Three Intraocular Lenses Designed to Improve Distance, Intermediate and Near Vision Following Lens Extraction

Sponsor
Bausch & Lomb Incorporated (Industry)
Overall Status
Completed
CT.gov ID
NCT01225952
Collaborator
(none)
93
1
3
12
7.8

Study Details

Study Description

Brief Summary

The objective of this study is to compare the contrast sensitivity, high and low contrast visual acuity (VA), glare meter outcomes, and subject satisfaction with three different FDA-approved intraocular lenses (IOLs) designed to improve distance, intermediate, and near vision following lens extraction in adults at least 40 years of age.

Condition or Disease Intervention/Treatment Phase
  • Device: Crystalens AO
  • Device: ReSTOR 3.0
  • Device: AMO Tecnis Multifocal
Phase 4

Study Design

Study Type:
Interventional
Actual Enrollment :
93 participants
Allocation:
Randomized
Intervention Model:
Parallel Assignment
Masking:
Double (Participant, Outcomes Assessor)
Primary Purpose:
Treatment
Official Title:
A Three Arm Prospective Clinical Evaluation of Three FDA-approved Intraocular Lenses Designed to Improve Distance, Intermediate and Near Vision Following Lens Extraction
Study Start Date :
Jul 1, 2010
Actual Primary Completion Date :
Jun 1, 2011
Actual Study Completion Date :
Jul 1, 2011

Arms and Interventions

Arm Intervention/Treatment
Experimental: Crystalens AO

Bausch & Lomb silicone multi-piece accommodating IOL is a modified plate haptic lens

Device: Crystalens AO
Bausch & Lomb model silicone multi-piece accommodating IOL.

Active Comparator: ReSTOR 3.0

An aspheric multifocal IOL (Alcon Laboratories) combines the functions of an apodized diffractive region and a refractive region.

Device: ReSTOR 3.0
An aspheric multifocal IOL (Alcon Laboratories) combines the functions of an apodized diffractive region and a refractive region.

Active Comparator: AMO Tecnis Multifocal

A foldable hydrophobic acrylic IOL,(Abbott Medical Optics), is an ultraviolet light-absorbing posterior chamber IOL

Device: AMO Tecnis Multifocal
The Tecnis Multifocal foldable hydrophobic acrylic IOL, Model ZMA00 (Abbott Medical Optics), is an ultraviolet light-absorbing posterior chamber IOL.

Outcome Measures

Primary Outcome Measures

  1. Monocular Mesopic Contrast Sensitivity Without Glare (1.5, 3 Cycles/Degree) [Postoperative visit 3 (day 30-60)]

    A single eye (monocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness). The participant's ability to detect changes in contrast is determined. Higher mean indicates improved contrast sensitivity.

  2. Monocular Mesopic Contrast Sensitivity Without Glare(6 Cycles/Degree) [Postoperative visit 3 (day 30-60)]

    A single eye (monocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness). The participant's ability to detect changes in contrast is determined. Higher mean indicates improved contrast sensitivity.

  3. Monocular Mesopic Contrast Sensitivity Without Glare(12 Cycles/Degree) [Postoperative visit 3 (day 30-60)]

    A single eye (monocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness). The participant's ability to detect changes in contrast is determined. Higher mean indicates improved contrast sensitivity.

  4. Monocular Mesopic Contrast Sensitivity Without Glare(18 Cycles/Degree) [Postoperative visit 3 (day 30-60)]

    A single eye (monocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness). The participant's ability to detect changes in contrast is determined. Higher mean indicates improved contrast sensitivity.

  5. Monocular Mesopic Contrast Sensitivity Without Glare(1.5, 3 Cycles/Degree) [Postoperative visit 4 (day 120-180)]

    A single eye (monocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness). The participant's ability to detect changes in contrast is determined. Higher mean indicates improved contrast sensitivity.

  6. Monocular Mesopic Contrast Sensitivity Without Glare(6 Cycles/Degree) [Postoperative visit 4 (day 120-180)]

    A single eye (monocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness). The participant's ability to detect changes in contrast is determined. Higher mean indicates improved contrast sensitivity.

  7. Monocular Mesopic Contrast Sensitivity Without Glare(12 Cycles/Degree) [Postoperative visit 4 (day 120-180)]

    A single eye (monocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness). The participant's ability to detect changes in contrast is determined. Higher mean indicates improved contrast sensitivity.

  8. Monocular Mesopic Contrast Sensitivity Without Glare(18 Cycles/Degree) [Postoperative visit 4 (day 120-180)]

    A single eye (monocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness). The participant's ability to detect changes in contrast is determined. Higher mean indicates improved contrast sensitivity.

Secondary Outcome Measures

  1. Binocular Mesopic Contrast Sensitivity Without Glare(1.5, 3 Cycles/Degree) [Postoperative visit 3 (day 30-60)]

    In two eye (Binocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness). The participant's ability to detect changes in contrast is determined. Higher mean indicates improved contrast sensitivity.

  2. Binocular Mesopic Contrast Sensitivity Without Glare(6 Cycles/Degree) [Postoperative visit 3 (day 30-60)]

    In two eyes (binocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness). The participant's ability to detect changes in contrast is determined. Higher mean indicates improved contrast sensitivity.

  3. Binocular Mesopic Contrast Sensitivity Without Glare(12 Cycles/Degree) [Postoperative visit 3 (day 30-60)]

    In two eyes (binocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness). The participant's ability to detect changes in contrast is determined. Higher mean indicates improved contrast sensitivity.

