Titan® One Touch Release Inflatable Penile Prosthesis
Study Details
Study Description
Brief Summary
The purpose of this study is to assess the ease of deflation of the Titan® OTR pump.
Condition or Disease | Intervention/Treatment | Phase |
---|---|---|
|
N/A |
Study Design
Arms and Interventions
Arm | Intervention/Treatment |
---|---|
Other: Titan® OTR IPP Subjects implanted with Titan® One Touch Release (OTR) Inflatable Penile Prosthesis (IPP) |
Device: Inflatable Penile Prosthesis
Hydraulic system designed to be surgically implanted into the penis for the management of erectile dysfunction.
Other Names:
|
Outcome Measures
Primary Outcome Measures
- Assess the Ease of Deflation of the Titan® OTR Pump [6 months]
The study's primary endpoint was to demonstrate that at least 64% of subjects were mostly or completely satisfied with the ability to deflate the device at the 6-month follow-up. The study's primary objective was to assess the ease of deflation of the Titan® OTR pump via subject questionnaire at 6-month follow-up. Subjects were asked via questionnaire how satisfied they were with ease of deflation of their implant. Success criteria was a response of "satisfactory" or "somewhat satisfactory". Other possible responses were "neither satisfactory nor unsatisfactory", "somewhat unsatisfactory", and "very unsatisfactory".
Secondary Outcome Measures
- Evaluate User Acceptance of Titan® OTR - Question 1 [3 months post-surgery]
Subject Satisfaction - overall function
- Evaluate User Acceptance of Titan® OTR - Question 1 [6 months post-surgery]
Subject Satisfaction - overall function
- Evaluate User Acceptance of Titan® OTR - Question 1 [12 months post-surgery]
Subject Satisfaction - overall function
- Evaluate User Acceptance of Titan® OTR - Question 2 [3 months post-surgery]
Subject Satisfaction - soft enough to conceal when deflated
- Evaluate User Acceptance of Titan® OTR - Question 2 [6 months post-surgery]
Subject Satisfaction - soft enough to conceal when deflated
- Evaluate User Acceptance of Titan® OTR - Question 2 [12 months post-surgery]
Subject Satisfaction - soft enough to conceal when deflated
- Evaluate User Acceptance of Titan® OTR - Question 3 [3 months post-surgery]
Subject Satisfaction - ease of locating the deflation touch pads
- Evaluate User Acceptance of Titan® OTR - Question 3 [6 months post-surgery]
Subject Satisfaction - ease of locating the deflation touch pads
- Evaluate User Acceptance of Titan® OTR - Question 3 [12 months post-surgery]
Subject Satisfaction - ease of locating the deflation touch pads
- Evaluate User Acceptance of Titan® OTR - Question 4 [3 months post-surgery]
Subject Satisfaction - ease of inflation
- Evaluate User Acceptance of Titan® OTR - Question 4 [6 months post-surgery]
Subject Satisfaction - ease of inflation
- Evaluate User Acceptance of Titan® OTR - Question 4 [12 months post-surgery]
Subject Satisfaction - ease of inflation
- Evaluate User Acceptance of Titan® OTR - Question 5 [3 months post-surgery]
Subject Satisfaction - ease of deflation
- Evaluate User Acceptance of Titan® OTR - Question 5 [6 months post-surgery]
Subject Satisfaction - ease of deflation
- Evaluate User Acceptance of Titan® OTR - Question 5 [12 months post-surgery]
Subject Satisfaction - ease of deflation
- Evaluate User Acceptance of Titan® OTR - Question 6 [3 months post-surgery]
Subject Satisfaction - hardness of erection when inflated
- Evaluate User Acceptance of Titan® OTR - Question 6 [6 months post-surgery]
Subject Satisfaction - hardness of erection when inflated
- Evaluate User Acceptance of Titan® OTR - Question 6 [12 months post-surgery]
Subject Satisfaction - hardness of erection when inflated
- Evaluate User Acceptance of Titan® OTR - Question 7 [3 months post-surgery]
Subject Satisfaction - width when inflated
- Evaluate User Acceptance of Titan® OTR - Question 7 [6 months post-surgery]
Subject Satisfaction - width when inflated
- Evaluate User Acceptance of Titan® OTR - Question 7 [12 months post-surgery]
Subject Satisfaction - width when inflated
- Evaluate User Acceptance of Titan® OTR - Question 8 [3 months post-surgery]
Subject Satisfaction - length when inflated
- Evaluate User Acceptance of Titan® OTR - Question 8 [6 months post-surgery]
Subject Satisfaction - length when inflated
- Evaluate User Acceptance of Titan® OTR - Question 8 [12 months post-surgery]
Subject Satisfaction - length when inflated
- Evaluate User Acceptance of Titan® OTR - Question 9 [3 months post-surgery]
Subject Satisfaction - would you recommend this penile implant device to men with the same erectile difficulty that you had?
