Three Intraocular Lenses Designed to Improve Distance, Intermediate and Near Vision Following Lens Extraction
Study Details
Study Description
Brief Summary
The objective of this study is to compare the contrast sensitivity, high and low contrast visual acuity (VA), glare meter outcomes, and subject satisfaction with three different FDA-approved intraocular lenses (IOLs) designed to improve distance, intermediate, and near vision following lens extraction in adults at least 40 years of age.
Condition or Disease | Intervention/Treatment | Phase |
---|---|---|
|
Phase 4 |
Study Design
Arms and Interventions
Arm | Intervention/Treatment |
---|---|
Experimental: Crystalens AO Bausch & Lomb silicone multi-piece accommodating IOL is a modified plate haptic lens |
Device: Crystalens AO
Bausch & Lomb model silicone multi-piece accommodating IOL.
|
Active Comparator: ReSTOR 3.0 An aspheric multifocal IOL (Alcon Laboratories) combines the functions of an apodized diffractive region and a refractive region. |
Device: ReSTOR 3.0
An aspheric multifocal IOL (Alcon Laboratories) combines the functions of an apodized diffractive region and a refractive region.
|
Active Comparator: AMO Tecnis Multifocal A foldable hydrophobic acrylic IOL,(Abbott Medical Optics), is an ultraviolet light-absorbing posterior chamber IOL |
Device: AMO Tecnis Multifocal
The Tecnis Multifocal foldable hydrophobic acrylic IOL, Model ZMA00 (Abbott Medical Optics), is an ultraviolet light-absorbing posterior chamber IOL.
|
Outcome Measures
Primary Outcome Measures
- Monocular Mesopic Contrast Sensitivity Without Glare (1.5, 3 Cycles/Degree) [Postoperative visit 3 (day 30-60)]
A single eye (monocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness). The participant's ability to detect changes in contrast is determined. Higher mean indicates improved contrast sensitivity.
- Monocular Mesopic Contrast Sensitivity Without Glare(6 Cycles/Degree) [Postoperative visit 3 (day 30-60)]
A single eye (monocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness). The participant's ability to detect changes in contrast is determined. Higher mean indicates improved contrast sensitivity.
- Monocular Mesopic Contrast Sensitivity Without Glare(12 Cycles/Degree) [Postoperative visit 3 (day 30-60)]
A single eye (monocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness). The participant's ability to detect changes in contrast is determined. Higher mean indicates improved contrast sensitivity.
- Monocular Mesopic Contrast Sensitivity Without Glare(18 Cycles/Degree) [Postoperative visit 3 (day 30-60)]
A single eye (monocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness). The participant's ability to detect changes in contrast is determined. Higher mean indicates improved contrast sensitivity.
- Monocular Mesopic Contrast Sensitivity Without Glare(1.5, 3 Cycles/Degree) [Postoperative visit 4 (day 120-180)]
A single eye (monocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness). The participant's ability to detect changes in contrast is determined. Higher mean indicates improved contrast sensitivity.
- Monocular Mesopic Contrast Sensitivity Without Glare(6 Cycles/Degree) [Postoperative visit 4 (day 120-180)]
A single eye (monocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness). The participant's ability to detect changes in contrast is determined. Higher mean indicates improved contrast sensitivity.
- Monocular Mesopic Contrast Sensitivity Without Glare(12 Cycles/Degree) [Postoperative visit 4 (day 120-180)]
A single eye (monocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness). The participant's ability to detect changes in contrast is determined. Higher mean indicates improved contrast sensitivity.
- Monocular Mesopic Contrast Sensitivity Without Glare(18 Cycles/Degree) [Postoperative visit 4 (day 120-180)]
A single eye (monocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness). The participant's ability to detect changes in contrast is determined. Higher mean indicates improved contrast sensitivity.
Secondary Outcome Measures
- Binocular Mesopic Contrast Sensitivity Without Glare(1.5, 3 Cycles/Degree) [Postoperative visit 3 (day 30-60)]
In two eye (Binocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness). The participant's ability to detect changes in contrast is determined. Higher mean indicates improved contrast sensitivity.
- Binocular Mesopic Contrast Sensitivity Without Glare(6 Cycles/Degree) [Postoperative visit 3 (day 30-60)]
In two eyes (binocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness). The participant's ability to detect changes in contrast is determined. Higher mean indicates improved contrast sensitivity.
- Binocular Mesopic Contrast Sensitivity Without Glare(12 Cycles/Degree) [Postoperative visit 3 (day 30-60)]
In two eyes (binocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness). The participant's ability to detect changes in contrast is determined. Higher mean indicates improved contrast sensitivity.
- Binocular Mesopic Contrast Sensitivity Without Glare(18 Cycles/Degree) [Postoperative visit 3 (day 30-60)]
In two eyes (binocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness). The participant's ability to detect changes in contrast is determined. Higher mean indicates improved contrast sensitivity.
- Binocular Mesopic Contrast Sensitivity Without Glare(1.5, 3 Cycles/Degree) [Postoperative visit 4 (day 120-180)]
In two eyes (binocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness). The participant's ability to detect changes in contrast is determined. Higher mean indicates improved contrast sensitivity.
- Binocular Mesopic Contrast Sensitivity Without Glare(6 Cycles/Degree) [Postoperative visit 4 (day 120-180)]
In two eyes (binocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness). The participant's ability to detect changes in contrast is determined. Higher mean indicates improved contrast sensitivity.
- Binocular Mesopic Contrast Sensitivity Without Glare(12 Cycles/Degree) [Postoperative visit 4 (day 120-180)]
In two eyes (binocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness). The participant's ability to detect changes in contrast is determined. Higher mean indicates improved contrast sensitivity.
