Accuracy of Pulse Oximeters With Profound Hypoxia

Sponsor
Mortara Instrument (Industry)
Overall Status
Completed
CT.gov ID
NCT02571686
Collaborator
(none)
10
1

Study Details

Study Description

Brief Summary

The aim of this project is to test the accuracy of pulse oximeters during mild, moderate and severe hypoxia. This is done by comparing the reading of the pulse oximeter during brief, steady state hypoxia with a gold-standard measurement of blood oxyhemoglobin saturation (arterial blood sample processed in a laboratory hemoximeter). The data obtained is submitted by pulse oximeter manufacturers to the FDA for device approval.

Condition or Disease Intervention/Treatment Phase
  • Device: Surveyor S4

Detailed Description

Overall design:

Studies normally involve 6-12 normal adult volunteer paid subjects, with 20-25 1-2 ml arterial blood samples from each subject obtained at different steady-state levels of hypoxia from 70-100%. Blood samples obtained from an arterial line are measured in a hemoximeter to determine true saturation value. Some sponsors may not require any blood sampling. Some sponsors may require additional blood samples, up to 35 samples may be drawn.

Readings from the test pulse oximeters are recorded and compared to these "gold-standard" blood values. The pulse oximeter probes may be located at a variety of sites on the subjects, including fingers, toes, ears, forehead, scalp and bridge of the nose. These detector-probes are all non-invasive. The level of hypoxia is measured and controlled by the investigator. A computer program that displays a prediction, breath by breath, the arterial oxygen saturation, (SaO2) of the study subject. SaO2 is computed from end-expired Po2 and Pco2 as determined by mass spectrometer gas analysis. This information permits the inspired gas mixture of air, plus CO2 and nitrogen, to be adjusted by an operator watching the value computed after each expiration on an analog meter. This computer-estimated saturation is adjusted by the operator to one of 6 levels of predicted saturation, and is held stable for about 30 seconds at each level. Two or three "runs" are conducted per subject. Each "run" lasts 8-12 min and 4 to 5 plateaus are tested per run. The manufacturer may choose the target values. Plateaus are typically sought at 92%, 86%, 80%, 74%, 68% and 62%. Other manufacturers have asked for an equal number of data points but with all points between 70% and 100%. Some sponsors may also request additional variables to be measured during testing including high or low Carbon Dioxide (CO2), where subjects will be asked to hyperventilate; low perfusion, where subjects will be asked to lay flat, or with their head up or down; testing pulse oximeters during motion, with the subject's hand fixed to a motion machine; or measure eye-tracking during hypoxia. The subject's will be informed by the study staff and in the consent form if any of these additional procedures apply.

.

Study Design

Study Type:
Observational
Actual Enrollment :
10 participants
Time Perspective:
Prospective
Official Title:
Accuracy of Pulse Oximeters With Profound Hypoxia
Study Start Date :
Oct 1, 2015
Actual Primary Completion Date :
Oct 1, 2015
Actual Study Completion Date :
Oct 1, 2015

Outcome Measures

Primary Outcome Measures

  1. Accuracy of Pulse Oximeter [2 months]

Eligibility Criteria

Criteria

Ages Eligible for Study:
18 Years to 50 Years
Sexes Eligible for Study:
All
Accepts Healthy Volunteers:
Yes
Inclusion Criteria:
  1. The subject is male or female, aged ≥18 and <50.

  2. The subject is in good general health with no evidence of any medical problems.

  3. The subject is fluent in both written and spoken English.

  4. The subject has provided informed consent and is willing to comply with the study procedures.

Exclusion Criteria:
  1. The subject is obese (BMI>30).

  2. The subject has a known history of heart disease, lung disease, kidney or liver disease.

  3. Diagnosis of asthma, sleep apnea, or use of CPAP(Continuous positive airway pressure).

  4. Subject has diabetes.

  5. Subject has a clotting disorder.

  6. The subject a hemoglobinopathy or history of anemia, per subject report or the first blood sample, that in the opinion of the investigator, would make them unsuitable for study participation.

  7. The subject has any other serious systemic illness.

  8. The subject is a current smoker.

  9. Any injury, deformity, or abnormality at the sensor sites that in the opinion of the investigators' would interfere with the sensors working correctly.

  10. The subject has a history of fainting or vasovagal response.

  11. The subject has a history of sensitivity to local anesthesia.

  12. The subject has a diagnosis of Raynaud's disease.

  13. The subject has unacceptable collateral circulation based on exam by the investigator (Allen's test).

  14. The subject is pregnant, lactating or trying to get pregnant.

  15. The subject is unable or unwilling to provide informed consent, or is unable or unwilling to comply with study procedures.

  16. The subject has any other condition, which in the opinion of the investigators' would make them unsuitable for the study -

Contacts and Locations

Locations

Site City State Country Postal Code
1 Department of Anesthesia Clinical Studies Facility, room A-65 Parnassus Campus. San Francisco California United States 94143

Sponsors and Collaborators

  • Mortara Instrument

Investigators

None specified.

Study Documents (Full-Text)

None provided.

More Information

Publications

None provided.
Responsible Party:
Mortara Instrument
ClinicalTrials.gov Identifier:
NCT02571686
Other Study ID Numbers:
  • 139253
First Posted:
Oct 8, 2015
Last Update Posted:
Dec 24, 2015
Last Verified:
Dec 1, 2015
Additional relevant MeSH terms:

Study Results

No Results Posted as of Dec 24, 2015