  4. Binocular Mesopic Contrast Sensitivity Without Glare(18 Cycles/Degree) [Postoperative visit 3 (day 30-60)]

    In two eyes (binocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness). The participant's ability to detect changes in contrast is determined. Higher mean indicates improved contrast sensitivity.

  5. Binocular Mesopic Contrast Sensitivity Without Glare(1.5, 3 Cycles/Degree) [Postoperative visit 4 (day 120-180)]

    In two eyes (binocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness). The participant's ability to detect changes in contrast is determined. Higher mean indicates improved contrast sensitivity.

  6. Binocular Mesopic Contrast Sensitivity Without Glare(6 Cycles/Degree) [Postoperative visit 4 (day 120-180)]

    In two eyes (binocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness). The participant's ability to detect changes in contrast is determined. Higher mean indicates improved contrast sensitivity.

  7. Binocular Mesopic Contrast Sensitivity Without Glare(12 Cycles/Degree) [Postoperative visit 4 (day 120-180)]

    In two eyes (binocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness). The participant's ability to detect changes in contrast is determined. Higher mean indicates improved contrast sensitivity.

  8. Binocular Mesopic Contrast Sensitivity Without Glare(18 Cycles/Degree) [Postoperative visit 4 (day 120-180)]

    In two eyes (binocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness). The participant's ability to detect changes in contrast is determined. Higher mean indicates improved contrast sensitivity.

  9. Monocular Mesopic Contrast Sensitivity With Glare(1.5 Cycles/Degree) [Postoperative visit 3 (day 30-60)]

    In one eye (monocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness). The participant's ability to detect changes in contrast is determined. Higher mean indicates improved contrast sensitivity.

  10. Monocular Mesopic Contrast Sensitivity With Glare (3 Cycles/Degree) [Postoperative visit 3 (day 30-60)]

    In one eye (monocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness). The participant's ability to detect changes in contrast is determined. Higher mean indicates improved contrast sensitivity.

  11. Monocular Mesopic Contrast Sensitivity With Glare(6 Cycles/Degree) [Postoperative visit 3 (day 30-60)]

    In one eye (monocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness). The participant's ability to detect changes in contrast is determined. Higher mean indicates improved contrast sensitivity.

  12. Monocular Mesopic Contrast Sensitivity With Glare(12 Cycles/Degree) [Postoperative visit 3 (day 30-60)]

    In one eye (monocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness). The participant's ability to detect changes in contrast is determined. Higher mean indicates improved contrast sensitivity.

  13. Monocular Mesopic Contrast Sensitivity With Glare(18 Cycles/Degree) [Postoperative visit 3 (day 30-60)]

    In one eye (monocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness). The participant's ability to detect changes in contrast is determined. Higher mean indicates improved contrast sensitivity.

  14. Monocular Mesopic Contrast Sensitivity With Glare(1.5 Cycles/Degree) [Postoperative visit 4 (day 120-180)]

    In one eye (monocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness). The participant's ability to detect changes in contrast is determined. Higher mean indicates improved contrast sensitivity.

  15. Monocular Mesopic Contrast Sensitivity With Glare(3 Cycles/Degree) [Postoperative visit 4 (day 120-180)]

    In one eye (monocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness). The participant's ability to detect changes in contrast is determined. Higher mean indicates improved contrast sensitivity.

  16. Monocular Mesopic Contrast Sensitivity With Glare(6 Cycles/Degree) [Postoperative visit 4 (day 120-180)]

    In one eye (monocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness). The participant's ability to detect changes in contrast is determined. Higher mean indicates improved contrast sensitivity.

  17. Monocular Mesopic Contrast Sensitivity With Glare(12 Cycles/Degree) [Postoperative visit 4 (day 120-180)]

    In one eye (monocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness). The participant's ability to detect changes in contrast is determined. Higher mean indicates improved contrast sensitivity.

  18. Monocular Mesopic Contrast Sensitivity With Glare(18 Cycles/Degree) [Postoperative visit 4 (day 120-180)]

    In one eye (monocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness). The participant's ability to detect changes in contrast is determined. Higher mean indicates improved contrast sensitivity.

Eligibility Criteria

Criteria

Ages Eligible for Study:
40 Years and Older
Sexes Eligible for Study:
All
Accepts Healthy Volunteers:
No
Inclusion Criteria:
  • Subjects must be undergoing primary IOL implantation for the correction of aphakia following continuous curvilinear anterior capsulotomy and phacoemulsification cataract extraction.

  • Subjects must require a spherical lens power from 10.00 diopters (D) to 33.00 D.

  • Subjects must have the potential for corrected distance visual acuity (CDVA) of 20/32 or better in both eyes.

  • Subjects must have stopped contact lens wear for at least two weeks for soft lens wearers or three weeks for gas permeable lens wearers prior to biometry and surgery.

  • At the time of surgery, subjects must have an intact centered capsulorhexis, intact posterior capsule, and no zonular rupture.

  • Subjects must have ≤ 1.25 D of preoperative corneal astigmatism in both eyes.

Exclusion Criteria:
  • Subjects with diagnoses of degenerative visual disorders (eg, macular degeneration, or other retinal disorders) that cause potential acuity losses to a level of 20/32 or worse.

  • Subjects with conditions with increased risk of zonular rupture, such as pseudoexfoliation syndrome.

  • Subjects who have any active inflammation or edema (swelling) of the cornea, including but not limited to the following: keratitis, keratoconjunctivitis, and keratouveitis.

  • Subjects with uncontrolled glaucoma.

  • Subjects with previous retinal detachment.