- Binocular Mesopic Contrast Sensitivity Without Glare(18 Cycles/Degree) [Postoperative visit 4 (day 120-180)]
In two eyes (binocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness). The participant's ability to detect changes in contrast is determined. Higher mean indicates improved contrast sensitivity.
- Monocular Mesopic Contrast Sensitivity With Glare(1.5 Cycles/Degree) [Postoperative visit 3 (day 30-60)]
In one eye (monocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness). The participant's ability to detect changes in contrast is determined. Higher mean indicates improved contrast sensitivity.
- Monocular Mesopic Contrast Sensitivity With Glare (3 Cycles/Degree) [Postoperative visit 3 (day 30-60)]
In one eye (monocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness). The participant's ability to detect changes in contrast is determined. Higher mean indicates improved contrast sensitivity.
- Monocular Mesopic Contrast Sensitivity With Glare(6 Cycles/Degree) [Postoperative visit 3 (day 30-60)]
In one eye (monocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness). The participant's ability to detect changes in contrast is determined. Higher mean indicates improved contrast sensitivity.
- Monocular Mesopic Contrast Sensitivity With Glare(12 Cycles/Degree) [Postoperative visit 3 (day 30-60)]
In one eye (monocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness). The participant's ability to detect changes in contrast is determined. Higher mean indicates improved contrast sensitivity.
- Monocular Mesopic Contrast Sensitivity With Glare(18 Cycles/Degree) [Postoperative visit 3 (day 30-60)]
In one eye (monocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness). The participant's ability to detect changes in contrast is determined. Higher mean indicates improved contrast sensitivity.
- Monocular Mesopic Contrast Sensitivity With Glare(1.5 Cycles/Degree) [Postoperative visit 4 (day 120-180)]
In one eye (monocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness). The participant's ability to detect changes in contrast is determined. Higher mean indicates improved contrast sensitivity.
- Monocular Mesopic Contrast Sensitivity With Glare(3 Cycles/Degree) [Postoperative visit 4 (day 120-180)]
In one eye (monocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness). The participant's ability to detect changes in contrast is determined. Higher mean indicates improved contrast sensitivity.
- Monocular Mesopic Contrast Sensitivity With Glare(6 Cycles/Degree) [Postoperative visit 4 (day 120-180)]
In one eye (monocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness). The participant's ability to detect changes in contrast is determined. Higher mean indicates improved contrast sensitivity.
- Monocular Mesopic Contrast Sensitivity With Glare(12 Cycles/Degree) [Postoperative visit 4 (day 120-180)]
In one eye (monocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness). The participant's ability to detect changes in contrast is determined. Higher mean indicates improved contrast sensitivity.
- Monocular Mesopic Contrast Sensitivity With Glare(18 Cycles/Degree) [Postoperative visit 4 (day 120-180)]
In one eye (monocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness). The participant's ability to detect changes in contrast is determined. Higher mean indicates improved contrast sensitivity.
Eligibility Criteria
Criteria
Inclusion Criteria:
-
Subjects must be undergoing primary IOL implantation for the correction of aphakia following continuous curvilinear anterior capsulotomy and phacoemulsification cataract extraction.
-
Subjects must require a spherical lens power from 10.00 diopters (D) to 33.00 D.
-
Subjects must have the potential for corrected distance visual acuity (CDVA) of 20/32 or better in both eyes.
-
Subjects must have stopped contact lens wear for at least two weeks for soft lens wearers or three weeks for gas permeable lens wearers prior to biometry and surgery.
-
At the time of surgery, subjects must have an intact centered capsulorhexis, intact posterior capsule, and no zonular rupture.
-
Subjects must have ≤ 1.25 D of preoperative corneal astigmatism in both eyes.
Exclusion Criteria:
-
Subjects with diagnoses of degenerative visual disorders (eg, macular degeneration, or other retinal disorders) that cause potential acuity losses to a level of 20/32 or worse.
-
Subjects with conditions with increased risk of zonular rupture, such as pseudoexfoliation syndrome.
-
Subjects who have any active inflammation or edema (swelling) of the cornea, including but not limited to the following: keratitis, keratoconjunctivitis, and keratouveitis.
-
Subjects with uncontrolled glaucoma.
-
Subjects with previous retinal detachment.
-
Subjects with visually significant diabetic retinopathy (proliferative or nonproliferative) which reduces potential acuity to 20/32 or worse.
-
Subjects with rubella, bilateral congenital, traumatic, complicated or polar cataract.
-
Subjects with marked microphthalmos or aniridia.
-
Subjects who have had previous corneal surgery.
-
Subjects with irregular corneal astigmatism.
-
Subjects with amblyopia which reduces potential acuity to 20/32 or worse.
-
Subjects with optic atrophy.
-
Subjects with iris neovascularization.
-
Subjects with clinically significant retinal pigment epithelium/macular changes which reduces potential acuity to 20/32 or worse.
-
Subjects with chronic use of systemic steroids or immunosuppressive medications.
