TIMAGING: Evaluation of Electrode-modiolus Distance and Cochlear Fibrosis Using Depth Sounding and Spectroscopy Tools

Sponsor
University Hospital, Montpellier (Other)
Overall Status
Not yet recruiting
CT.gov ID
NCT05402969
Collaborator
(none)
180
24

Study Details

Study Description

Brief Summary

Cochlear fibrosis development can compromise the success and the outcomes of the cochlear implantation (CI) thus affecting the quality of life of the implanted patient. Correlating the results of the Transimpedance Matrix (TIM) measurements to the implant electrode location determined by the Cone Beam Computer Tomography (CBCT), this study aims to identify a range of TIM profiles within the implanted population, certain profiles suggesting the growth of the fibrosis tissue in cochlea

Condition or Disease Intervention/Treatment Phase
  • Procedure: Transimedance Matrix measurements
  • Procedure: Depth sounding and spectroscopy
  • Procedure: Cone Beam Computer Tomography

Detailed Description

Fibrosis tissue growth after the CI is an inflammatory reaction to a foreign body occurring around the electrode array. It is considered responsible for a larger current diffusion, less precise stimulation of auditory nerve fibres and elevated electrical impedances thus affecting the functioning and the battery life of an implant. Fibrosis has been also proposed as a factor reducing residual hearing. Hence, its early detection and prevention can significantly improve the outcomes of CI.

To date, no reliable technique to visualise cochlear fibrosis in vivo exists. However, the TIM measurement developed by Cochlear® including depth sounding and spectroscopy tools could possibly provide a better "view" of the fibrosis apparition after CI. This approach has been already successfully applied in the University Hospital of Montpellier, France, to the patients-users of the Advanced Bionics cochlear implants.

One of the factors provoking the fibrosis tissue development is an electrode positioning into the scala vestibule (SV) versus scala tympani (ST). CBCT will be used to visualise the electrode location as well as possible damages of the inner ear caused by CI. The imaging data will be then linked to the TIM measurements results.

Potential participants will be recruited from the French population of the Cochlear® implant users either coming for their regular checks-up at the ENT service of the University Hospital of Montpellier and implanted for more than 6 month (time necessary for the fibrosis tissue to develop) or recently implanted at the same hospital.

For the patients with long-term implantation (160 participants planned) the TIM measurements will be performed only once whereas the newly implanted group (20 participants planned) will require these measurements to be repeated several times: on the day of the surgery as well as 15 days, 1 month, 2 months, 3 months and 6 months after it. For the first group the post operational CBCT imaging file will be extracted from the patient medical record; for the second group CBCT will be performed during the surgery.

The main objective of this study is:

For the group with long-term implantation - to show the difference in TIM profiles according to the electrode scala position (ST or SV) determined by CBCT 6 months or more after the implantation For the group with recent implantation - to evaluate the changes in TIM profiles with time according to the electrode scala position (ST or SV) determined by CBCT

The secondary objectives are common for both groups:
  • To analyse various patient subgroups defined by the electrode type (Contour Advanced - only for long-term implantation, Slim Straight, Slim Modiolar) and the inner ear access (round window, enlarged round window or cochleostomy)

  • To identify the TIM profile threshold values, potential indicators of the cochlear fibrosis apparition, depending on the electrode type

  • To evaluate the depth sounding tool accuracy for the measurement of the distance between the electrode and the spiral ganglion in comparison to CBCT depending on the type of the electrode array.

  • To evaluate the spectroscopy tool accuracy for fibrosis detection in specific cochlear areas (around electrodes dislocated to SV or basal electrodes in case of cohleostomy)

Study Design

Study Type:
Observational
Anticipated Enrollment :
180 participants
Observational Model:
Cohort
Time Perspective:
Other
Official Title:
Evaluation of Electrode-modiolus Distance and Cochlear Fibrosis Using Depth Sounding and Spectroscopy Tools: Comparison With Postoperative Cone Beam CT
Anticipated Study Start Date :
Jun 1, 2022
Anticipated Primary Completion Date :
Dec 1, 2023
Anticipated Study Completion Date :
Jun 1, 2024

Arms and Interventions

Arm Intervention/Treatment
Patients with long-term implantation

Patients undergone CI more than 6 months ago. After the study is presented to the patient and the egligibility criterias are verified, the patient is proposed to sign a consent form (visit 0). Once the consent form is signed, the visit 1 is scheduled (0 to 90 days after the visit 0). During the visit 1 TIM measurements are performed, including depth sounding and spectroscopy, and adversed affects if any are collected.

Procedure: Transimedance Matrix measurements
TIM measurements are based on cochlear implant back-telemetry function. Each electrode contact is stimulated in monopolar mode. Transimpedance is represented by the ratio between the voltage measured at each recording electrode and the current injected at the stimulation contact. The result of the measurements is a matrix of 21x21 transimpedances. Stimulation current cannot be perceived by a patient thus making TIM measurements a non-invasive and painless procedure. The test duration is only 2-3 minutes
Other Names:
  • TIM measurements
  • Procedure: Depth sounding and spectroscopy
    Depth sounding and spectroscopy are based on the matrixes obtained after TIM imaging. Deep sounding measures the capacitances throughout the electrode array whereas spectroscopy determines the phase of every transimpedance value. These tests take 10 to 15 minutes together. The stimulation intensity used by depth sounding technique can be perceived by a patient. If it is considered as unpleasant, the test will be stopped.