  • Subjects with visually significant diabetic retinopathy (proliferative or nonproliferative) which reduces potential acuity to 20/32 or worse.

  • Subjects with rubella, bilateral congenital, traumatic, complicated or polar cataract.

  • Subjects with marked microphthalmos or aniridia.

  • Subjects who have had previous corneal surgery.

  • Subjects with irregular corneal astigmatism.

  • Subjects with amblyopia which reduces potential acuity to 20/32 or worse.

  • Subjects with optic atrophy.

  • Subjects with iris neovascularization.

  • Subjects with clinically significant retinal pigment epithelium/macular changes which reduces potential acuity to 20/32 or worse.

  • Subjects with chronic use of systemic steroids or immunosuppressive medications.

Contacts and Locations

Locations

Site City State Country Postal Code
1 Asian Eye Institute Makati City Philippines 1200

Sponsors and Collaborators

  • Bausch & Lomb Incorporated

Investigators

  • Study Director: Tes Ignacio, MD, Bausch & Lomb Incorporated

Study Documents (Full-Text)

None provided.

More Information

Publications

None provided.
Responsible Party:
Bausch & Lomb Incorporated
ClinicalTrials.gov Identifier:
NCT01225952
Other Study ID Numbers:
  • 656
First Posted:
Oct 21, 2010
Last Update Posted:
Jan 18, 2013
Last Verified:
Dec 1, 2012
Keywords provided by Bausch & Lomb Incorporated
Additional relevant MeSH terms:

Study Results

Participant Flow

Recruitment Details This study was an evaluation of three United States Food and Drug Administration (FDA) approved IOLs. A total of 93 participants(186 eyes) were enrolled at one investigative site in the Philippines. First participant was enrolled 7/15/2010 and last participant exited the study 4/28/2011.
Pre-assignment Detail 186 participant eyes were enrolled in this study 154 participant eyes completed the study. Participants underwent bilateral phacoemulsification and IOL implantation. Participants were randomized to treatment groups prior to surgery.
Arm/Group Title Crystalens AO ReSTOR 3.0 AMO Tecnis Multifocal
Arm/Group Description Bausch & Lomb silicone multi-piece accommodating IOL is a modified plate haptic lens An aspheric multifocal IOL (Alcon Laboratories) combines the functions of an apodized diffractive region and a refractive region. A foldable hydrophobic acrylic IOL,(Abbott Medical Optics), is an ultraviolet light-absorbing posterior chamber IOL
Period Title: Overall Study
STARTED 31 31 31
COMPLETED 25 27 25
NOT COMPLETED 6 4 6

Baseline Characteristics

Arm/Group Title Crystalens AO ReSTOR 3.0 AMO Tecnis Multifocal Total
Arm/Group Description Bausch & Lomb silicone multi-piece accommodating IOL is a modified plate haptic lens An aspheric multifocal IOL (Alcon Laboratories) combines the functions of an apodized diffractive region and a refractive region. A foldable hydrophobic acrylic IOL,(Abbott Medical Optics), is an ultraviolet light-absorbing posterior chamber IOL Total of all reporting groups
Overall Participants 31 31 31 93
Age, Customized (participants) [Number]
< 40 years
0
0%
0
0%
0
0%
0
0%
40 - 49 years
2
6.5%
1
3.2%
0
0%
3
3.2%
50 - 59 years
6
19.4%
8
25.8%
7
22.6%
21
22.6%
60 - 69 years
13
41.9%
14
45.2%
16
51.6%
43
46.2%
70 - 79 years
10
32.3%
8
25.8%
7
22.6%
25
26.9%
>/= 80 years
0
0%
0
0%
1
3.2%
1
1.1%
Sex: Female, Male (Count of Participants)
Female
23
74.2%
25
80.6%
22
71%
70
75.3%
Male
8
25.8%
6
19.4%
9
29%
23
24.7%
Region of Enrollment (participants) [Number]
Philippines
31
100%
31
100%
31
100%
93
100%

Outcome Measures

1. Primary Outcome
Title Monocular Mesopic Contrast Sensitivity Without Glare (1.5, 3 Cycles/Degree)
Description A single eye (monocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness). The participant's ability to detect changes in contrast is determined. Higher mean indicates improved contrast sensitivity.
Time Frame Postoperative visit 3 (day 30-60)

Outcome Measure Data

Analysis Population Description
Full Analysis Set
Arm/Group Title Crystalens AO ReSTOR 3.0 AMO Tecnis Multifocal
Arm/Group Description Bausch & Lomb silicone multi-piece accommodating IOL is a modified plate haptic lens An aspheric multifocal IOL (Alcon Laboratories) combines the functions of an apodized diffractive region and a refractive region. A foldable hydrophobic acrylic IOL,(Abbott Medical Optics), is an ultraviolet light-absorbing posterior chamber IOL
Measure Participants 26 27 25
Measure eyes 52 54 50
1.5 cycles/degree
1.612
(0.229)
1.582
(0.231)
1.527
(0.248)
3 cycles/degree
1.658
(0.195)
1.627
(0.224)
1.629
(0.240)
2. Primary Outcome
Title Monocular Mesopic Contrast Sensitivity Without Glare(6 Cycles/Degree)
Description A single eye (monocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness). The participant's ability to detect changes in contrast is determined. Higher mean indicates improved contrast sensitivity.
Time Frame Postoperative visit 3 (day 30-60)