Contacts and Locations
Locations
Site | City | State | Country | Postal Code | |
---|---|---|---|---|---|
1 | Asian Eye Institute | Makati City | Philippines | 1200 |
Sponsors and Collaborators
- Bausch & Lomb Incorporated
Investigators
- Study Director: Tes Ignacio, MD, Bausch & Lomb Incorporated
Study Documents (Full-Text)
None provided.More Information
Publications
None provided.- 656
Study Results
Participant Flow
Recruitment Details | This study was an evaluation of three United States Food and Drug Administration (FDA) approved IOLs. A total of 93 participants(186 eyes) were enrolled at one investigative site in the Philippines. First participant was enrolled 7/15/2010 and last participant exited the study 4/28/2011. |
---|---|
Pre-assignment Detail | 186 participant eyes were enrolled in this study 154 participant eyes completed the study. Participants underwent bilateral phacoemulsification and IOL implantation. Participants were randomized to treatment groups prior to surgery. |
Arm/Group Title | Crystalens AO | ReSTOR 3.0 | AMO Tecnis Multifocal |
---|---|---|---|
Arm/Group Description | Bausch & Lomb silicone multi-piece accommodating IOL is a modified plate haptic lens | An aspheric multifocal IOL (Alcon Laboratories) combines the functions of an apodized diffractive region and a refractive region. | A foldable hydrophobic acrylic IOL,(Abbott Medical Optics), is an ultraviolet light-absorbing posterior chamber IOL |
Period Title: Overall Study | |||
STARTED | 31 | 31 | 31 |
COMPLETED | 25 | 27 | 25 |
NOT COMPLETED | 6 | 4 | 6 |
Baseline Characteristics
Arm/Group Title | Crystalens AO | ReSTOR 3.0 | AMO Tecnis Multifocal | Total |
---|---|---|---|---|
Arm/Group Description | Bausch & Lomb silicone multi-piece accommodating IOL is a modified plate haptic lens | An aspheric multifocal IOL (Alcon Laboratories) combines the functions of an apodized diffractive region and a refractive region. | A foldable hydrophobic acrylic IOL,(Abbott Medical Optics), is an ultraviolet light-absorbing posterior chamber IOL | Total of all reporting groups |
Overall Participants | 31 | 31 | 31 | 93 |
Age, Customized (participants) [Number] | ||||
< 40 years |
0
0%
|
0
0%
|
0
0%
|
0
0%
|
40 - 49 years |
2
6.5%
|
1
3.2%
|
0
0%
|
3
3.2%
|
50 - 59 years |
6
19.4%
|
8
25.8%
|
7
22.6%
|
21
22.6%
|
60 - 69 years |
13
41.9%
|
14
45.2%
|
16
51.6%
|
43
46.2%
|
70 - 79 years |
10
32.3%
|
8
25.8%
|
7
22.6%
|
25
26.9%
|
>/= 80 years |
0
0%
|
0
0%
|
1
3.2%
|
1
1.1%
|
Sex: Female, Male (Count of Participants) | ||||
Female |
23
74.2%
|
25
80.6%
|
22
71%
|
70
75.3%
|
Male |
8
25.8%
|
6
19.4%
|
9
29%
|
23
24.7%
|
Region of Enrollment (participants) [Number] | ||||
Philippines |
31
100%
|
31
100%
|
31
100%
|
93
100%
|
Outcome Measures
Title | Monocular Mesopic Contrast Sensitivity Without Glare (1.5, 3 Cycles/Degree) |
---|---|
Description | A single eye (monocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness). The participant's ability to detect changes in contrast is determined. Higher mean indicates improved contrast sensitivity. |
Time Frame | Postoperative visit 3 (day 30-60) |
Outcome Measure Data
Analysis Population Description |
---|
Full Analysis Set |
Arm/Group Title | Crystalens AO | ReSTOR 3.0 | AMO Tecnis Multifocal |
---|---|---|---|
Arm/Group Description | Bausch & Lomb silicone multi-piece accommodating IOL is a modified plate haptic lens | An aspheric multifocal IOL (Alcon Laboratories) combines the functions of an apodized diffractive region and a refractive region. | A foldable hydrophobic acrylic IOL,(Abbott Medical Optics), is an ultraviolet light-absorbing posterior chamber IOL |
Measure Participants | 26 | 27 | 25 |
Measure eyes | 52 | 54 | 50 |
1.5 cycles/degree |
1.612
(0.229)
|
1.582
(0.231)
|
1.527
(0.248)
|
3 cycles/degree |
1.658
(0.195)
|
1.627
(0.224)
|
1.629
(0.240)
|
Title | Monocular Mesopic Contrast Sensitivity Without Glare(6 Cycles/Degree) |
---|---|
Description | A single eye (monocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness). The participant's ability to detect changes in contrast is determined. Higher mean indicates improved contrast sensitivity. |
Time Frame | Postoperative visit 3 (day 30-60) |
Outcome Measure Data
Analysis Population Description |
---|
Full Analysis Set, non-missing |
Arm/Group Title | Crystalens AO | ReSTOR 3.0 | AMO Tecnis Multifocal |
---|---|---|---|
Arm/Group Description | Bausch & Lomb silicone multi-piece accommodating IOL is a modified plate haptic lens | An aspheric multifocal IOL (Alcon Laboratories) combines the functions of an apodized diffractive region and a refractive region. | A foldable hydrophobic acrylic IOL,(Abbott Medical Optics), is an ultraviolet light-absorbing posterior chamber IOL |
Measure Participants | 26 | 25 | 24 |
Measure eyes | 51 | 50 | 48 |
Mean (Standard Deviation) [log contrast sensitivity units] |
1.641
(0.210)
|
1.611
(0.237)
|
1.641
(0.286)
|
Title | Binocular Mesopic Contrast Sensitivity Without Glare(1.5, 3 Cycles/Degree) |
---|---|
Description | In two eye (Binocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness). The participant's ability to detect changes in contrast is determined. Higher mean indicates improved contrast sensitivity. |
Time Frame | Postoperative visit 3 (day 30-60) |
Outcome Measure Data
Analysis Population Description |
---|
Full Analysis Set |
