The primary objective of this study is to evaluate the potential of a new large field CZT gamma camera to estimate the thyroid uptake (TU) on Single photon emission computed tomography (SPECT) and SPECT/CT images in comparison with standard planar scintigraphy.
The secondary objective is to analyze the diagnostic contribution of SPECT/CT imaging.
Condition or Disease | Intervention/Treatment | Phase |
---|---|---|
|
Detailed Description
Planar and SPECT/CT scintigraphy will be performed on a new generation large field CZT gamma camera and thyroid uptake (TU) value will be calculated on planar acquisition for 23 consecutive patients referred in our department for hyperthyroidism.
With a quantitative approach using attenuation, scatter correction and iterative algorithm with resolution recovery we will have the possibility to calculate the TU value with SPECT or SPECT/CT acquisitions.
Our primary objective will be to evaluate the potential of a new large field CZT gamma camera to estimate the thyroid uptake on SPECT (Tomography without attenuation correction) and SPECT/CT (tomography with attenuation correction) acquisitions compared with standard planar scintigraphy as a gold standard.
The secondary objective will be to analyze the diagnostic contribution of SPECT/CT images.
To established this study it will be necessary to introduce cross-calibration factors performed from a thyroid phantom, for planar and SPECT acquisitions, to estimate the thyroid uptake values.
Study Design
Outcome Measures
Primary Outcome Measures
- Thyroid uptake [Inclusion]
Quantification of thyroid uptake
Secondary Outcome Measures
- SPECT/CT [Inclusion]
Diagnostic contribution of SPECT/CT
Eligibility Criteria
Criteria
Inclusion Criteria:
- Hyperthyroidism confirmed by blood test (TSH) and thyroid ultrasound
Exclusion Criteria:
- Pregnant women
Contacts and Locations
Locations
Site | City | State | Country | Postal Code | |
---|---|---|---|---|---|
1 | Centre Hospitalier Princesse Grace | Monaco | Monaco | 98000 |
Sponsors and Collaborators
- Centre Hospitalier Princesse Grace
Investigators
None specified.Study Documents (Full-Text)
None provided.More Information
Publications
- Giovanella L, Avram AM, Iakovou I, Kwak J, Lawson SA, Lulaj E, Luster M, Piccardo A, Schmidt M, Tulchinsky M, Verburg FA, Wolin E. EANM practice guideline/SNMMI procedure standard for RAIU and thyroid scintigraphy. Eur J Nucl Med Mol Imaging. 2019 Nov;46(12):2514-2525. doi: 10.1007/s00259-019-04472-8. Epub 2019 Aug 7.
- Gnesin S, Leite Ferreira P, Malterre J, Laub P, Prior JO, Verdun FR. Phantom Validation of Tc-99m Absolute Quantification in a SPECT/CT Commercial Device. Comput Math Methods Med. 2016;2016:4360371. doi: 10.1155/2016/4360371. Epub 2016 Dec 14.
- Mariani G, Tonacchera M, Grosso M, Orsolini F, Vitti P, Strauss HW. The Role of Nuclear Medicine in the Clinical Management of Benign Thyroid Disorders, Part 1: Hyperthyroidism. J Nucl Med. 2021 Mar;62(3):304-312. doi: 10.2967/jnumed.120.243170. Epub 2020 Oct 2. Review.
- Ritt P, Vija H, Hornegger J, Kuwert T. Absolute quantification in SPECT. Eur J Nucl Med Mol Imaging. 2011 May;38 Suppl 1:S69-77. doi: 10.1007/s00259-011-1770-8. Epub 2011 Apr 12. Review.
- 21-09