Outcome Measure Data

Analysis Population Description
Full Analysis Set, non-missing
Arm/Group Title Crystalens AO ReSTOR 3.0 AMO Tecnis Multifocal
Arm/Group Description Bausch & Lomb silicone multi-piece accommodating IOL is a modified plate haptic lens An aspheric multifocal IOL (Alcon Laboratories) combines the functions of an apodized diffractive region and a refractive region. A foldable hydrophobic acrylic IOL,(Abbott Medical Optics), is an ultraviolet light-absorbing posterior chamber IOL
Measure Participants 26 25 24
Measure eyes 51 50 48
Mean (Standard Deviation) [log contrast sensitivity units]
1.641
(0.210)
1.611
(0.237)
1.641
(0.286)
3. Secondary Outcome
Title Binocular Mesopic Contrast Sensitivity Without Glare(1.5, 3 Cycles/Degree)
Description In two eye (Binocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness). The participant's ability to detect changes in contrast is determined. Higher mean indicates improved contrast sensitivity.
Time Frame Postoperative visit 3 (day 30-60)

Outcome Measure Data

Analysis Population Description
Full Analysis Set
Arm/Group Title Crystalens AO ReSTOR 3.0 AMO Tecnis Multifocal
Arm/Group Description Bausch & Lomb silicone multi-piece accommodating IOL is a modified plate haptic lens An aspheric multifocal IOL (Alcon Laboratories) combines the functions of an apodized diffractive region and a refractive region. A foldable hydrophobic acrylic IOL,(Abbott Medical Optics), is an ultraviolet light-absorbing posterior chamber IOL
Measure Participants 26 27 25
1.5 cycles/degree
1.770
(0.187)
1.656
(0.238)
1.605
(0.203)
3 cycles/degree
1.807
(0.199)
1.679
(0.275)
1.689
(0.189)
4. Secondary Outcome
Title Binocular Mesopic Contrast Sensitivity Without Glare(6 Cycles/Degree)
Description In two eyes (binocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness). The participant's ability to detect changes in contrast is determined. Higher mean indicates improved contrast sensitivity.
Time Frame Postoperative visit 3 (day 30-60)

Outcome Measure Data

Analysis Population Description
Full Analysis Set
Arm/Group Title Crystalens AO ReSTOR 3.0 AMO Tecnis Multifocal
Arm/Group Description Bausch & Lomb silicone multi-piece accommodating IOL is a modified plate haptic lens An aspheric multifocal IOL (Alcon Laboratories) combines the functions of an apodized diffractive region and a refractive region. A foldable hydrophobic acrylic IOL,(Abbott Medical Optics), is an ultraviolet light-absorbing posterior chamber IOL
Measure Participants 25 26 25
Mean (Standard Deviation) [log contrast sensitivity units]
1.825
(0.216)
1.591
(0.274)
1.700
(0.284)
5. Primary Outcome
Title Monocular Mesopic Contrast Sensitivity Without Glare(12 Cycles/Degree)
Description A single eye (monocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness). The participant's ability to detect changes in contrast is determined. Higher mean indicates improved contrast sensitivity.
Time Frame Postoperative visit 3 (day 30-60)

Outcome Measure Data

Analysis Population Description
Full Analysis Set, non-missing
Arm/Group Title Crystalens AO ReSTOR 3.0 AMO Tecnis Multifocal
Arm/Group Description Bausch & Lomb silicone multi-piece accommodating IOL is a modified plate haptic lens An aspheric multifocal IOL (Alcon Laboratories) combines the functions of an apodized diffractive region and a refractive region. A foldable hydrophobic acrylic IOL,(Abbott Medical Optics), is an ultraviolet light-absorbing posterior chamber IOL
Measure Participants 24 22 21
Measure eyes 48 43 42
Mean (Standard Deviation) [log contrast sensitivity units]
1.325
(0.226)
1.227
(0.198)
1.324
(0.281)
6. Primary Outcome
Title Monocular Mesopic Contrast Sensitivity Without Glare(18 Cycles/Degree)
Description A single eye (monocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness). The participant's ability to detect changes in contrast is determined. Higher mean indicates improved contrast sensitivity.
Time Frame Postoperative visit 3 (day 30-60)

Outcome Measure Data

Analysis Population Description
Full Analysis Set, non-missing
Arm/Group Title Crystalens AO ReSTOR 3.0 AMO Tecnis Multifocal
Arm/Group Description Bausch & Lomb silicone multi-piece accommodating IOL is a modified plate haptic lens An aspheric multifocal IOL (Alcon Laboratories) combines the functions of an apodized diffractive region and a refractive region. A foldable hydrophobic acrylic IOL,(Abbott Medical Optics), is an ultraviolet light-absorbing posterior chamber IOL
Measure Participants 17 13 15
Measure eyes 34 26 30
Mean (Standard Deviation) [log contrast sensitivity units]
0.977
(0.266)
0.985
(0.185)
0.973
(0.342)
7. Primary Outcome
Title Monocular Mesopic Contrast Sensitivity Without Glare(1.5, 3 Cycles/Degree)
Description A single eye (monocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness). The participant's ability to detect changes in contrast is determined. Higher mean indicates improved contrast sensitivity.
Time Frame Postoperative visit 4 (day 120-180)

Outcome Measure Data

Analysis Population Description
[Not Specified]
Arm/Group Title Crystalens AO ReSTOR 3.0 AMO Tecnis Multifocal
Arm/Group Description Bausch & Lomb silicone multi-piece accommodating IOL is a modified plate haptic lens An aspheric multifocal IOL (Alcon Laboratories) combines the functions of an apodized diffractive region and a refractive region. A foldable hydrophobic acrylic IOL,(Abbott Medical Optics), is an ultraviolet light-absorbing posterior chamber IOL
Measure Participants 25 27 25
Measure eyes 50 54 50
1.5 cycles/degree
1.696
(0.213)
1.621
(0.222)
1.527
(0.190)
3 cycles/degree
1.678
(0.159)
1.670
(0.236)
1.630
(0.203)
8. Primary Outcome
Title Monocular Mesopic Contrast Sensitivity Without Glare(6 Cycles/Degree)
Description A single eye (monocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness). The participant's ability to detect changes in contrast is determined. Higher mean indicates improved contrast sensitivity.
Time Frame Postoperative visit 4 (day 120-180)