Arm/Group Title | Crystalens AO | ReSTOR 3.0 | AMO Tecnis Multifocal |
---|---|---|---|
Arm/Group Description | Bausch & Lomb silicone multi-piece accommodating IOL is a modified plate haptic lens | An aspheric multifocal IOL (Alcon Laboratories) combines the functions of an apodized diffractive region and a refractive region. | A foldable hydrophobic acrylic IOL,(Abbott Medical Optics), is an ultraviolet light-absorbing posterior chamber IOL |
Measure Participants | 26 | 27 | 25 |
1.5 cycles/degree |
1.770
(0.187)
|
1.656
(0.238)
|
1.605
(0.203)
|
3 cycles/degree |
1.807
(0.199)
|
1.679
(0.275)
|
1.689
(0.189)
|
Title | Binocular Mesopic Contrast Sensitivity Without Glare(6 Cycles/Degree) |
---|---|
Description | In two eyes (binocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness). The participant's ability to detect changes in contrast is determined. Higher mean indicates improved contrast sensitivity. |
Time Frame | Postoperative visit 3 (day 30-60) |
Outcome Measure Data
Analysis Population Description |
---|
Full Analysis Set |
Arm/Group Title | Crystalens AO | ReSTOR 3.0 | AMO Tecnis Multifocal |
---|---|---|---|
Arm/Group Description | Bausch & Lomb silicone multi-piece accommodating IOL is a modified plate haptic lens | An aspheric multifocal IOL (Alcon Laboratories) combines the functions of an apodized diffractive region and a refractive region. | A foldable hydrophobic acrylic IOL,(Abbott Medical Optics), is an ultraviolet light-absorbing posterior chamber IOL |
Measure Participants | 25 | 26 | 25 |
Mean (Standard Deviation) [log contrast sensitivity units] |
1.825
(0.216)
|
1.591
(0.274)
|
1.700
(0.284)
|
Title | Monocular Mesopic Contrast Sensitivity Without Glare(12 Cycles/Degree) |
---|---|
Description | A single eye (monocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness). The participant's ability to detect changes in contrast is determined. Higher mean indicates improved contrast sensitivity. |
Time Frame | Postoperative visit 3 (day 30-60) |
Outcome Measure Data
Analysis Population Description |
---|
Full Analysis Set, non-missing |
Arm/Group Title | Crystalens AO | ReSTOR 3.0 | AMO Tecnis Multifocal |
---|---|---|---|
Arm/Group Description | Bausch & Lomb silicone multi-piece accommodating IOL is a modified plate haptic lens | An aspheric multifocal IOL (Alcon Laboratories) combines the functions of an apodized diffractive region and a refractive region. | A foldable hydrophobic acrylic IOL,(Abbott Medical Optics), is an ultraviolet light-absorbing posterior chamber IOL |
Measure Participants | 24 | 22 | 21 |
Measure eyes | 48 | 43 | 42 |
Mean (Standard Deviation) [log contrast sensitivity units] |
1.325
(0.226)
|
1.227
(0.198)
|
1.324
(0.281)
|
Title | Monocular Mesopic Contrast Sensitivity Without Glare(18 Cycles/Degree) |
---|---|
Description | A single eye (monocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness). The participant's ability to detect changes in contrast is determined. Higher mean indicates improved contrast sensitivity. |
Time Frame | Postoperative visit 3 (day 30-60) |
Outcome Measure Data
Analysis Population Description |
---|
Full Analysis Set, non-missing |
Arm/Group Title | Crystalens AO | ReSTOR 3.0 | AMO Tecnis Multifocal |
---|---|---|---|
Arm/Group Description | Bausch & Lomb silicone multi-piece accommodating IOL is a modified plate haptic lens | An aspheric multifocal IOL (Alcon Laboratories) combines the functions of an apodized diffractive region and a refractive region. | A foldable hydrophobic acrylic IOL,(Abbott Medical Optics), is an ultraviolet light-absorbing posterior chamber IOL |
Measure Participants | 17 | 13 | 15 |
Measure eyes | 34 | 26 | 30 |
Mean (Standard Deviation) [log contrast sensitivity units] |
0.977
(0.266)
|
0.985
(0.185)
|
0.973
(0.342)
|
Title | Monocular Mesopic Contrast Sensitivity Without Glare(1.5, 3 Cycles/Degree) |
---|---|
Description | A single eye (monocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness). The participant's ability to detect changes in contrast is determined. Higher mean indicates improved contrast sensitivity. |
Time Frame | Postoperative visit 4 (day 120-180) |
Outcome Measure Data
Analysis Population Description |
---|
[Not Specified] |
Arm/Group Title | Crystalens AO | ReSTOR 3.0 | AMO Tecnis Multifocal |
---|---|---|---|
Arm/Group Description | Bausch & Lomb silicone multi-piece accommodating IOL is a modified plate haptic lens | An aspheric multifocal IOL (Alcon Laboratories) combines the functions of an apodized diffractive region and a refractive region. | A foldable hydrophobic acrylic IOL,(Abbott Medical Optics), is an ultraviolet light-absorbing posterior chamber IOL |
Measure Participants | 25 | 27 | 25 |
Measure eyes | 50 | 54 | 50 |
1.5 cycles/degree |
1.696
(0.213)
|
1.621
(0.222)
|
1.527
(0.190)
|
3 cycles/degree |
1.678
(0.159)
|
1.670
(0.236)
|
1.630
(0.203)
|
Title | Monocular Mesopic Contrast Sensitivity Without Glare(6 Cycles/Degree) |
---|---|
Description | A single eye (monocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness). The participant's ability to detect changes in contrast is determined. Higher mean indicates improved contrast sensitivity. |
Time Frame | Postoperative visit 4 (day 120-180) |
Outcome Measure Data
Analysis Population Description |
---|
[Not Specified] |
Arm/Group Title | Crystalens AO | ReSTOR 3.0 | AMO Tecnis Multifocal |
