RIVAL - AR EFS: HLT® Meridian® TAVR Valve System Early Feasibility Study for Aortic Regurgitation

Sponsor
HLT Inc. (Industry)
Overall Status
Recruiting
CT.gov ID
NCT05409378
Collaborator
(none)
10
3
1
71.8
3.3
0

Study Details

Study Description

Brief Summary

To evaluate the safety and performance of The Meridian® TAVR Valve in patients with severe aortic regurgitation who present at high risk for aortic valve replacement surgery.

Condition or Disease Intervention/Treatment Phase
  • Device: The HLT® Meridian® TAVR Valve (Meridian® with TriVentTM Anticalcification Treatment Valve and Pathfinder® Delivery System)
N/A

Study Design

Study Type:
Interventional
Anticipated Enrollment :
10 participants
Allocation:
N/A
Intervention Model:
Single Group Assignment
Masking:
None (Open Label)
Primary Purpose:
Treatment
Official Title:
HLT® Meridian® TAVR Valve System Early Feasibility Study for Aortic Regurgitation
Actual Study Start Date :
Jul 8, 2022
Anticipated Primary Completion Date :
Jul 1, 2023
Anticipated Study Completion Date :
Jul 1, 2028

Arms and Interventions

Arm Intervention/Treatment
Experimental: The HLT® Meridian® TAVR Valve

Transcatheter aortic valve replacement with the HLT Meridian TAVR Valve

Device: The HLT® Meridian® TAVR Valve (Meridian® with TriVentTM Anticalcification Treatment Valve and Pathfinder® Delivery System)
Aortic valve replacement with the HLT Meridian TAVR Valve to treat aortic regurgitation

Outcome Measures

Primary Outcome Measures

  1. Mortality at 30 days [30-Days]

    The primary safety endpoint is all-cause mortality at 30 days

Secondary Outcome Measures

  1. Technical Success - A composite endpoint per VARC-3, defined as meeting the following: [Immediately after procedure]

    Freedom from Mortality Successful access, delivery of the device, and retrieval of the delivery system Correct positioning of a single prosthetic heart valve into the proper anatomical location Freedom from surgery or intervention related to the device or to a major vascular or access-related or cardiac structural complication

  2. Device Success- A composite endpoint per VARC-3, defined as meeting the following: [30 Days]

    Technical Success Freedom from mortality Freedom from surgery or intervention related to the device a or to a major vascular or access-related or cardiac structural complication Intended performance of the valve (mean gradient < 20 mmHg, peak velocity < 3 m/s, Doppler velocity index ≥ 0.25, and less than moderate aortic regurgitation.

  3. Post-procedural Valve Performance - EOA assessed via Echo [Discharge (or day 7), 30 days, 6, 12, 24, 36, 48 and 60 months]

    • Aortic valve effective orifice area (EOA; cm^2)

  4. Post-procedural Valve Performance- Aortic Regurgitation (AR) assessed via Echo [Discharge (or day 7), 30 days, 6, 12, 24, 36, 48 and 60 months]

    • Severity of aortic valve regurgitation (AR; graded scale)

  5. Post-procedural Valve Performance - Aortic Valve Gradient assessed via Echo [Discharge (or day 7), 30 days, 6, 12, 24, 36, 48 and 60 months]

    • Aortic valve gradient (mean and peak; mmHg)

  6. Adverse Events [Up to 5 years following enrollment]

    All adverse events will be assessed throughout the five (5) year follow up period per VARC-3 definitions

Eligibility Criteria

Criteria

Ages Eligible for Study:
N/A and Older
Sexes Eligible for Study:
All
Accepts Healthy Volunteers:
No
Inclusion Criteria:
  1. Subject has evidence of severe aortic regurgitation as assessed by the independent echo core lab according to an integrative multiparametric based on ASE (American Society of Echocardiography) guidelines considering qualitative and quantitative measures.

  2. Symptoms due to severe aortic regurgitation with a NYHA Functional Classification of II or greater

  3. Documented aortic valve annular diameter ≥ 21 and ≤ 26 mm (associated perimeter range is 66.0 - 81.7 mm or associated area range of 346 - 531 mm^2) measured by the MSCT Core Lab assessment of pre-procedure imaging

  4. Documented heart team agreement of high risk for surgical aortic valve replacement (SAVR)

  5. Geographically available, willing to comply with follow-up and able to provide written informed consent.

Exclusion Criteria:
  1. Congenital, unicuspid or bicuspid aortic valve which could compromise procedural success

  2. Severe mitral or tricuspid valve regurgitation

  3. Moderate to severe mitral stenosis

  4. Myocardial infarction within the past 30 days*

  5. Echocardiographic evidence of intracardiac mass, thrombus or vegetation

  6. Left Ventricular Ejection Fraction (LVEF) < 35%

  7. Severe pulmonary hypertension with pulmonary systolic pressure greater than two-thirds of systemic pressure

  8. Hemodynamic instability requiring inotropic drug therapy within the past 30 days

  9. Untreated clinically significant coronary artery disease requiring revascularization

  10. Any therapeutic invasive cardiac procedure performed or planned to perform within 30 days of the index procedure, except for PCI which is performed within 7 days prior to the index procedure*

  11. Presence of significant aortic disease such as atheroma, thrombus, dissection (> 50 mm in diameter or which could be considered high risk for dissection) or aneurysm which precludes safe implant delivery

  12. Blood dyscrasias defined as leukopenia (White blood cell count < 1,000 cell/mm3), anemia (hemoglobin < 9.0 g/dL), thrombocytopenia (platelet count < 50,000 cells/mm3), history of bleeding diathesis or coagulopathy

  13. Patient ineligible for or refuses blood transfusions

  14. Unfavorable peripheral vascular anatomy or disease (e.g. severe obstructive calcification, severe tortuosity) that would preclude passage of 18F catheters from the femoral arterial access to the aorta by the MSCT Core Lab assessment of pre-procedure imaging

  15. Active peptic ulcer or gastrointestinal bleeding requiring hospitalization or transfusion within the past 90 days, that would not allow treatment with the protocol antiplatelet regimen post implant *

  16. Symptoms of carotid or vertebral artery disease (e.g. stroke or transient ischemic attack) within past 6 months, or treatment of carotid stenosis within past two months*

  • At the time of procedure, if a subject's medical status has changed since enrollment, the subject shall be re-evaluated for eligibility

Contacts and Locations

Locations

Site City State Country Postal Code
1 Northwestern University Chicago Illinois United States 60611
2 The Christ Hospital Cincinnati Ohio United States 45219
3 Ohio State University Wexner Medical Center Columbus Ohio United States 43210

Sponsors and Collaborators

  • HLT Inc.

Investigators

  • Principal Investigator: Kendra Grubb, MD, Emory University
  • Principal Investigator: Charles Davidson, MD, Northwestern University

Study Documents (Full-Text)

None provided.

More Information

Publications

None provided.
Responsible Party:
HLT Inc.
ClinicalTrials.gov Identifier:
NCT05409378
Other Study ID Numbers:
  • HLT2101
First Posted:
Jun 8, 2022
Last Update Posted:
Jul 13, 2022
Last Verified:
Jul 1, 2022
Individual Participant Data (IPD) Sharing Statement:
No
Plan to Share IPD:
No
Studies a U.S. FDA-regulated Drug Product:
No
Studies a U.S. FDA-regulated Device Product:
Yes
Keywords provided by HLT Inc.
Additional relevant MeSH terms:

Study Results

No Results Posted as of Jul 13, 2022