Outcome Measure Data

Analysis Population Description
[Not Specified]
Arm/Group Title Crystalens AO ReSTOR 3.0 AMO Tecnis Multifocal
Arm/Group Description Bausch & Lomb silicone multi-piece accommodating IOL is a modified plate haptic lens An aspheric multifocal IOL (Alcon Laboratories) combines the functions of an apodized diffractive region and a refractive region. A foldable hydrophobic acrylic IOL,(Abbott Medical Optics), is an ultraviolet light-absorbing posterior chamber IOL
Measure Participants 24 26 24
Measure eyes 48 51 48
Mean (Standard Deviation) [log contrast sensitivity units]
1.579
(0.239)
1.499
(0.230)
1.610
(0.228)
9. Primary Outcome
Title Monocular Mesopic Contrast Sensitivity Without Glare(12 Cycles/Degree)
Description A single eye (monocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness). The participant's ability to detect changes in contrast is determined. Higher mean indicates improved contrast sensitivity.
Time Frame Postoperative visit 4 (day 120-180)

Outcome Measure Data

Analysis Population Description
[Not Specified]
Arm/Group Title Crystalens AO ReSTOR 3.0 AMO Tecnis Multifocal
Arm/Group Description Bausch & Lomb silicone multi-piece accommodating IOL is a modified plate haptic lens An aspheric multifocal IOL (Alcon Laboratories) combines the functions of an apodized diffractive region and a refractive region. A foldable hydrophobic acrylic IOL,(Abbott Medical Optics), is an ultraviolet light-absorbing posterior chamber IOL
Measure Participants 20 18 20
Measure eyes 39 35 40
Mean (Standard Deviation) [log contrast sensitivity units]
1.280
(0.245)
1.225
(0.253)
1.230
(0.248)
10. Primary Outcome
Title Monocular Mesopic Contrast Sensitivity Without Glare(18 Cycles/Degree)
Description A single eye (monocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness). The participant's ability to detect changes in contrast is determined. Higher mean indicates improved contrast sensitivity.
Time Frame Postoperative visit 4 (day 120-180)

Outcome Measure Data

Analysis Population Description
[Not Specified]
Arm/Group Title Crystalens AO ReSTOR 3.0 AMO Tecnis Multifocal
Arm/Group Description Bausch & Lomb silicone multi-piece accommodating IOL is a modified plate haptic lens An aspheric multifocal IOL (Alcon Laboratories) combines the functions of an apodized diffractive region and a refractive region. A foldable hydrophobic acrylic IOL,(Abbott Medical Optics), is an ultraviolet light-absorbing posterior chamber IOL
Measure Participants 13 10 9
Measure eyes 25 20 18
Mean (Standard Deviation) [log contrast sensitivity units]
0.974
(0.303)
0.980
(0.246)
0.976
(0.207)
11. Secondary Outcome
Title Binocular Mesopic Contrast Sensitivity Without Glare(12 Cycles/Degree)
Description In two eyes (binocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness). The participant's ability to detect changes in contrast is determined. Higher mean indicates improved contrast sensitivity.
Time Frame Postoperative visit 3 (day 30-60)

Outcome Measure Data

Analysis Population Description
Full Analysis Set
Arm/Group Title Crystalens AO ReSTOR 3.0 AMO Tecnis Multifocal
Arm/Group Description Bausch & Lomb silicone multi-piece accommodating IOL is a modified plate haptic lens An aspheric multifocal IOL (Alcon Laboratories) combines the functions of an apodized diffractive region and a refractive region. A foldable hydrophobic acrylic IOL,(Abbott Medical Optics), is an ultraviolet light-absorbing posterior chamber IOL
Measure Participants 25 21 22
Mean (Standard Deviation) [log contrast sensitivity units]
1.425
(0.221)
1.335
(0.218)
1.361
(0.278)
12. Secondary Outcome
Title Binocular Mesopic Contrast Sensitivity Without Glare(18 Cycles/Degree)
Description In two eyes (binocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness). The participant's ability to detect changes in contrast is determined. Higher mean indicates improved contrast sensitivity.
Time Frame Postoperative visit 3 (day 30-60)

Outcome Measure Data

Analysis Population Description
Full Analysis Set
Arm/Group Title Crystalens AO ReSTOR 3.0 AMO Tecnis Multifocal
Arm/Group Description Bausch & Lomb silicone multi-piece accommodating IOL is a modified plate haptic lens An aspheric multifocal IOL (Alcon Laboratories) combines the functions of an apodized diffractive region and a refractive region. A foldable hydrophobic acrylic IOL,(Abbott Medical Optics), is an ultraviolet light-absorbing posterior chamber IOL
Measure Participants 19 15 16
Mean (Standard Deviation) [log contrast sensitivity units]
1.029
(0.250)
1.001
(0.213)
0.981
(0.362)
13. Secondary Outcome
Title Binocular Mesopic Contrast Sensitivity Without Glare(1.5, 3 Cycles/Degree)
Description In two eyes (binocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness). The participant's ability to detect changes in contrast is determined. Higher mean indicates improved contrast sensitivity.
Time Frame Postoperative visit 4 (day 120-180)

