Safety and Effectiveness of the External Fastening System for Temporary Active Fixation Pacemakers (TEMPACE)
Study Details
Study Description
Brief Summary
Temporary cardiac pacing with active-fixation leads (TPAFL) using a reusable permanent pacemaker generator has been shown to be safer than lead systems without fixation. However, TPAFL requires the off-label (non-certificated) use of Pacemaker leads and generators. It is designed a fastening system to ensure the safety and efficacy of the procedure: the KronoSafe System®. The aim of this study is to demonstrate the safety and effectiveness of the KronoSafe System® for temporary pacing in a series of patients receiving TPAFL.
Condition or Disease | Intervention/Treatment | Phase |
---|---|---|
|
N/A |
Study Design
Arms and Interventions
Arm | Intervention/Treatment |
---|---|
Experimental: Temporary Cardiac Pacing with Active-Fixation Leads
|
Device: KronoSafe® System
All patients included in the study will have a temporary pacemaker with active fixation leads, held by the KronoSafe® fixation system.
|
Outcome Measures
Primary Outcome Measures
- Ventricular R wave detection [Immediately after the procedure of lead implantation]
Ventricular R wave, measured in millivolts with pacemaker generator programmer
- Lead impedance detection [Immediately after the procedure of lead implantation]
Lead impedance, measured in ohms with pacemaker generator programmer
- Right ventricle threshold detection [Immediately after the procedure of lead implantation]
Right ventricle threshold, measured in volts with pacemaker generator programmer
- Ventricular R wave detection [24 hours after the procedure of lead implantation]
Ventricular R wave, measured in millivolts with pacemaker generator programmer
- Ventricular R wave detection [day 3 after the procedure of lead implantation]
Ventricular R wave, measured in millivolts with pacemaker generator programmer
- Ventricular R wave detection [day 5 after the procedure of lead implantation]
Ventricular R wave, measured in millivolts with pacemaker generator programmer
- Ventricular R wave detection [day 7 after the procedure of lead implantation]
Ventricular R wave, measured in millivolts with pacemaker generator programmer
- Ventricular R wave detection [day 9 after the procedure of lead implantation]
Ventricular R wave, measured in millivolts with pacemaker generator programmer
- Ventricular R wave detection [day 11 after the procedure of lead implantation]
Ventricular R wave, measured in millivolts with pacemaker generator programmer
- Ventricular R wave detection [day 13 after the procedure of lead implantation]
Ventricular R wave, measured in millivolts with pacemaker generator programmer
- Ventricular R wave detection [day 15 after the procedure of lead implantation]
Ventricular R wave, measured in millivolts with pacemaker generator programmer
- Ventricular R wave detection [day 17 after the procedure of lead implantation]
Ventricular R wave, measured in millivolts with pacemaker generator programmer
- Ventricular R wave detection [day 19 after the procedure of lead implantation]
Ventricular R wave, measured in millivolts with pacemaker generator programmer
- Ventricular R wave detection [day 21 after the procedure of lead implantation]
Ventricular R wave, measured in millivolts with pacemaker generator programmer
- Ventricular R wave detection [day 23 after the procedure of lead implantation]
Ventricular R wave, measured in millivolts with pacemaker generator programmer
- Ventricular R wave detection [day 25 after the procedure of lead implantation]
Ventricular R wave, measured in millivolts with pacemaker generator programmer
- Ventricular R wave detection [day 27 after the procedure of lead implantation]
Ventricular R wave, measured in millivolts with pacemaker generator programmer
- Ventricular R wave detection [day 29 after the procedure of lead implantation]
Ventricular R wave, measured in millivolts with pacemaker generator programmer
- Ventricular R wave detection [Immediately before the procedure of lead removal]
Ventricular R wave, measured in millivolts with pacemaker generator programmer
- Lead impedance detection [24 hours after the procedure of lead implantation]
Lead impedance, measured in ohms with pacemaker generator programmer
- Lead impedance detection [day 3 after the procedure of lead implantation]
Lead impedance, measured in ohms with pacemaker generator programmer
- Lead impedance detection [day 5 after the procedure of lead implantation]
Lead impedance, measured in ohms with pacemaker generator programmer
- Lead impedance detection [day 7 after the procedure of lead implantation]
Lead impedance, measured in ohms with pacemaker generator programmer
- Lead impedance detection [day 9 after the procedure of lead implantation]
Lead impedance, measured in ohms with pacemaker generator programmer
- Lead impedance detection [day 11 after the procedure of lead implantation]
Lead impedance, measured in ohms with pacemaker generator programmer
- Lead impedance detection [day 13 after the procedure of lead implantation]
Lead impedance, measured in ohms with pacemaker generator programmer
- Lead impedance detection [day 15 after the procedure of lead implantation]
Lead impedance, measured in ohms with pacemaker generator programmer
- Lead impedance detection [day 17 after the procedure of lead implantation]
Lead impedance, measured in ohms with pacemaker generator programmer
- Lead impedance detection [day 19 after the procedure of lead implantation]
Lead impedance, measured in ohms with pacemaker generator programmer
- Lead impedance detection [day 21 after the procedure of lead implantation]
Lead impedance, measured in ohms with pacemaker generator programmer
- Lead impedance detection [day 23 after the procedure of lead implantation]
Lead impedance, measured in ohms with pacemaker generator programmer
- Lead impedance detection [day 25 after the procedure of lead implantation]
Lead impedance, measured in ohms with pacemaker generator programmer
