA, Vienna Institute for Research in Ocular Surgery
    Additional relevant MeSH terms:

    Study Results

    No Results Posted as of Jun 4, 2014