PSG Validation Study of Zensorium Biosensing Wearable Device

Sponsor
University of California, San Diego (Other)
Overall Status
Completed
CT.gov ID
NCT04200495
Collaborator
Nitto Denko Asia Technical Centre (NAT) (Other)
68
1
11.3
6

Study Details

Study Description

Brief Summary

The objective of this study is to collect data comparing the Zensorium Biosensing Wearable Device (zBWD) to polysomnography (PSG) in individuals with Sleep-Wake Disorders. zBWD is similar to other wearable devices (e.g., fitbit and apple watch) that use an optical light sensor and Tri-axis Accelerometer to record steps, activity, sleep/awake states, sleep staging, and statistical variation of pulse pressure to track stress. Specifically, it is designed to report sleep duration, non-rapid eye movement (NREM) and REM sleep duration. However, zBWD has not been validated against PSG for sleep tracking in healthy controls or in individuals with sleep-wake disorders.

Condition or Disease Intervention/Treatment Phase

    Detailed Description

    Subjects known to have or not have sleep-wake disorder, as well as healthy subjects, will be enrolled if they otherwise meet all the inclusion criteria, and none of the exclusion criteria. Subject will be asked to complete several questionnaires and wear the zBWD for 7-days, in an off-site (Home) study, followed by a single overnight PSG study to compare the zBWD data to the current gold standard of PSG. Subjects who have completed the study will also be sent a letter thanking them for their participation and offering a summary of the results of their sleep study.

    Study Design

    Study Type:
    Observational
    Actual Enrollment :
    68 participants
    Observational Model:
    Cohort
    Time Perspective:
    Prospective
    Official Title:
    PSG Validation Study of Zensorium Biosensing Wearable Device
    Actual Study Start Date :
    Nov 20, 2019
    Actual Primary Completion Date :
    Oct 28, 2020
    Actual Study Completion Date :
    Oct 28, 2020

    Arms and Interventions

    Arm Intervention/Treatment
    Healthy Controls

    No sleep-wake disorder present

    Active Sleep-Wake Disorder

    Presence of Sleep-Wake Disorder

    Outcome Measures

    Primary Outcome Measures

    1. polysomnography derived percent REM and percent non-REM sleep [One overnight stay at Visit 2 (7 - 10 days following Baseline)]

      A full clinical overnight polysomnography (PSG) study will be performed 7 - 10 days following visit 1. The subject will have EEG, EMG, EOG, and ECG electrodes, an adhesive body position sensor placed in standard locations. This equipment is standard for diagnostic PSG. The variables of interest from the PSG are stages of sleep as indicated by PSG including % rapid eye movement (REM; amount of time in REM sleep/Total sleep time) and % non-REM (amount of time in non-REM sleep/Total sleep time)

    2. Zensorium derived percent REM and percent Non-REM sleep [To be worn everyday between Baseline to visit 2 (7 -10 days)]

      Zensorium Biosensing Wearable Device (zBWD) will be worn for 7 to 10 days following visit 1. The variables of interest are stages of sleep as indicated by the zBWD including % rapid eye movement (REM; amount of time in REM sleep/Total sleep time) and % non-REM (amount of time in non-REM sleep/Total sleep time).

    Secondary Outcome Measures

    1. Sleep Diary [To be filled out everyday between Baseline to visit 2 (7 -10 days)]

      Tracking off-site (home) sleep for 7 to 10 days via self report diary - Sleep Efficiency, Total Sleep Time, Sleep latency, and Wake after sleep onset

    2. Insomnia Severity Index [Baseline]

      Insomnia Severity Index (ISI) is a widely used measure of insomnia with well-established reliability and validity. The ISI consists of seven items, three of which assess severity of insomnia (i.e., degree of difficulty falling asleep, staying asleep, and waking too early). The remaining questions tap satisfaction with sleep pattern, effect of sleep on daytime and social functioning, and concern about current sleep difficulties. Ranges from 0 (no insomnia) to 28 (Severe insomnia)

    3. Pre-Sleep Arousal Scale [Baseline]

      Pre-sleep arousal scale (PSAS) is a self-administered measure in which participants rate the intensity of experienced arousal for somatic (8 items) and cognitive (8 items) subscales. The PSAS shows strong internal consistency and reliability. Higher scores indicate worse pre-sleep arousal. Ratings range from 1 (not at all) to 5 (extremely).

    4. Berlin Questionnaire [Baseline]

      Berlin Questionnaire is a validated screening tool for sleep disordered breather. One question and four follow-up questions concern snoring; three questions address daytime sleepiness, with a subquestion about sleepiness behind the wheel. Lastly, the measure addresses history of high blood pressure, age, weight, height, sex, neck circumference, and ethnicity. High Risk is considered if there are 2 or more categories where the score is positive, where more categories endorsed equates to higher risk.

    5. Epworth Sleepiness Scale [Baseline]

      Epworth Sleepiness Scale (ESS) is a self-administered questionnaire with 8 questions. The questions measure participants experiences of having dozed off or fallen asleep while engaged in eight different activities that differ widely in their somnificity (including sitting in traffic or watching television). can range from 0 to 24. The higher the ESS score, the higher that person's average sleep propensity in daily life (ASP), or their 'daytime sleepiness'

    Eligibility Criteria

    Criteria

    Ages Eligible for Study:
    18 Years to 65 Years
    Sexes Eligible for Study:
    All
    Inclusion Criteria:
      1. Health status: Healthy, or those with sleep-wake disorders
    Exclusion Criteria:
    1. Pregnancy

    2. Currently smoking

    3. Any respiratory disorder other than associated with Sleep/Wake Disorder or well-controlled asthma

    4. Habitual use of medications known to affect respiratory function (e.g. opioids, benzodiazepines, etc).

    5. Use of medication known to affect pulse pressure (Beta Blockers)

    6. Atrial Fibrillation

    7. Tattoos located on wrist

    Contacts and Locations

    Locations

    Site City State Country Postal Code
    1 University of California San Diego San Diego California United States 92093

    Sponsors and Collaborators

    • University of California, San Diego
    • Nitto Denko Asia Technical Centre (NAT)

    Investigators

    None specified.

    Study Documents (Full-Text)

    None provided.

    More Information

    Publications

    None provided.
    Responsible Party:
    Peter Colvonen, Assistant Clinical Professor, University of California, San Diego
    ClinicalTrials.gov Identifier:
    NCT04200495
    Other Study ID Numbers:
    • 190705
    First Posted:
    Dec 16, 2019
    Last Update Posted:
    Feb 4, 2021
    Last Verified:
    Feb 1, 2021
    Individual Participant Data (IPD) Sharing Statement:
    No
    Plan to Share IPD:
    No
    Studies a U.S. FDA-regulated Drug Product:
    No
    Studies a U.S. FDA-regulated Device Product:
    No
    Additional relevant MeSH terms:

    Study Results

    No Results Posted as of Feb 4, 2021