- Lead impedance detection [day 27 after the procedure of lead implantation]
Lead impedance, measured in ohms with pacemaker generator programmer
- Lead impedance detection [day 29 after the procedure of lead implantation]
Lead impedance, measured in ohms with pacemaker generator programmer
- Lead impedance detection [Immediately before the procedure of lead removal]
Lead impedance, measured in ohms with pacemaker generator programmer
- Right ventricle threshold detection [24 hours after the procedure of lead implantation]
Right ventricle threshold, measured in volts with pacemaker generator programmer
- Right ventricle threshold detection [day 3 after the procedure of lead implantation]
Right ventricle threshold, measured in volts with pacemaker generator programmer
- Right ventricle threshold detection [day 5 after the procedure of lead implantation]
Right ventricle threshold, measured in volts with pacemaker generator programmer
- Right ventricle threshold detection [day 7 after the procedure of lead implantation]
Right ventricle threshold, measured in volts with pacemaker generator programmer
- Right ventricle threshold detection [day 9 after the procedure of lead implantation]
Right ventricle threshold, measured in volts with pacemaker generator programmer
- Right ventricle threshold detection [day 11 after the procedure of lead implantation]
Right ventricle threshold, measured in volts with pacemaker generator programmer
- Right ventricle threshold detection [day 13 after the procedure of lead implantation]
Right ventricle threshold, measured in volts with pacemaker generator programmer
- Right ventricle threshold detection [day 15 after the procedure of lead implantation]
Right ventricle threshold, measured in volts with pacemaker generator programmer
- Right ventricle threshold detection [day 17 after the procedure of lead implantation]
Right ventricle threshold, measured in volts with pacemaker generator programmer
- Right ventricle threshold detection [day 19 after the procedure of lead implantation]
Right ventricle threshold, measured in volts with pacemaker generator programmer
- Right ventricle threshold detection [day 21 after the procedure of lead implantation]
Right ventricle threshold, measured in volts with pacemaker generator programmer
- Right ventricle threshold detection [day 23 after the procedure of lead implantation]
Right ventricle threshold, measured in volts with pacemaker generator programmer
- Right ventricle threshold detection [day 25 after the procedure of lead implantation]
Right ventricle threshold, measured in volts with pacemaker generator programmer
- Right ventricle threshold detection [day 27 after the procedure of lead implantation]
Right ventricle threshold, measured in volts with pacemaker generator programmer
- Right ventricle threshold detection [day 29 after the procedure of lead implantation]
Right ventricle threshold, measured in volts with pacemaker generator programmer
- Right ventricle threshold detection [Immediately before the procedure of lead removal]
Right ventricle threshold, measured in volts with pacemaker generator programmer
- Lead displacement detection [Immediately before the procedure of lead removal]
Visual observation of 2 cm electrode displacement through vein acces by clinical examination
- Capture failure detection [Immediately before the procedure of lead removal]
Right ventricle threshold above 5 volts during 1 millisecond, measured with pacemaker generator programmer.
- Sensing failure detection [Immediately before the procedure of lead removal]
Ventricular R wave below 1 volt, measured with pacemaker generator programmer
- Lead insulation breakdown detection [Immediately before the procedure of lead removal]
Visual observation of lead insulation breakdown by clinical examination
- Generator detachment detection [Immediately before the procedure of lead removal]
Visual observation of generator detachment by clinical examination. removal of the generator from the casing.
- Cardiac peroration detection [Immediately before the procedure of lead removal]
New pericardial effusion detected by Echocardiography with signs of cardiac tamponade: respiratory variation of transmitral filling flow greater than 25%
Secondary Outcome Measures
- Detection of Bacteremia [Immediately after the procedure of lead removal]
Isolation of the same bacteria in the pacemaker lead and 2 blood cultures from the patient
- Detection of Agitation [Immediately after the procedure of lead removal]
Record of antipsychotic and anxiolytic treatment of the patient
Eligibility Criteria
Criteria
Inclusion Criteria:
-
Patient older than 18 years.
-
Need for temporary cardiac stimulation
-
Temporary active fixation pacemaker implantation through left or right jugular access.
-
Signature of informed consent
-
Correct functioning of the electrode after the implantation: sensing of the R wave above 5 millivolts. Ventricular pacing threshold below 1 volt. Electrode impedance between 400-1200 Ohm.
-
Pacemaker programming with stimulation output at 5 volts intensity and 1 millisecond pulse duration.
-
Placement of the KronoSafe® fastening system after electrode implantation.
Exclusion Criteria:
-
Complication during the procedure: pneumothorax, hematoma at the puncture site, cardiac perforation, accidental arterial puncture or other that, in the investigator's opinion, poses an added risk of pacemaker dysfunction.
-
Need to transfer the patient to another center with loss of follow-up.
-
Absence of surveillance through continuous electrocardiographic monitoring with detection of arrhythmic events.
Contacts and Locations
Locations
No locations specified.Sponsors and Collaborators
- ICU Medical Technologies S.L.
- Universidad Miguel Hernandez de Elche
Investigators
- Principal Investigator: Raimundo Vicente, MD, Intensive Care Physician at Dr. Balmis University Hospital (Alicante)
Study Documents (Full-Text)
None provided.More Information
Publications
None provided.- TEMPACE