---|---|---|---|
Arm/Group Description | Bausch & Lomb silicone multi-piece accommodating IOL is a modified plate haptic lens | An aspheric multifocal IOL (Alcon Laboratories) combines the functions of an apodized diffractive region and a refractive region. | A foldable hydrophobic acrylic IOL,(Abbott Medical Optics), is an ultraviolet light-absorbing posterior chamber IOL |
Measure Participants | 24 | 26 | 24 |
Measure eyes | 48 | 51 | 48 |
Mean (Standard Deviation) [log contrast sensitivity units] |
1.579
(0.239)
|
1.499
(0.230)
|
1.610
(0.228)
|
Title | Monocular Mesopic Contrast Sensitivity Without Glare(12 Cycles/Degree) |
---|---|
Description | A single eye (monocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness). The participant's ability to detect changes in contrast is determined. Higher mean indicates improved contrast sensitivity. |
Time Frame | Postoperative visit 4 (day 120-180) |
Outcome Measure Data
Analysis Population Description |
---|
[Not Specified] |
Arm/Group Title | Crystalens AO | ReSTOR 3.0 | AMO Tecnis Multifocal |
---|---|---|---|
Arm/Group Description | Bausch & Lomb silicone multi-piece accommodating IOL is a modified plate haptic lens | An aspheric multifocal IOL (Alcon Laboratories) combines the functions of an apodized diffractive region and a refractive region. | A foldable hydrophobic acrylic IOL,(Abbott Medical Optics), is an ultraviolet light-absorbing posterior chamber IOL |
Measure Participants | 20 | 18 | 20 |
Measure eyes | 39 | 35 | 40 |
Mean (Standard Deviation) [log contrast sensitivity units] |
1.280
(0.245)
|
1.225
(0.253)
|
1.230
(0.248)
|
Title | Monocular Mesopic Contrast Sensitivity Without Glare(18 Cycles/Degree) |
---|---|
Description | A single eye (monocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness). The participant's ability to detect changes in contrast is determined. Higher mean indicates improved contrast sensitivity. |
Time Frame | Postoperative visit 4 (day 120-180) |
Outcome Measure Data
Analysis Population Description |
---|
[Not Specified] |
Arm/Group Title | Crystalens AO | ReSTOR 3.0 | AMO Tecnis Multifocal |
---|---|---|---|
Arm/Group Description | Bausch & Lomb silicone multi-piece accommodating IOL is a modified plate haptic lens | An aspheric multifocal IOL (Alcon Laboratories) combines the functions of an apodized diffractive region and a refractive region. | A foldable hydrophobic acrylic IOL,(Abbott Medical Optics), is an ultraviolet light-absorbing posterior chamber IOL |
Measure Participants | 13 | 10 | 9 |
Measure eyes | 25 | 20 | 18 |
Mean (Standard Deviation) [log contrast sensitivity units] |
0.974
(0.303)
|
0.980
(0.246)
|
0.976
(0.207)
|
Title | Binocular Mesopic Contrast Sensitivity Without Glare(12 Cycles/Degree) |
---|---|
Description | In two eyes (binocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness). The participant's ability to detect changes in contrast is determined. Higher mean indicates improved contrast sensitivity. |
Time Frame | Postoperative visit 3 (day 30-60) |
Outcome Measure Data
Analysis Population Description |
---|
Full Analysis Set |
Arm/Group Title | Crystalens AO | ReSTOR 3.0 | AMO Tecnis Multifocal |
---|---|---|---|
Arm/Group Description | Bausch & Lomb silicone multi-piece accommodating IOL is a modified plate haptic lens | An aspheric multifocal IOL (Alcon Laboratories) combines the functions of an apodized diffractive region and a refractive region. | A foldable hydrophobic acrylic IOL,(Abbott Medical Optics), is an ultraviolet light-absorbing posterior chamber IOL |
Measure Participants | 25 | 21 | 22 |
Mean (Standard Deviation) [log contrast sensitivity units] |
1.425
(0.221)
|
1.335
(0.218)
|
1.361
(0.278)
|
Title | Binocular Mesopic Contrast Sensitivity Without Glare(18 Cycles/Degree) |
---|---|
Description | In two eyes (binocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness). The participant's ability to detect changes in contrast is determined. Higher mean indicates improved contrast sensitivity. |
Time Frame | Postoperative visit 3 (day 30-60) |
Outcome Measure Data
Analysis Population Description |
---|
Full Analysis Set |
Arm/Group Title | Crystalens AO | ReSTOR 3.0 | AMO Tecnis Multifocal |
---|---|---|---|
Arm/Group Description | Bausch & Lomb silicone multi-piece accommodating IOL is a modified plate haptic lens | An aspheric multifocal IOL (Alcon Laboratories) combines the functions of an apodized diffractive region and a refractive region. | A foldable hydrophobic acrylic IOL,(Abbott Medical Optics), is an ultraviolet light-absorbing posterior chamber IOL |
Measure Participants | 19 | 15 | 16 |
Mean (Standard Deviation) [log contrast sensitivity units] |
1.029
(0.250)
|
1.001
(0.213)
|
0.981
(0.362)
|
Title | Binocular Mesopic Contrast Sensitivity Without Glare(1.5, 3 Cycles/Degree) |
---|---|
Description | In two eyes (binocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness). The participant's ability to detect changes in contrast is determined. Higher mean indicates improved contrast sensitivity. |
Time Frame | Postoperative visit 4 (day 120-180) |
Outcome Measure Data
Analysis Population Description |
---|
Full Analysis Set |
Arm/Group Title | Crystalens AO | ReSTOR 3.0 | AMO Tecnis Multifocal |
---|---|---|---|