Outcome Measure Data

Analysis Population Description
Full Analysis Set
Arm/Group Title Crystalens AO ReSTOR 3.0 AMO Tecnis Multifocal
Arm/Group Description Bausch & Lomb silicone multi-piece accommodating IOL is a modified plate haptic lens An aspheric multifocal IOL (Alcon Laboratories) combines the functions of an apodized diffractive region and a refractive region. A foldable hydrophobic acrylic IOL,(Abbott Medical Optics), is an ultraviolet light-absorbing posterior chamber IOL
Measure Participants 25 27 25
1.5 cycles/degree
1.797
(0.183)
1.751
(0.238)
1.577
(0.206)
3 cycles/degree
1.851
(0.178)
1.723
(0.239)
1.702
(0.219)
14. Secondary Outcome
Title Binocular Mesopic Contrast Sensitivity Without Glare(6 Cycles/Degree)
Description In two eyes (binocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness). The participant's ability to detect changes in contrast is determined. Higher mean indicates improved contrast sensitivity.
Time Frame Postoperative visit 4 (day 120-180)

Outcome Measure Data

Analysis Population Description
Full Analysis Set
Arm/Group Title Crystalens AO ReSTOR 3.0 AMO Tecnis Multifocal
Arm/Group Description Bausch & Lomb silicone multi-piece accommodating IOL is a modified plate haptic lens An aspheric multifocal IOL (Alcon Laboratories) combines the functions of an apodized diffractive region and a refractive region. A foldable hydrophobic acrylic IOL,(Abbott Medical Optics), is an ultraviolet light-absorbing posterior chamber IOL
Measure Participants 24 25 24
Mean (Standard Deviation) [log contrast sensitivity units]
1.758
(0.280)
1.640
(0.236)
1.690
(0.263)
15. Secondary Outcome
Title Binocular Mesopic Contrast Sensitivity Without Glare(12 Cycles/Degree)
Description In two eyes (binocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness). The participant's ability to detect changes in contrast is determined. Higher mean indicates improved contrast sensitivity.
Time Frame Postoperative visit 4 (day 120-180)

Outcome Measure Data

Analysis Population Description
Full Analysis Set
Arm/Group Title Crystalens AO ReSTOR 3.0 AMO Tecnis Multifocal
Arm/Group Description Bausch & Lomb silicone multi-piece accommodating IOL is a modified plate haptic lens An aspheric multifocal IOL (Alcon Laboratories) combines the functions of an apodized diffractive region and a refractive region. A foldable hydrophobic acrylic IOL,(Abbott Medical Optics), is an ultraviolet light-absorbing posterior chamber IOL
Measure Participants 21 18 22
Mean (Standard Deviation) [log contrast sensitivity units]
1.371
(0.247)
1.269
(0.257)
1.310
(0.286)
16. Secondary Outcome
Title Binocular Mesopic Contrast Sensitivity Without Glare(18 Cycles/Degree)
Description In two eyes (binocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness). The participant's ability to detect changes in contrast is determined. Higher mean indicates improved contrast sensitivity.
Time Frame Postoperative visit 4 (day 120-180)

Outcome Measure Data

Analysis Population Description
Full Analysis Set
Arm/Group Title Crystalens AO ReSTOR 3.0 AMO Tecnis Multifocal
Arm/Group Description Bausch & Lomb silicone multi-piece accommodating IOL is a modified plate haptic lens An aspheric multifocal IOL (Alcon Laboratories) combines the functions of an apodized diffractive region and a refractive region. A foldable hydrophobic acrylic IOL,(Abbott Medical Optics), is an ultraviolet light-absorbing posterior chamber IOL
Measure Participants 16 12 12
Mean (Standard Deviation) [log contrast sensitivity units]
1.115
(0.299)
0.995
(0.384)
1.041
(0.330)
17. Secondary Outcome
Title Monocular Mesopic Contrast Sensitivity With Glare(1.5 Cycles/Degree)
Description In one eye (monocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness). The participant's ability to detect changes in contrast is determined. Higher mean indicates improved contrast sensitivity.
Time Frame Postoperative visit 3 (day 30-60)

Outcome Measure Data

Analysis Population Description
Full analysis set
Arm/Group Title Crystalens AO ReSTOR 3.0 AMO Tecnis Multifocal
Arm/Group Description Bausch & Lomb silicone multi-piece accommodating IOL is a modified plate haptic lens An aspheric multifocal IOL (Alcon Laboratories) combines the functions of an apodized diffractive region and a refractive region. A foldable hydrophobic acrylic IOL,(Abbott Medical Optics), is an ultraviolet light-absorbing posterior chamber IOL
Measure Participants 26 25 25
Measure eyes 52 50 49
Mean (Standard Deviation) [log contrast sensitivity units]
1.591
(0.215)
1.519
(0.218)
1.482
(0.268)
18. Secondary Outcome
Title Monocular Mesopic Contrast Sensitivity With Glare (3 Cycles/Degree)
Description In one eye (monocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness). The participant's ability to detect changes in contrast is determined. Higher mean indicates improved contrast sensitivity.
Time Frame Postoperative visit 3 (day 30-60)

