Noninvasive Ventilation Masks Carbon Dioxide Clearance in Normal Volunteers

Sponsor
Rush University Medical Center (Other)
Overall Status
Completed
CT.gov ID
NCT03882723
Collaborator
(none)
8
1
4
17
0.5

Study Details

Study Description

Brief Summary

This study will be a randomized crossover trial that will be conducted at Rush University. The study will be conducted with 20 healthy volunteers who will be placed on ICU ventilator operating in the NIPPV mode. All subjects will perform 15 minutes of breathing on NIPPV on each mask of the 4 different masks (2 oronasal and 2 full face masks) that are randomly selected. Breathing through each mask will be followed by a 5- minute wash out period between masks. End tidal Carbon Dioxide (EtCO2) will be sampled nasal/oral. Different levels of EPAP will be set and CO2 clearance will be monitored. Additionally, subjective mask comfort will be assessed via visual analog scale (VAS) with 1 referring to least comfortable and 5 being the most comfortable.

Condition or Disease Intervention/Treatment Phase
  • Device: Evaluated masks used for NIV with mechanical ventilators
N/A

Detailed Description

This study will be a randomized crossover trial that will be conducted at Rush University.

Normal volunteers will be recruited from Rush University Medical Centers. Volunteers will be screened with inclusion and exclusion criteria and an informed consent is required in order for them to participate. Demographic (age, gender, ethnicity) will be obtained.

Also, initial baseline vital signs (heart rate, blood pressure, respiratory rate, and oxygen saturation) will be obtained. EtCO2 will be obtained at baseline and periodically using an oral/nasal sample line with the CapnostreamTM 20p monitor (Medtronic, Minneapolis, MN). The study will use a double limb circuit ICU ventilator Puritan Bennett™ (Medtronic, Dublin, Ireland). Investigators will evaluate 4 NIPPV masks (2 oronasal masks and 2 full face masks).

Oronasal masks will be labeled as group A and numbered as A1 and A2; while full face masks will be labeled group B and numbered as B1 and B2. Order of the masks will be randomly chosen in two steps. First steps will be choosing A or B then randomly assign mask in each of these groups. Investigators will randomly assign mask in group A or B by drawing an initial mask from the group and then continue the sequence of masks from that point. For example, investigators randomly chose group A, then randomly chose A2, the sequence will be A2, A1. Then Investigators would randomly choose B2, then the sequence would be B2, B1. All subjects will perform 15 minutes on each mask followed by 5 minutes wash out interval between masks. EPAP levels will be 0, 2, 4, & 5 while IPAP remains at 5 higher than EPAP. EtCO2 will be collected at 4:00, 4:30 and 5:00 minute mark. These three measures should differ by less than 20% of their average. Also, subjective mask ORA: 18011502-IRB01 Date IRB Approved: 3/30/2018 Amendment Date: 5/11/2018 comfort will be assessed via visual analog scale (VAS) with 1 referring to least comfortable and 5 being the most comfortable after 5 each EPAP setting.

Study Design

Study Type:
Interventional
Actual Enrollment :
8 participants
Allocation:
Randomized
Intervention Model:
Crossover Assignment
Intervention Model Description:
randomized crossover trialrandomized crossover trial
Masking:
None (Open Label)
Primary Purpose:
Prevention
Official Title:
Carbon Dioxide (CO2) Clearance From Various Face Masks in Normal Volunteers at Various Level of Expiratory Positive Airway Pressure (EPAP) When Using a Non-Invasive Positive Pressure Ventilation (NIPPV)
Actual Study Start Date :
Mar 30, 2018
Actual Primary Completion Date :
Aug 30, 2019
Actual Study Completion Date :
Aug 30, 2019

Arms and Interventions

Arm Intervention/Treatment
Active Comparator: BiTrac MaxShield™ with standard elbow (FFM)

Order of the masks were randomly chosen. Investigators randomly assigned all masks for paper raffling from container. All subjects performed 20 minutes on each mask followed by 5 minutes wash out interval between masks. EPAP levels will be 0, 2, 4, & 5 (5 minutes for each level) while IPAP remains at 5 higher than EPAP. FiCO2 and EtCO2 were collected at 4:00, 4:30 and 5:00 minute mark for each EPAP level. Also, subjective mask comfort was assessed via visual analog scale (VAS) with 1 referring to least comfortable and 5 being the most comfortable after 5 each EPAP setting.

Device: Evaluated masks used for NIV with mechanical ventilators
Order of the masks were randomly chosen. Investigators randomly assigned all masks for paper raffling from container. All subjects performed 20 minutes on each mask followed by 5 minutes wash out interval between masks. EPAP levels will be 0, 2, 4, & 5 (5 minutes for each level) while IPAP remains at 5 higher than EPAP. FiCO2 and EtCO2 were collected at 4:00, 4:30 and 5:00 minute mark for each EPAP level. Also, subjective mask comfort was assessed via visual analog scale (VAS) with 1 referring to least comfortable and 5 being the most comfortable after 5 each EPAP setting.