    Procedure: Cone Beam Computer Tomography
    CBCT is a routine postoperative assessment after CI, non-invasive and painless for patients.
    Other Names:
  • CBCT
  • Recently implanted patients

    Patients enrolled at their pre-op appointments (visit 0 - egligibilty criterias are verified and the consent collected). Visit 1 (0 to 90 days after visit 0) is a surgery. TIM measurements, icluding deep sound and spectroscopy, are performed after the CI and all adverse events are noted. CBCT is routinely done the next day after CI. TIM measurements (as well as adverse event collection) are repeated during visits 2, 3, 4, 5 and 6 (appoitments 15 days, 1, 2, 3 and 6 months after the surgery).

    Procedure: Transimedance Matrix measurements
    TIM measurements are based on cochlear implant back-telemetry function. Each electrode contact is stimulated in monopolar mode. Transimpedance is represented by the ratio between the voltage measured at each recording electrode and the current injected at the stimulation contact. The result of the measurements is a matrix of 21x21 transimpedances. Stimulation current cannot be perceived by a patient thus making TIM measurements a non-invasive and painless procedure. The test duration is only 2-3 minutes
    Other Names:
  • TIM measurements
  • Procedure: Depth sounding and spectroscopy
    Depth sounding and spectroscopy are based on the matrixes obtained after TIM imaging. Deep sounding measures the capacitances throughout the electrode array whereas spectroscopy determines the phase of every transimpedance value. These tests take 10 to 15 minutes together. The stimulation intensity used by depth sounding technique can be perceived by a patient. If it is considered as unpleasant, the test will be stopped.

    Procedure: Cone Beam Computer Tomography
    CBCT is a routine postoperative assessment after CI, non-invasive and painless for patients.
    Other Names:
  • CBCT
  • Outcome Measures

    Primary Outcome Measures

    1. TIM profiles in patients with long-term implantation [Visit 1 (0 to 90 days after inclusion)]

      Comparison of TIM profiles of patients with ST electrode position versus patients with SV electrode dislocateon in the long-term implantation group

    2. TIM profiles time dynamics in the recently implanted patients [Visit 1 - surgery (0 to 90 days after inclusion), visit 3, 4, 5, 6 (1, 2, 3 and 6 months after surgery)]

      Assessment of the TIM profile changes with time and depending on electrode scala position in newly-implanted patients

    Secondary Outcome Measures

    1. TIM profiles depending on electrode array and inner ear access [Visit 1 (0 to 90 days after inclusion)]

      Comparison of the TIM profiles depending on the electrode array type and inner ear access

    2. Depth sounding versus the distance electrode - spiral ganglion [Visit 1 (0 to 90 days after inclusion)]

      Correlation of the depth sounding measures with the electrode-spiral ganglion distance assesed through CBCT imaging

    3. Spectroscopy of the specific electrode insertion zones [Visit 1 (0 to 90 days after inclusion)]

      Comparison of the spectroscopy measures in the specific zones prone to electrode insertion damages (180° turn and basal part) between groups with ST electrode positioning and SV electrode dislocation

    Eligibility Criteria

    Criteria

    Ages Eligible for Study:
    18 Years to 120 Years
    Sexes Eligible for Study:
    All
    Accepts Healthy Volunteers:
    No
    Inclusion Criteria:
    Group with long-term implantation:
    • Age18 years or older

    • Affiliation to the French social security

    • Signed and dated informed consent form

    • Users of uni or bilateral cochlear implants for more than 6 months (CI24CA, CI422, CI512, CI522, CI532, CI612; CI622; CI632), in primo-implantation or reimplantation, with more than 18 electrodes active

    • Postoperation CBCT imaging results available

    Recently-implanted group:
    • Age18 years or older

    • Affiliation to the French social security

    • Signed and dated informed consent form

    • Indication for CI622 or CI632 cochlear implants

    • Postoperation CBCT imaging results available

    Exclusion Criteria:
    • Vulnerable individuals (minors, adults protected in accordance to the French Public Health code, patients deprived of liberty by court decision)

    • Daily usage of the cochlear implant for less than 4 hours or weekly usage for less than 5 days

    • Failure or disfunction of the cochlear implant

    Contacts and Locations

    Locations

    No locations specified.

    Sponsors and Collaborators

    • University Hospital, Montpellier

    Investigators

    None specified.

    Study Documents (Full-Text)

    None provided.

    More Information

    Publications

    None provided.
    Responsible Party:
    University Hospital, Montpellier
    ClinicalTrials.gov Identifier:
    NCT05402969
    Other Study ID Numbers:
    • RECHMPL20_0666
    First Posted:
    Jun 2, 2022
    Last Update Posted:
    Jun 2, 2022
    Last Verified:
    May 1, 2022
    Studies a U.S. FDA-regulated Drug Product:
    No
    Studies a U.S. FDA-regulated Device Product:
    No