Arm/Group Description | Bausch & Lomb silicone multi-piece accommodating IOL is a modified plate haptic lens | An aspheric multifocal IOL (Alcon Laboratories) combines the functions of an apodized diffractive region and a refractive region. | A foldable hydrophobic acrylic IOL,(Abbott Medical Optics), is an ultraviolet light-absorbing posterior chamber IOL |
Measure Participants | 25 | 27 | 25 |
1.5 cycles/degree |
1.797
(0.183)
|
1.751
(0.238)
|
1.577
(0.206)
|
3 cycles/degree |
1.851
(0.178)
|
1.723
(0.239)
|
1.702
(0.219)
|
Title | Binocular Mesopic Contrast Sensitivity Without Glare(6 Cycles/Degree) |
---|---|
Description | In two eyes (binocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness). The participant's ability to detect changes in contrast is determined. Higher mean indicates improved contrast sensitivity. |
Time Frame | Postoperative visit 4 (day 120-180) |
Outcome Measure Data
Analysis Population Description |
---|
Full Analysis Set |
Arm/Group Title | Crystalens AO | ReSTOR 3.0 | AMO Tecnis Multifocal |
---|---|---|---|
Arm/Group Description | Bausch & Lomb silicone multi-piece accommodating IOL is a modified plate haptic lens | An aspheric multifocal IOL (Alcon Laboratories) combines the functions of an apodized diffractive region and a refractive region. | A foldable hydrophobic acrylic IOL,(Abbott Medical Optics), is an ultraviolet light-absorbing posterior chamber IOL |
Measure Participants | 24 | 25 | 24 |
Mean (Standard Deviation) [log contrast sensitivity units] |
1.758
(0.280)
|
1.640
(0.236)
|
1.690
(0.263)
|
Title | Binocular Mesopic Contrast Sensitivity Without Glare(12 Cycles/Degree) |
---|---|
Description | In two eyes (binocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness). The participant's ability to detect changes in contrast is determined. Higher mean indicates improved contrast sensitivity. |
Time Frame | Postoperative visit 4 (day 120-180) |
Outcome Measure Data
Analysis Population Description |
---|
Full Analysis Set |
Arm/Group Title | Crystalens AO | ReSTOR 3.0 | AMO Tecnis Multifocal |
---|---|---|---|
Arm/Group Description | Bausch & Lomb silicone multi-piece accommodating IOL is a modified plate haptic lens | An aspheric multifocal IOL (Alcon Laboratories) combines the functions of an apodized diffractive region and a refractive region. | A foldable hydrophobic acrylic IOL,(Abbott Medical Optics), is an ultraviolet light-absorbing posterior chamber IOL |
Measure Participants | 21 | 18 | 22 |
Mean (Standard Deviation) [log contrast sensitivity units] |
1.371
(0.247)
|
1.269
(0.257)
|
1.310
(0.286)
|
Title | Binocular Mesopic Contrast Sensitivity Without Glare(18 Cycles/Degree) |
---|---|
Description | In two eyes (binocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness). The participant's ability to detect changes in contrast is determined. Higher mean indicates improved contrast sensitivity. |
Time Frame | Postoperative visit 4 (day 120-180) |
Outcome Measure Data
Analysis Population Description |
---|
Full Analysis Set |
Arm/Group Title | Crystalens AO | ReSTOR 3.0 | AMO Tecnis Multifocal |
---|---|---|---|
Arm/Group Description | Bausch & Lomb silicone multi-piece accommodating IOL is a modified plate haptic lens | An aspheric multifocal IOL (Alcon Laboratories) combines the functions of an apodized diffractive region and a refractive region. | A foldable hydrophobic acrylic IOL,(Abbott Medical Optics), is an ultraviolet light-absorbing posterior chamber IOL |
Measure Participants | 16 | 12 | 12 |
Mean (Standard Deviation) [log contrast sensitivity units] |
1.115
(0.299)
|
0.995
(0.384)
|
1.041
(0.330)
|
Title | Monocular Mesopic Contrast Sensitivity With Glare(1.5 Cycles/Degree) |
---|---|
Description | In one eye (monocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness). The participant's ability to detect changes in contrast is determined. Higher mean indicates improved contrast sensitivity. |
Time Frame | Postoperative visit 3 (day 30-60) |
Outcome Measure Data
Analysis Population Description |
---|
Full analysis set |
Arm/Group Title | Crystalens AO | ReSTOR 3.0 | AMO Tecnis Multifocal |
---|---|---|---|
Arm/Group Description | Bausch & Lomb silicone multi-piece accommodating IOL is a modified plate haptic lens | An aspheric multifocal IOL (Alcon Laboratories) combines the functions of an apodized diffractive region and a refractive region. | A foldable hydrophobic acrylic IOL,(Abbott Medical Optics), is an ultraviolet light-absorbing posterior chamber IOL |
Measure Participants | 26 | 25 | 25 |
Measure eyes | 52 | 50 | 49 |
Mean (Standard Deviation) [log contrast sensitivity units] |
1.591
(0.215)
|
1.519
(0.218)
|
1.482
(0.268)
|
Title | Monocular Mesopic Contrast Sensitivity With Glare (3 Cycles/Degree) |
---|---|
Description | In one eye (monocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness). The participant's ability to detect changes in contrast is determined. Higher mean indicates improved contrast sensitivity. |
Time Frame | Postoperative visit 3 (day 30-60) |
Outcome Measure Data
Analysis Population Description |
---|
Full analysis set |
Arm/Group Title | Crystalens AO | ReSTOR 3.0 | AMO Tecnis Multifocal |
---|---|---|---|
Arm/Group Description | Bausch & Lomb silicone multi-piece accommodating IOL is a modified plate haptic lens | An aspheric multifocal IOL (Alcon Laboratories) combines the functions of an apodized diffractive region and a refractive region. | A foldable hydrophobic acrylic IOL,(Abbott Medical Optics), is an ultraviolet light-absorbing posterior chamber IOL |