- Evaluate User Acceptance of Titan® OTR - Question 9 [6 months post-surgery]
Subject Satisfaction - would you recommend this penile implant device to men with the same erectile difficulty that you had?
- Evaluate User Acceptance of Titan® OTR - Question 9 [12 months post-surgery]
Subject Satisfaction - would you recommend this penile implant device to men with the same erectile difficulty that you had?
- Evaluate User Acceptance of Titan® OTR - Question 10 [3 months post-surgery]
Subject Satisfaction - if you had the decision to make again, would you undergo this penile implant procedure again?
- Evaluate User Acceptance of Titan® OTR - Question 10 [6 months post-surgery]
Subject Satisfaction - if you had the decision to make again, would you undergo this penile implant procedure again?
- Evaluate User Acceptance of Titan® OTR - Question 10 [12 months post-surgery]
Subject Satisfaction - if you had the decision to make again, would you undergo this penile implant procedure again?
- Assess the Ease of Implant of the Titan® OTR Question 1 [At implant]
Titan OTR pre-implant product preparation was straightforward/simple?
- Assess the Ease of Implant of the Titan® OTR - Question 2 [At implant]
Titan OTR pre-implant product preparation was easier than your usual pump of choice?
- Assess the Ease of Implant of the Titan® OTR - Question 3 [At implant]
The subject was easily able to accommodate the OTR pump?
- Assess the Ease of Training for Titan® OTR - Question 1 [6 weeks]
It was easy for the subject to find the inflation bulb?
- Assess the Ease of Training for Titan® OTR - Question 2 [6 weeks]
It was easy for the subject to find the deflation touch pads?
- Assess the Ease of Training for Titan® OTR - Question 3 [6 weeks]
It was easy for the subject to inflate the device?
- Assess the Ease of Training for Titan® OTR - Question 4 [6 weeks]
It was easy for the subject to compress the deflation touch pads?
- Assess the Ease of Training for Titan® OTR - Question 5 [6 weeks]
Subject training with OTR pump was easier than with previous pump?
- Assess the Ease of Training for Titan® OTR - Question 6 [6 weeks]
The OTR pump was easy to use at 1st cycling?
- Assess the Ease of Training for Titan® OTR - Question 7 [6 weeks]
The subject likes the OTR pump?
- Assess the Ease of Training for Titan® OTR - Question 8 [6 weeks]
How easy was it for the subject to learn?
- Assess the Ease of Training for Titan® OTR - Question 9 [6 weeks]
How likely the subject will need continued training or retraining?
- Assess Partner Satisfaction (Where Applicable) [Baseline]
The Partner Treatment Satisfaction Scale (TSS) was used. The TSS provides scores ranging from 0 to 100 in 5 domains of Ease of erection, Erectile function, Pleasure from sexual activity, Satisfaction with orgasm and Confidence to complete sexual activity with higher scores indicative of worse symptoms. Available data is summarized for each domain at each visit along with change from baseline.
- Assess Partner Satisfaction (Where Applicable) [6 months]
The Partner Treatment Satisfaction Scale (TSS) was used. The TSS provides scores ranging from 0 to 100 in 5 domains of Ease of erection, Erectile function, Pleasure from sexual activity, Satisfaction with orgasm and Confidence to complete sexual activity with higher scores indicative of worse symptoms. Available data is summarized for each domain at each visit along with change from baseline.
