Micrima MARIA Data Collection for Machine Learning Study
Study Details
Study Description
Brief Summary
The MARIA® breast imaging system is a CE-marked radiofrequency (RF) medical imaging device. The system employs an electromagnetic imaging technique that exploits the dielectric contrast between normal and cancerous tissues. The use of artificial intelligence will provide additional, novel functionality to the MARIA® system but requires participant data in order to develop and validate the machine learning algorithms that aim to increase the accuracy and overall clinical utility of the device. This study aims to collect data from sites for this purpose.
Condition or Disease | Intervention/Treatment | Phase |
---|---|---|
|
N/A |
Detailed Description
Previous and existing trials have been focused on and designed to assess sensitivity of different versions of MARIA® when used on symptomatic subjects. This study will consider those measures, but the primary endpoint will be the collection of the data required for developing and validating machine learning-based enhancements of the MARIA® system. Data that are diverse in terms of subject characteristics, including disease state, presentation of disease, and variation of normal anatomy are requisite for the creation of accurate and well-characterized machine intelligence. As such, the recruitment criteria is broad in order to encompass all kinds of subjects that are representative as possible of the general population.
At the current stage in development of the MARIA® system a need has been identified to acquire additional data for validation of a lesion classification algorithm specific to women with larger breasts. This need has guided the study size calculation; however, additional data from all shell sizes continues to be important for further lesion classifier development and the development of additional functionality, including automated breast density assessment. Therefore, as part of this study subjects across the entire intended use population will be recruited.
The data points that are collected via this study have been selected as those that are required for the purposes of machine learning first and foremost, with any secondary endpoints utilising only existing available information. In addition to this data, a questionnaire will be provided to each participant to obtain valuable feedback on the scan experience.
All subjects will undertake the same scan procedure, requiring them to have both breasts scanned using the MARIA® system. The MARIA® scan will be in addition to any standard diagnostic procedures, including imaging, that comprise the standard of care and will have no effect on the participant's diagnosis or treatment. While there is no direct benefit nor detrimental effect from this study to the participants, the study has potentially significant implications for future learning and for the breast imaging landscape.
Study Design
Arms and Interventions
Arm | Intervention/Treatment |
---|---|
Experimental: Main study group
|
Device: MARIA
After consent has been given, the participant's breast will be fitted by the operator using a clear plastic cup to ascertain their corresponding MARIA® cup insert size. Once this has been determined, participants will be required to lie prone with their breast pendent through an aperture in the MARIA® participant bed. The cup insert will contain a layer of coupling fluid. The hemispherical cup inserts and array are brought into contact with the breast. A scan sequence comprising of two complete scan rotations will then be performed, which takes around 5 minutes. The participant will then be assisted to adjust their positioning, and the scan sequence performed on the other breast.
|
Outcome Measures
Primary Outcome Measures
- Validation of MARIA breast classifier for malignant/benign tissue using clinical information [12 months]
The ground truth diagnosis for each participant will be obtained from clinical diagnosis information pertaining to histology and imaging. This will ultimately lead to a binary classification for that patient of either a benign or malignant diagnosis. This will then be compared to the classification that the AI has determined for that area of the breast volume.
Secondary Outcome Measures
- - Proportion of participants with invasive breast cancer in whom breast cancer is correctly identified by MARIA® [12 months]
This will compare the number of participants that have been formally diagnosed, via biopsy, with breast cancer to those identified with breast cancer by MARIA
- - Proportion of participants in whom breast cancer is correctly detected using MARIA® compared with mammography, in the overall participants, and stratified by mammographic breast density (BIRADS A-D) and by histological tumour type (DCIS, IDC, ILC) [12 months]
Further analysis on participants identified with breast cancer, but comparing directly to mammography, and stratifying with breast density and histology.
- - Participant tolerance using data obtained via a questionnaire [12 months]
Participants asked for their opinion on the MARIA scan experience, likes and dislikes, and also comparison to other modalities.
Eligibility Criteria
Criteria
Inclusion Criteria
To be eligible to participate in this study, a subject must meet all the following criteria:
-
Attending a symptomatic clinic or other appointment at a participating site
-
Female sex
-
18 years or older
-
Able to provide informed consent
-
Not in any identified, vulnerable group
Exclusion Criteria
A potential subject who meets any of the following criteria will be excluded from participation in this study:
-
Unable to mount MARIA® participant bed
-
Unable to lie in the prone position for a period of up to 5 minutes
-
Participants who have undergone biopsy less than 5 days before the MARIA® scan
-
Participants with implanted electronics
-
Participants with breast implants
-
Participants with nipple piercings (unless they are removed prior to the MARIA® scan)
-
Breast sizes too small or too large to be suitable
Contacts and Locations
Locations
Site | City | State | Country | Postal Code | |
---|---|---|---|---|---|
1 | Leeds Teaching Hospitals NHS Foundation Trust | Leeds | United Kingdom |
Sponsors and Collaborators
- Micrima, Ltd.
Investigators
None specified.Study Documents (Full-Text)
None provided.More Information
Publications
None provided.- PROTOCOL-M7-037