Evaluation of the Motus Gl CleanUp System During Screening Colonoscopy

Sponsor
Motus GI Medical Technologies Ltd (Industry)
Overall Status
Completed
CT.gov ID
NCT02161549
Collaborator
(none)
71
1
15
4.7

Study Details

Study Description

Brief Summary

Motus Gl is developing a single-use device that attaches to a standard colonoscope, thereby integrating colon cleansing, screening and treatment into one procedure.

The device creates turbulence within the colon using water jets. The colon's fecal fluids are evacuated by the system.

Condition or Disease Intervention/Treatment Phase
  • Device: Motus Gl CleanUp System

Detailed Description

63 eligible subjects who are allocated for a diagnostic screening colonoscopy were enrolled in the study. The assignment of the level of the bowel preparation agent was according to the physician decision. Following preparation, enrolled subjects were intubated and cleansed using a standard colonoscope with the Motus GI CleanUp device attached to it. At the end of the procedure, subjects were followed through discharge and at 48 hours post procedure for occurrence of adverse events.

40 subjects were enrolled under protocol Rev 1.0 used Motus GI CleanUp System Rev 1.0 31 subjects were enrolled under protocol Rev 2 used Motus GI CleanUp System Rev 1.5, due to improvement made to the study device including the following:

  1. One head that irrigates and evacuates instead of two heads

  2. Elimination of the grinding mechanism and the motor to reduce the pipes' diameters while keeping the same flow rate.

  3. System adjustment to the above changes including updated software version.

Study Design

Study Type:
Observational [Patient Registry]
Actual Enrollment :
71 participants
Observational Model:
Cohort
Time Perspective:
Prospective
Official Title:
Evaluation of the Motus Gl CleanUp System During Screening Colonoscopy
Actual Study Start Date :
Jul 1, 2012
Actual Primary Completion Date :
Oct 1, 2013
Actual Study Completion Date :
Oct 1, 2013

Arms and Interventions

Arm Intervention/Treatment
Colonoscopy with MotusGi CleanUp System Rev 1.0

subjects indicated for colonoscopy procedure with CleanUp System Rev 1.0 , enrolled under protocol Rev 1.0

Device: Motus Gl CleanUp System

Colonoscopy with MotusGi CleanUp System Rev 1.5

subjects indicated for colonoscopy procedure with CleanUp System Rev 1.5 , enrolled under protocol Rev 2.0

Device: Motus Gl CleanUp System

Outcome Measures

Primary Outcome Measures

  1. Mean Score of Boston Bowel Preparation Scale (BBPS) Index Prior and After the Use of MotusGI CleanUp System [during colonoscopy procedure after cleansing with CleanUp System]

    Human colon has 3 segments and each segment can be scored 0 (unprepared colon) - 3 (clean colon). average of all colon segments BBPS score after the use of MotusGI CleanUp System

Eligibility Criteria

Criteria

Ages Eligible for Study:
50 Years to 69 Years
Sexes Eligible for Study:
All
Accepts Healthy Volunteers:
Yes
Inclusion Criteria:
  • Subjects in the age range of 50-69 years

  • Subjects who are considered for routine screening colonoscopy

  • Subjects with BMI within the range of 18.5-30

  • Subject is willing to sign informed consent form

Exclusion Criteria:
  • History of G diseases or history of Colorectal cancer (personal or family)

  • Previous major abdominal surgeries

  • Medical status ASA grater or equal to 3

  • Acute Gl symptoms, such as bleeding, unexplained weight loss, inflammatory bowel disease (lBD).

  • Patients taking anticoagulants (e.g., Coumadin, Heparin, Clopidrogrel)

  • Life-threatening condition

  • Subjects who are unable or unwilling to cooperate with study procedures

  • Severe Diverticula

Contacts and Locations

Locations

Site City State Country Postal Code
1 Clinica San Rafael Cadiz Spain 11002

Sponsors and Collaborators

  • Motus GI Medical Technologies Ltd

Investigators

None specified.

Study Documents (Full-Text)

None provided.

More Information

Publications

None provided.
Responsible Party:
Motus GI Medical Technologies Ltd
ClinicalTrials.gov Identifier:
NCT02161549
Other Study ID Numbers:
  • CP-MCU-SP-1211
First Posted:
Jun 11, 2014
Last Update Posted:
Sep 20, 2017
Last Verified:
Mar 1, 2017
Individual Participant Data (IPD) Sharing Statement:
No
Plan to Share IPD:
No
Studies a U.S. FDA-regulated Drug Product:
No
Studies a U.S. FDA-regulated Device Product:
No

Study Results

Participant Flow

Recruitment Details The recruitment was done in two phases: Phase 1- under protocol Rev 1.0 used MotusGi CleanUp System Rev 1.0 , study period Aug- Jan 2012. 40 subjects enrolled. Phase 2- under protocol Rev 2.0 used MotusGi CleanUp System Rev 1.5 , study period Jul-Oct 2013.31 subjects enrolled.
Pre-assignment Detail
Arm/Group Title Colonoscopy With MotusGi CleanUp System Rev 1.0 Colonoscopy With MotusGi CleanUp System Rev 1.5
Arm/Group Description subjects indicated for colonoscopy procedure with CleanUp System Rev 1.0 , enrolled under protocol Rev 1.0 Motus Gl CleanUp System subjects indicated for colonoscopy procedure with CleanUp System Rev 1.5 , enrolled under protocol Rev 2.0 Motus Gl CleanUp System
Period Title: Overall Study
STARTED 40 31
COMPLETED 34 29
NOT COMPLETED 6 2