Eligibility Criteria
Criteria
Inclusion Criteria:
-
Participant has an estimated life expectancy of more than 5 years
-
Participant has been diagnosed with erectile dysfunction (impotence)
-
participant is willing to have the Titan OTR IPP implanted
-
Participant is able and willing to complete all follow-up visits and procedures indicated in the protocol
Exclusion Criteria:
-
Participant is unable or unwilling to sign the Informed Consent Form and /or comply with all follow-up requirements according to the study protocol
-
Participant had a previous penile prosthesis or prior penile enlargement surgeries
-
Participant has a compromised immune system, such as systemic lupus erythematosus, discoid lupus, or scleroderma
-
Participant has had a myocardial infarction or coronary artery stent placement within 2 months prior to treatment
-
Participant currently has uncontrolled or severe diabetes or other disease significantly reducing wound healing ability
-
Participant does not have the manual dexterity or mental ability to operate the pump
-
Participant has chordee or priapism
-
Participant has penile sensory neuropathy
-
Participant has serious bleeding disorder
Contacts and Locations
Locations
Site | City | State | Country | Postal Code | |
---|---|---|---|---|---|
1 | Emory Healthcare | Altanta | Georgia | United States | 30322 |
2 | Rush Presbyterian | Chicago | Illinois | United States | 60612 |
3 | Boston Medical Center | Boston | Massachusetts | United States | 02118 |
4 | Michigan Urology Center | Ann Arbor | Michigan | United States | 48109 |
5 | Carolina Urologic Research Center | Myrtle Beach | South Carolina | United States | 29572 |
6 | Urology San Antonio | San Antonio | Texas | United States | 78229 |
7 | St Joseph's Hospital | London | Ontario | Canada | N6A 4V2 |
8 | University College London Hospital | London | UK | United Kingdom | W1G 6BJ |
Sponsors and Collaborators
- Coloplast A/S
Investigators
- Principal Investigator: Gerald Brock, MD, St Joseph's Hospital
- Principal Investigator: Ricardo Munarriz, MD, Boston Medical Center
- Principal Investigator: Lawrence Levine, MD, Rush Presbyterian
- Principal Investigator: Chad Ritenour, MD, Emory Healthcare
- Principal Investigator: LeRoy Jones, MD, Urology San Antonio
- Principal Investigator: William Bogache, MD, Carolina Urologic Research Center
- Principal Investigator: Dana Ohl, MD, Michigan Urology Center
- Principal Investigator: David Ralph, MD, University College London Hospital
Study Documents (Full-Text)
None provided.More Information
Publications
None provided.- DK057CC
Study Results
Participant Flow
Recruitment Details | |
---|---|
Pre-assignment Detail |
Arm/Group Title | Titan® OTR IPP |
---|---|
Arm/Group Description | Subjects implanted with Titan® One Touch Release (OTR) Inflatable Penile Prosthesis (IPP) Inflatable Penile Prosthesis : Hydraulic system designed to be surgically implanted into the penis for the management of erectile dysfunction. |
Period Title: Overall Study | |
STARTED | 124 |
COMPLETED | 90 |
NOT COMPLETED | 34 |
Baseline Characteristics
Arm/Group Title | Titan® OTR IPP |
---|---|
Arm/Group Description | Subjects implanted with Titan® One Touch Release (OTR) Inflatable Penile Prosthesis (IPP) Inflatable Penile Prosthesis : Hydraulic system designed to be surgically implanted into the penis for the management of erectile dysfunction. |
Overall Participants | 124 |
Age (Count of Participants) | |
<=18 years |
0
0%
|
Between 18 and 65 years |
76
61.3%
|
>=65 years |
48
38.7%
|
Age (years) [Mean (Standard Deviation) ] | |
Mean (Standard Deviation) [years] |
61.0
(9.1)
|
Sex: Female, Male (Count of Participants) | |
Female |
0
0%
|
Male |
124
100%
|
Region of Enrollment (participants) [Number] | |
United States |
71
57.3%
|
Canada |
27
21.8%
|
United Kingdom |
26
21%
|
Outcome Measures
Title | Assess the Ease of Deflation of the Titan® OTR Pump |
---|---|
Description | The study's primary endpoint was to demonstrate that at least 64% of subjects were mostly or completely satisfied with the ability to deflate the device at the 6-month follow-up. The study's primary objective was to assess the ease of deflation of the Titan® OTR pump via subject questionnaire at 6-month follow-up. Subjects were asked via questionnaire how satisfied they were with ease of deflation of their implant. Success criteria was a response of "satisfactory" or "somewhat satisfactory". Other possible responses were "neither satisfactory nor unsatisfactory", "somewhat unsatisfactory", and "very unsatisfactory". |
Time Frame | 6 months |
Outcome Measure Data
Analysis Population Description |
---|
[Not Specified] |
Arm/Group Title | Titan® OTR IPP |
---|---|
Arm/Group Description | Subjects implanted with Titan® One Touch Release (OTR) Inflatable Penile Prosthesis (IPP) Inflatable Penile Prosthesis : Hydraulic system designed to be surgically implanted into the penis for the management of erectile dysfunction. |
Measure Participants | 96 |
Number [% of participants meeting success criter] |
70.8
57.1%
|
Title | Evaluate User Acceptance of Titan® OTR - Question 1 |
---|---|