Measure Participants | 26 | 25 | 24 |
Measure eyes | 52 | 50 | 47 |
Mean (Standard Deviation) [log contrast sensitivity units] |
1.687
(0.229)
|
1.581
(0.204)
|
1.588
(0.264)
|
Title | Monocular Mesopic Contrast Sensitivity With Glare(6 Cycles/Degree) |
---|---|
Description | In one eye (monocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness). The participant's ability to detect changes in contrast is determined. Higher mean indicates improved contrast sensitivity. |
Time Frame | Postoperative visit 3 (day 30-60) |
Outcome Measure Data
Analysis Population Description |
---|
Full analysis set |
Arm/Group Title | Crystalens AO | ReSTOR 3.0 | AMO Tecnis Multifocal |
---|---|---|---|
Arm/Group Description | Bausch & Lomb silicone multi-piece accommodating IOL is a modified plate haptic lens | An aspheric multifocal IOL (Alcon Laboratories) combines the functions of an apodized diffractive region and a refractive region. | A foldable hydrophobic acrylic IOL,(Abbott Medical Optics), is an ultraviolet light-absorbing posterior chamber IOL |
Measure Participants | 25 | 25 | 23 |
Measure eyes | 50 | 49 | 45 |
Mean (Standard Deviation) [log contrast sensitivity units] |
1.674
(0.237)
|
1.531
(0.238)
|
1.626
(0.293)
|
Title | Monocular Mesopic Contrast Sensitivity With Glare(12 Cycles/Degree) |
---|---|
Description | In one eye (monocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness). The participant's ability to detect changes in contrast is determined. Higher mean indicates improved contrast sensitivity. |
Time Frame | Postoperative visit 3 (day 30-60) |
Outcome Measure Data
Analysis Population Description |
---|
Full analysis set |
Arm/Group Title | Crystalens AO | ReSTOR 3.0 | AMO Tecnis Multifocal |
---|---|---|---|
Arm/Group Description | Bausch & Lomb silicone multi-piece accommodating IOL is a modified plate haptic lens | An aspheric multifocal IOL (Alcon Laboratories) combines the functions of an apodized diffractive region and a refractive region. | A foldable hydrophobic acrylic IOL,(Abbott Medical Optics), is an ultraviolet light-absorbing posterior chamber IOL |
Measure Participants | 23 | 18 | 17 |
Measure eyes | 45 | 35 | 34 |
Mean (Standard Deviation) [log contrast sensitivity units] |
1.328
(0.224)
|
1.230
(0.223)
|
1.340
(0.268)
|
Title | Monocular Mesopic Contrast Sensitivity With Glare(18 Cycles/Degree) |
---|---|
Description | In one eye (monocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness). The participant's ability to detect changes in contrast is determined. Higher mean indicates improved contrast sensitivity. |
Time Frame | Postoperative visit 3 (day 30-60) |
Outcome Measure Data
Analysis Population Description |
---|
Full analysis set |
Arm/Group Title | Crystalens AO | ReSTOR 3.0 | AMO Tecnis Multifocal |
---|---|---|---|
Arm/Group Description | Bausch & Lomb silicone multi-piece accommodating IOL is a modified plate haptic lens | An aspheric multifocal IOL (Alcon Laboratories) combines the functions of an apodized diffractive region and a refractive region. | A foldable hydrophobic acrylic IOL,(Abbott Medical Optics), is an ultraviolet light-absorbing posterior chamber IOL |
Measure Participants | 14 | 12 | 13 |
Measure eyes | 27 | 24 | 25 |
Mean (Standard Deviation) [log contrast sensitivity units] |
1.024
(0.196)
|
0.945
(0.158)
|
0.834
(0.283)
|
Title | Monocular Mesopic Contrast Sensitivity With Glare(1.5 Cycles/Degree) |
---|---|
Description | In one eye (monocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness). The participant's ability to detect changes in contrast is determined. Higher mean indicates improved contrast sensitivity. |
Time Frame | Postoperative visit 4 (day 120-180) |
Outcome Measure Data
Analysis Population Description |
---|
Full analysis set |
Arm/Group Title | Crystalens AO | ReSTOR 3.0 | AMO Tecnis Multifocal |
---|---|---|---|
Arm/Group Description | Bausch & Lomb silicone multi-piece accommodating IOL is a modified plate haptic lens | An aspheric multifocal IOL (Alcon Laboratories) combines the functions of an apodized diffractive region and a refractive region. | A foldable hydrophobic acrylic IOL,(Abbott Medical Optics), is an ultraviolet light-absorbing posterior chamber IOL |
Measure Participants | 25 | 25 | 24 |
Measure eyes | 50 | 50 | 48 |
Mean (Standard Deviation) [log contrast sensitivity units] |
1.550
(0.263)
|
1.488
(0.226)
|
1.449
(0.206)
|
Title | Monocular Mesopic Contrast Sensitivity With Glare(3 Cycles/Degree) |
---|---|
Description | In one eye (monocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness). The participant's ability to detect changes in contrast is determined. Higher mean indicates improved contrast sensitivity. |
Time Frame | Postoperative visit 4 (day 120-180) |
Outcome Measure Data
Analysis Population Description |
---|
Full analysis set |
Arm/Group Title | Crystalens AO | ReSTOR 3.0 | AMO Tecnis Multifocal |
---|---|---|---|
Arm/Group Description | Bausch & Lomb silicone multi-piece accommodating IOL is a modified plate haptic lens | An aspheric multifocal IOL (Alcon Laboratories) combines the functions of an apodized diffractive region and a refractive region. | A foldable hydrophobic acrylic IOL,(Abbott Medical Optics), is an ultraviolet light-absorbing posterior chamber IOL |