Active Comparator: BiTrac™ Full Face with standard elbow

Order of the masks were randomly chosen. Investigators randomly assigned all masks for paper raffling from container. All subjects performed 20 minutes on each mask followed by 5 minutes wash out interval between masks. EPAP levels will be 0, 2, 4, & 5 (5 minutes for each level) while IPAP remains at 5 higher than EPAP. FiCO2 and EtCO2 were collected at 4:00, 4:30 and 5:00 minute mark for each EPAP level. Also, subjective mask comfort was assessed via visual analog scale (VAS) with 1 referring to least comfortable and 5 being the most comfortable after 5 each EPAP setting.

Device: Evaluated masks used for NIV with mechanical ventilators
Order of the masks were randomly chosen. Investigators randomly assigned all masks for paper raffling from container. All subjects performed 20 minutes on each mask followed by 5 minutes wash out interval between masks. EPAP levels will be 0, 2, 4, & 5 (5 minutes for each level) while IPAP remains at 5 higher than EPAP. FiCO2 and EtCO2 were collected at 4:00, 4:30 and 5:00 minute mark for each EPAP level. Also, subjective mask comfort was assessed via visual analog scale (VAS) with 1 referring to least comfortable and 5 being the most comfortable after 5 each EPAP setting.

Active Comparator: Respironics PerforMax with standard elbow

Order of the masks were randomly chosen. Investigators randomly assigned all masks for paper raffling from container. All subjects performed 20 minutes on each mask followed by 5 minutes wash out interval between masks. EPAP levels will be 0, 2, 4, & 5 (5 minutes for each level) while IPAP remains at 5 higher than EPAP. FiCO2 and EtCO2 were collected at 4:00, 4:30 and 5:00 minute mark for each EPAP level. Also, subjective mask comfort was assessed via visual analog scale (VAS) with 1 referring to least comfortable and 5 being the most comfortable after 5 each EPAP setting.

Device: Evaluated masks used for NIV with mechanical ventilators
Order of the masks were randomly chosen. Investigators randomly assigned all masks for paper raffling from container. All subjects performed 20 minutes on each mask followed by 5 minutes wash out interval between masks. EPAP levels will be 0, 2, 4, & 5 (5 minutes for each level) while IPAP remains at 5 higher than EPAP. FiCO2 and EtCO2 were collected at 4:00, 4:30 and 5:00 minute mark for each EPAP level. Also, subjective mask comfort was assessed via visual analog scale (VAS) with 1 referring to least comfortable and 5 being the most comfortable after 5 each EPAP setting.

Active Comparator: Philips Respironics AF531 with standard elbow

Order of the masks were randomly chosen. Investigators randomly assigned all masks for paper raffling from container. All subjects performed 20 minutes on each mask followed by 5 minutes wash out interval between masks. EPAP levels will be 0, 2, 4, & 5 (5 minutes for each level) while IPAP remains at 5 higher than EPAP. FiCO2 and EtCO2 were collected at 4:00, 4:30 and 5:00 minute mark for each EPAP level. Also, subjective mask comfort was assessed via visual analog scale (VAS) with 1 referring to least comfortable and 5 being the most comfortable after 5 each EPAP setting.

Device: Evaluated masks used for NIV with mechanical ventilators
Order of the masks were randomly chosen. Investigators randomly assigned all masks for paper raffling from container. All subjects performed 20 minutes on each mask followed by 5 minutes wash out interval between masks. EPAP levels will be 0, 2, 4, & 5 (5 minutes for each level) while IPAP remains at 5 higher than EPAP. FiCO2 and EtCO2 were collected at 4:00, 4:30 and 5:00 minute mark for each EPAP level. Also, subjective mask comfort was assessed via visual analog scale (VAS) with 1 referring to least comfortable and 5 being the most comfortable after 5 each EPAP setting.

Outcome Measures

Primary Outcome Measures

  1. Carbon dioxide clearance from the Masks [5 min]

    Measured the amount or fraction of CO2 being re-breathed

Eligibility Criteria

Criteria

Ages Eligible for Study:
18 Years and Older
Sexes Eligible for Study:
All
Accepts Healthy Volunteers:
Yes
Inclusion Criteria:
  • Healthy subjects greater than 18 years old.
Exclusion Criteria:
  • Prior history of NIPPV as a patient.

  • Facial surgery or deformity

  • Ear infection

  • History of pulmonary or cardiac disease

Contacts and Locations

Locations

Site City State Country Postal Code
1 Rush University Medical Center Chicago Illinois United States 60612

Sponsors and Collaborators

  • Rush University Medical Center

Investigators

  • Study Chair: David Vines, MHS, Rush University Medical Center

Study Documents (Full-Text)

None provided.

More Information

Publications

None provided.
Responsible Party:
Rush University Medical Center
ClinicalTrials.gov Identifier:
NCT03882723
Other Study ID Numbers:
  • ORA: 18011502-IRB01
First Posted:
Mar 20, 2019
Last Update Posted:
Nov 10, 2020
Last Verified:
Nov 1, 2020
Individual Participant Data (IPD) Sharing Statement:
Undecided
Plan to Share IPD:
Undecided
Studies a U.S. FDA-regulated Drug Product:
No
Studies a U.S. FDA-regulated Device Product:
Yes
Product Manufactured in and Exported from the U.S.:
No

Study Results

No Results Posted as of Nov 10, 2020