Outcome Measure Data

Analysis Population Description
Full analysis set
Arm/Group Title Crystalens AO ReSTOR 3.0 AMO Tecnis Multifocal
Arm/Group Description Bausch & Lomb silicone multi-piece accommodating IOL is a modified plate haptic lens An aspheric multifocal IOL (Alcon Laboratories) combines the functions of an apodized diffractive region and a refractive region. A foldable hydrophobic acrylic IOL,(Abbott Medical Optics), is an ultraviolet light-absorbing posterior chamber IOL
Measure Participants 26 25 24
Measure eyes 52 50 47
Mean (Standard Deviation) [log contrast sensitivity units]
1.687
(0.229)
1.581
(0.204)
1.588
(0.264)
19. Secondary Outcome
Title Monocular Mesopic Contrast Sensitivity With Glare(6 Cycles/Degree)
Description In one eye (monocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness). The participant's ability to detect changes in contrast is determined. Higher mean indicates improved contrast sensitivity.
Time Frame Postoperative visit 3 (day 30-60)

Outcome Measure Data

Analysis Population Description
Full analysis set
Arm/Group Title Crystalens AO ReSTOR 3.0 AMO Tecnis Multifocal
Arm/Group Description Bausch & Lomb silicone multi-piece accommodating IOL is a modified plate haptic lens An aspheric multifocal IOL (Alcon Laboratories) combines the functions of an apodized diffractive region and a refractive region. A foldable hydrophobic acrylic IOL,(Abbott Medical Optics), is an ultraviolet light-absorbing posterior chamber IOL
Measure Participants 25 25 23
Measure eyes 50 49 45
Mean (Standard Deviation) [log contrast sensitivity units]
1.674
(0.237)
1.531
(0.238)
1.626
(0.293)
20. Secondary Outcome
Title Monocular Mesopic Contrast Sensitivity With Glare(12 Cycles/Degree)
Description In one eye (monocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness). The participant's ability to detect changes in contrast is determined. Higher mean indicates improved contrast sensitivity.
Time Frame Postoperative visit 3 (day 30-60)

Outcome Measure Data

Analysis Population Description
Full analysis set
Arm/Group Title Crystalens AO ReSTOR 3.0 AMO Tecnis Multifocal
Arm/Group Description Bausch & Lomb silicone multi-piece accommodating IOL is a modified plate haptic lens An aspheric multifocal IOL (Alcon Laboratories) combines the functions of an apodized diffractive region and a refractive region. A foldable hydrophobic acrylic IOL,(Abbott Medical Optics), is an ultraviolet light-absorbing posterior chamber IOL
Measure Participants 23 18 17
Measure eyes 45 35 34
Mean (Standard Deviation) [log contrast sensitivity units]
1.328
(0.224)
1.230
(0.223)
1.340
(0.268)
21. Secondary Outcome
Title Monocular Mesopic Contrast Sensitivity With Glare(18 Cycles/Degree)
Description In one eye (monocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness). The participant's ability to detect changes in contrast is determined. Higher mean indicates improved contrast sensitivity.
Time Frame Postoperative visit 3 (day 30-60)

Outcome Measure Data

Analysis Population Description
Full analysis set
Arm/Group Title Crystalens AO ReSTOR 3.0 AMO Tecnis Multifocal
Arm/Group Description Bausch & Lomb silicone multi-piece accommodating IOL is a modified plate haptic lens An aspheric multifocal IOL (Alcon Laboratories) combines the functions of an apodized diffractive region and a refractive region. A foldable hydrophobic acrylic IOL,(Abbott Medical Optics), is an ultraviolet light-absorbing posterior chamber IOL
Measure Participants 14 12 13
Measure eyes 27 24 25
Mean (Standard Deviation) [log contrast sensitivity units]
1.024
(0.196)
0.945
(0.158)
0.834
(0.283)
22. Secondary Outcome
Title Monocular Mesopic Contrast Sensitivity With Glare(1.5 Cycles/Degree)
Description In one eye (monocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness). The participant's ability to detect changes in contrast is determined. Higher mean indicates improved contrast sensitivity.
Time Frame Postoperative visit 4 (day 120-180)

Outcome Measure Data

Analysis Population Description
Full analysis set
Arm/Group Title Crystalens AO ReSTOR 3.0 AMO Tecnis Multifocal
Arm/Group Description Bausch & Lomb silicone multi-piece accommodating IOL is a modified plate haptic lens An aspheric multifocal IOL (Alcon Laboratories) combines the functions of an apodized diffractive region and a refractive region. A foldable hydrophobic acrylic IOL,(Abbott Medical Optics), is an ultraviolet light-absorbing posterior chamber IOL
Measure Participants 25 25 24
Measure eyes 50 50 48
Mean (Standard Deviation) [log contrast sensitivity units]
1.550
(0.263)
1.488
(0.226)
1.449
(0.206)
23. Secondary Outcome
Title Monocular Mesopic Contrast Sensitivity With Glare(3 Cycles/Degree)
Description In one eye (monocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness). The participant's ability to detect changes in contrast is determined. Higher mean indicates improved contrast sensitivity.
Time Frame Postoperative visit 4 (day 120-180)

Outcome Measure Data

Analysis Population Description
Full analysis set
Arm/Group Title Crystalens AO ReSTOR 3.0 AMO Tecnis Multifocal
Arm/Group Description Bausch & Lomb silicone multi-piece accommodating IOL is a modified plate haptic lens An aspheric multifocal IOL (Alcon Laboratories) combines the functions of an apodized diffractive region and a refractive region. A foldable hydrophobic acrylic IOL,(Abbott Medical Optics), is an ultraviolet light-absorbing posterior chamber IOL
Measure Participants 25 25 25
Measure eyes 50 50 49
Mean (Standard Deviation) [log contrast sensitivity units]
1.664
(0.258)
1.572
(0.225)
1.535
(0.220)
24. Secondary Outcome
Title Monocular Mesopic Contrast Sensitivity With Glare(6 Cycles/Degree)
Description In one eye (monocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness). The participant's ability to detect changes in contrast is determined. Higher mean indicates improved contrast sensitivity.
Time Frame Postoperative visit 4 (day 120-180)