Measure Participants | 25 | 25 | 25 |
Measure eyes | 50 | 50 | 49 |
Mean (Standard Deviation) [log contrast sensitivity units] |
1.664
(0.258)
|
1.572
(0.225)
|
1.535
(0.220)
|
Title | Monocular Mesopic Contrast Sensitivity With Glare(6 Cycles/Degree) |
---|---|
Description | In one eye (monocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness). The participant's ability to detect changes in contrast is determined. Higher mean indicates improved contrast sensitivity. |
Time Frame | Postoperative visit 4 (day 120-180) |
Outcome Measure Data
Analysis Population Description |
---|
Full analysis set |
Arm/Group Title | Crystalens AO | ReSTOR 3.0 | AMO Tecnis Multifocal |
---|---|---|---|
Arm/Group Description | Bausch & Lomb silicone multi-piece accommodating IOL is a modified plate haptic lens | An aspheric multifocal IOL (Alcon Laboratories) combines the functions of an apodized diffractive region and a refractive region. | A foldable hydrophobic acrylic IOL,(Abbott Medical Optics), is an ultraviolet light-absorbing posterior chamber IOL |
Measure Participants | 24 | 23 | 23 |
Measure eyes | 48 | 46 | 45 |
Mean (Standard Deviation) [log contrast sensitivity units] |
1.578
(0.252)
|
1.493
(0.267)
|
1.519
(0.260)
|
Title | Monocular Mesopic Contrast Sensitivity With Glare(12 Cycles/Degree) |
---|---|
Description | In one eye (monocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness). The participant's ability to detect changes in contrast is determined. Higher mean indicates improved contrast sensitivity. |
Time Frame | Postoperative visit 4 (day 120-180) |
Outcome Measure Data
Analysis Population Description |
---|
Full analysis set |
Arm/Group Title | Crystalens AO | ReSTOR 3.0 | AMO Tecnis Multifocal |
---|---|---|---|
Arm/Group Description | Bausch & Lomb silicone multi-piece accommodating IOL is a modified plate haptic lens | An aspheric multifocal IOL (Alcon Laboratories) combines the functions of an apodized diffractive region and a refractive region. | A foldable hydrophobic acrylic IOL,(Abbott Medical Optics), is an ultraviolet light-absorbing posterior chamber IOL |
Measure Participants | 20 | 15 | 16 |
Measure eyes | 39 | 30 | 31 |
Mean (Standard Deviation) [log contrast sensitivity units] |
1.335
(0.298)
|
1.246
(0.278)
|
1.194
(0.237)
|
Title | Monocular Mesopic Contrast Sensitivity With Glare(18 Cycles/Degree) |
---|---|
Description | In one eye (monocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness). The participant's ability to detect changes in contrast is determined. Higher mean indicates improved contrast sensitivity. |
Time Frame | Postoperative visit 4 (day 120-180) |
Outcome Measure Data
Analysis Population Description |
---|
Full analysis set |
Arm/Group Title | Crystalens AO | ReSTOR 3.0 | AMO Tecnis Multifocal |
---|---|---|---|
Arm/Group Description | Bausch & Lomb silicone multi-piece accommodating IOL is a modified plate haptic lens | An aspheric multifocal IOL (Alcon Laboratories) combines the functions of an apodized diffractive region and a refractive region. | A foldable hydrophobic acrylic IOL,(Abbott Medical Optics), is an ultraviolet light-absorbing posterior chamber IOL |
Measure Participants | 13 | 8 | 8 |
Measure eyes | 25 | 16 | 16 |
Mean (Standard Deviation) [log contrast sensitivity units] |
0.919
(0.283)
|
1.019
(0.367)
|
0.935
(0.207)
|
Adverse Events
Time Frame | 120-180 days | |||||
---|---|---|---|---|---|---|
Adverse Event Reporting Description | Adverse events measured by eye | |||||
Arm/Group Title | Crystalens AO | ReSTOR 3.0 | AMO Tecnis Multifocal | |||
Arm/Group Description | Bausch & Lomb silicone multi-piece accommodating IOL is a modified plate haptic lens | An aspheric multifocal IOL (Alcon Laboratories) combines the functions of an apodized diffractive region and a refractive region. | A foldable hydrophobic acrylic IOL,(Abbott Medical Optics), is an ultraviolet light-absorbing posterior chamber IOL | |||
All Cause Mortality |
||||||
Crystalens AO | ReSTOR 3.0 | AMO Tecnis Multifocal | ||||
Affected / at Risk (%) | # Events | Affected / at Risk (%) | # Events | Affected / at Risk (%) | # Events | |
Total | / (NaN) | / (NaN) | / (NaN) | |||
Serious Adverse Events |
||||||
Crystalens AO | ReSTOR 3.0 | AMO Tecnis Multifocal | ||||
Affected / at Risk (%) | # Events | Affected / at Risk (%) | # Events | Affected / at Risk (%) | # Events | |
Total | 0/53 (0%) | 0/54 (0%) | 0/51 (0%) | |||
Other (Not Including Serious) Adverse Events |
||||||
Crystalens AO | ReSTOR 3.0 | AMO Tecnis Multifocal | ||||
Affected / at Risk (%) | # Events | Affected / at Risk (%) | # Events | Affected / at Risk (%) | # Events | |
Total | 12/53 (22.6%) | 5/54 (9.3%) | 6/51 (11.8%) | |||
Eye disorders | ||||||
Increased Intraocular Pressure | 11/53 (20.8%) | 11 | 5/54 (9.3%) | 5 | 3/51 (5.9%) | 3 |
Cystoid Macular Edema | 1/53 (1.9%) | 1 | 0/54 (0%) | 0 | 3/51 (5.9%) | 3 |
Limitations/Caveats
More Information
Certain Agreements
Principal Investigators are NOT employed by the organization sponsoring the study.
All study data generated as a result of this study will be regarded as confidential, until appropriate analysis and review by the Sponsor or its designee and the Investigator(s) are completed. The results of the study may be published or presented by the Investigator(s)after the review by, and in consultation and agreement with the Sponsor.