Baseline Characteristics

Arm/Group Title Colonoscopy With MotusGi CleanUp System Rev 1.0 Colonoscopy With MotusGi CleanUp System Rev 1.5 Total
Arm/Group Description subjects indicated for colonoscopy procedure with CleanUp System Rev 1.0 , enrolled under protocol Rev 1.0 Motus Gl CleanUp System subjects indicated for colonoscopy procedure with CleanUp System Rev 1.5 , enrolled under protocol Rev 2.0 Motus Gl CleanUp System Total of all reporting groups
Overall Participants 34 29 63
Age (years) [Mean (Standard Deviation) ]
Mean (Standard Deviation) [years]
55
(3.8)
53
(4.1)
54
(4.0)
Sex: Female, Male (Count of Participants)
Female
13
38.2%
14
48.3%
27
42.9%
Male
21
61.8%
15
51.7%
36
57.1%

Outcome Measures

1. Primary Outcome
Title Mean Score of Boston Bowel Preparation Scale (BBPS) Index Prior and After the Use of MotusGI CleanUp System
Description Human colon has 3 segments and each segment can be scored 0 (unprepared colon) - 3 (clean colon). average of all colon segments BBPS score after the use of MotusGI CleanUp System
Time Frame during colonoscopy procedure after cleansing with CleanUp System

Outcome Measure Data

Analysis Population Description
the mean and standard deviation of the average score after cleansing the colon using the Motus CleanUp System
Arm/Group Title Colonoscopy With MotusGi CleanUp System Rev 1.0 Colonoscopy With MotusGi CleanUp System Rev 1.5
Arm/Group Description subjects indicated for colonoscopy procedure with CleanUp System Rev 1.0 , enrolled under protocol Rev 1.0 Motus Gl CleanUp System subjects indicated for colonoscopy procedure with CleanUp System Rev 1.5 , enrolled under protocol Rev 2.0 Motus Gl CleanUp System
Measure Participants 34 29
Mean (Standard Deviation) [units on a scale]
2.9
(0.4)
2.9
(0.2)

Adverse Events

Time Frame
Adverse Event Reporting Description
Arm/Group Title Colonoscopy With MotusGi CleanUp System Rev 1.0 Colonoscopy With MotusGi CleanUp System Rev 1.5
Arm/Group Description subjects indicated for colonoscopy procedure with CleanUp System Rev 1.0 , enrolled under protocol Rev 1.0 Motus Gl CleanUp System subjects indicated for colonoscopy procedure with CleanUp System Rev 1.5 , enrolled under protocol Rev 2.0 Motus Gl CleanUp System
All Cause Mortality
Colonoscopy With MotusGi CleanUp System Rev 1.0 Colonoscopy With MotusGi CleanUp System Rev 1.5
Affected / at Risk (%) # Events Affected / at Risk (%) # Events
Total 0/40 (0%) 0/31 (0%)
Serious Adverse Events
Colonoscopy With MotusGi CleanUp System Rev 1.0 Colonoscopy With MotusGi CleanUp System Rev 1.5
Affected / at Risk (%) # Events Affected / at Risk (%) # Events
Total 0/40 (0%) 0/31 (0%)
Other (Not Including Serious) Adverse Events
Colonoscopy With MotusGi CleanUp System Rev 1.0 Colonoscopy With MotusGi CleanUp System Rev 1.5
Affected / at Risk (%) # Events Affected / at Risk (%) # Events
Total 17/40 (42.5%) 17/31 (54.8%)
Gastrointestinal disorders
Erosion 5/40 (12.5%) 7 13/31 (41.9%) 13
Hematoma 5/40 (12.5%) 11 7/31 (22.6%) 8
Vesicles 3/40 (7.5%) 3 0/31 (0%) 0
Laceration 0/40 (0%) 0 6/31 (19.4%) 6
Papules 4/40 (10%) 5 0/31 (0%) 0

Limitations/Caveats

[Not Specified]

More Information

Certain Agreements

Principal Investigators are NOT employed by the organization sponsoring the study.

There is NOT an agreement between Principal Investigators and the Sponsor (or its agents) that restricts the PI's rights to discuss or publish trial results after the trial is completed.

Results Point of Contact

Name/Title Ravit Peled -Director of Clinical Trial
Organization MotusGI
Phone +972 (0)4 6214447 ext 203
Email ravit@motusgi.com
Responsible Party:
Motus GI Medical Technologies Ltd
ClinicalTrials.gov Identifier:
NCT02161549
Other Study ID Numbers:
  • CP-MCU-SP-1211
First Posted:
Jun 11, 2014
Last Update Posted:
Sep 20, 2017
Last Verified:
Mar 1, 2017