Outcome Measure Data

Analysis Population Description
Full analysis set
Arm/Group Title Crystalens AO ReSTOR 3.0 AMO Tecnis Multifocal
Arm/Group Description Bausch & Lomb silicone multi-piece accommodating IOL is a modified plate haptic lens An aspheric multifocal IOL (Alcon Laboratories) combines the functions of an apodized diffractive region and a refractive region. A foldable hydrophobic acrylic IOL,(Abbott Medical Optics), is an ultraviolet light-absorbing posterior chamber IOL
Measure Participants 24 23 23
Measure eyes 48 46 45
Mean (Standard Deviation) [log contrast sensitivity units]
1.578
(0.252)
1.493
(0.267)
1.519
(0.260)
25. Secondary Outcome
Title Monocular Mesopic Contrast Sensitivity With Glare(12 Cycles/Degree)
Description In one eye (monocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness). The participant's ability to detect changes in contrast is determined. Higher mean indicates improved contrast sensitivity.
Time Frame Postoperative visit 4 (day 120-180)

Outcome Measure Data

Analysis Population Description
Full analysis set
Arm/Group Title Crystalens AO ReSTOR 3.0 AMO Tecnis Multifocal
Arm/Group Description Bausch & Lomb silicone multi-piece accommodating IOL is a modified plate haptic lens An aspheric multifocal IOL (Alcon Laboratories) combines the functions of an apodized diffractive region and a refractive region. A foldable hydrophobic acrylic IOL,(Abbott Medical Optics), is an ultraviolet light-absorbing posterior chamber IOL
Measure Participants 20 15 16
Measure eyes 39 30 31
Mean (Standard Deviation) [log contrast sensitivity units]
1.335
(0.298)
1.246
(0.278)
1.194
(0.237)
26. Secondary Outcome
Title Monocular Mesopic Contrast Sensitivity With Glare(18 Cycles/Degree)
Description In one eye (monocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness). The participant's ability to detect changes in contrast is determined. Higher mean indicates improved contrast sensitivity.
Time Frame Postoperative visit 4 (day 120-180)

Outcome Measure Data

Analysis Population Description
Full analysis set
Arm/Group Title Crystalens AO ReSTOR 3.0 AMO Tecnis Multifocal
Arm/Group Description Bausch & Lomb silicone multi-piece accommodating IOL is a modified plate haptic lens An aspheric multifocal IOL (Alcon Laboratories) combines the functions of an apodized diffractive region and a refractive region. A foldable hydrophobic acrylic IOL,(Abbott Medical Optics), is an ultraviolet light-absorbing posterior chamber IOL
Measure Participants 13 8 8
Measure eyes 25 16 16
Mean (Standard Deviation) [log contrast sensitivity units]
0.919
(0.283)
1.019
(0.367)
0.935
(0.207)

Adverse Events

Time Frame 120-180 days
Adverse Event Reporting Description Adverse events measured by eye
Arm/Group Title Crystalens AO ReSTOR 3.0 AMO Tecnis Multifocal
Arm/Group Description Bausch & Lomb silicone multi-piece accommodating IOL is a modified plate haptic lens An aspheric multifocal IOL (Alcon Laboratories) combines the functions of an apodized diffractive region and a refractive region. A foldable hydrophobic acrylic IOL,(Abbott Medical Optics), is an ultraviolet light-absorbing posterior chamber IOL
All Cause Mortality
Crystalens AO ReSTOR 3.0 AMO Tecnis Multifocal
Affected / at Risk (%) # Events Affected / at Risk (%) # Events Affected / at Risk (%) # Events
Total / (NaN) / (NaN) / (NaN)
Serious Adverse Events
Crystalens AO ReSTOR 3.0 AMO Tecnis Multifocal
Affected / at Risk (%) # Events Affected / at Risk (%) # Events Affected / at Risk (%) # Events
Total 0/53 (0%) 0/54 (0%) 0/51 (0%)
Other (Not Including Serious) Adverse Events
Crystalens AO ReSTOR 3.0 AMO Tecnis Multifocal
Affected / at Risk (%) # Events Affected / at Risk (%) # Events Affected / at Risk (%) # Events
Total 12/53 (22.6%) 5/54 (9.3%) 6/51 (11.8%)
Eye disorders
Increased Intraocular Pressure 11/53 (20.8%) 11 5/54 (9.3%) 5 3/51 (5.9%) 3
Cystoid Macular Edema 1/53 (1.9%) 1 0/54 (0%) 0 3/51 (5.9%) 3

Limitations/Caveats

[Not Specified]

More Information

Certain Agreements

Principal Investigators are NOT employed by the organization sponsoring the study.

All study data generated as a result of this study will be regarded as confidential, until appropriate analysis and review by the Sponsor or its designee and the Investigator(s) are completed. The results of the study may be published or presented by the Investigator(s)after the review by, and in consultation and agreement with the Sponsor.

Results Point of Contact

Name/Title Omid Khodai, OD,RAC
Organization Bausch & Lomb Incorporated
Phone (949) 521-7894
Email omid.khodai@bausch.com
Responsible Party:
Bausch & Lomb Incorporated
ClinicalTrials.gov Identifier:
NCT01225952
Other Study ID Numbers:
  • 656
First Posted:
Oct 21, 2010
Last Update Posted:
Jan 18, 2013
Last Verified:
Dec 1, 2012