Description | Subject Satisfaction - overall function |
Time Frame | 3 months post-surgery |
Outcome Measure Data
Analysis Population Description |
---|
[Not Specified] |
Arm/Group Title | Titan® OTR IPP |
---|---|
Arm/Group Description | Subjects implanted with Titan® One Touch Release (OTR) Inflatable Penile Prosthesis (IPP) Inflatable Penile Prosthesis : Hydraulic system designed to be surgically implanted into the penis for the management of erectile dysfunction. |
Measure Participants | 94 |
Number [% satisfactory or somewhat satisfactory] |
72.4
|
Title | Evaluate User Acceptance of Titan® OTR - Question 1 |
---|---|
Description | Subject Satisfaction - overall function |
Time Frame | 6 months post-surgery |
Outcome Measure Data
Analysis Population Description |
---|
[Not Specified] |
Arm/Group Title | Titan® OTR IPP |
---|---|
Arm/Group Description | Subjects implanted with Titan® One Touch Release (OTR) Inflatable Penile Prosthesis (IPP) Inflatable Penile Prosthesis : Hydraulic system designed to be surgically implanted into the penis for the management of erectile dysfunction. |
Measure Participants | 96 |
Number [% satisfactory or somewhat satisfactory] |
90.6
|
Title | Evaluate User Acceptance of Titan® OTR - Question 1 |
---|---|
Description | Subject Satisfaction - overall function |
Time Frame | 12 months post-surgery |
Outcome Measure Data
Analysis Population Description |
---|
[Not Specified] |
Arm/Group Title | Titan® OTR IPP |
---|---|
Arm/Group Description | Subjects implanted with Titan® One Touch Release (OTR) Inflatable Penile Prosthesis (IPP) Inflatable Penile Prosthesis : Hydraulic system designed to be surgically implanted into the penis for the management of erectile dysfunction. |
Measure Participants | 90 |
Number [% satisfactory or somewhat satisfactory] |
90.0
|
Title | Evaluate User Acceptance of Titan® OTR - Question 2 |
---|---|
Description | Subject Satisfaction - soft enough to conceal when deflated |
Time Frame | 3 months post-surgery |
Outcome Measure Data
Analysis Population Description |
---|
[Not Specified] |
Arm/Group Title | Titan® OTR IPP |
---|---|
Arm/Group Description | Subjects implanted with Titan® One Touch Release (OTR) Inflatable Penile Prosthesis (IPP) Inflatable Penile Prosthesis : Hydraulic system designed to be surgically implanted into the penis for the management of erectile dysfunction. |
Measure Participants | 94 |
Number [% satisfactory or somewhat satisfactory] |
62.8
|
Title | Evaluate User Acceptance of Titan® OTR - Question 2 |
---|---|
Description | Subject Satisfaction - soft enough to conceal when deflated |
Time Frame | 6 months post-surgery |
Outcome Measure Data
Analysis Population Description |
---|
[Not Specified] |
Arm/Group Title | Titan® OTR IPP |
---|---|
Arm/Group Description | Subjects implanted with Titan® One Touch Release (OTR) Inflatable Penile Prosthesis (IPP) Inflatable Penile Prosthesis : Hydraulic system designed to be surgically implanted into the penis for the management of erectile dysfunction. |
Measure Participants | 96 |
Number [% satisfactory or somewhat satisfactory] |
76.0
|
Title | Evaluate User Acceptance of Titan® OTR - Question 2 |
---|---|
Description | Subject Satisfaction - soft enough to conceal when deflated |
Time Frame | 12 months post-surgery |
Outcome Measure Data
Analysis Population Description |
---|
[Not Specified] |
Arm/Group Title | Titan® OTR IPP |
---|---|
Arm/Group Description | Subjects implanted with Titan® One Touch Release (OTR) Inflatable Penile Prosthesis (IPP) Inflatable Penile Prosthesis : Hydraulic system designed to be surgically implanted into the penis for the management of erectile dysfunction. |
Measure Participants | 90 |
Number [% satisfactory or somewhat satisfactory] |
82.2
|
Title | Evaluate User Acceptance of Titan® OTR - Question 3 |
---|---|
Description | Subject Satisfaction - ease of locating the deflation touch pads |
Time Frame | 3 months post-surgery |
Outcome Measure Data
Analysis Population Description |
---|
[Not Specified] |
Arm/Group Title | Titan® OTR IPP |
---|---|
Arm/Group Description | Subjects implanted with Titan® One Touch Release (OTR) Inflatable Penile Prosthesis (IPP) Inflatable Penile Prosthesis : Hydraulic system designed to be surgically implanted into the penis for the management of erectile dysfunction. |
Measure Participants | 94 |
Number [% satisfactory or somewhat satisfactory] |
60.6
|
Title | Evaluate User Acceptance of Titan® OTR - Question 3 |
---|---|
Description | Subject Satisfaction - ease of locating the deflation touch pads |
Time Frame | 6 months post-surgery |
Outcome Measure Data
Analysis Population Description |
---|
[Not Specified] |
Arm/Group Title | Titan® OTR IPP |
---|---|
Arm/Group Description | Subjects implanted with Titan® One Touch Release (OTR) Inflatable Penile Prosthesis (IPP) Inflatable Penile Prosthesis : Hydraulic system designed to be surgically implanted into the penis for the management of erectile dysfunction. |
Measure Participants | 96 |
Number [% satisfactory or somewhat satisfactory] |
78.2
|
Title | Evaluate User Acceptance of Titan® OTR - Question 3 |
---|---|
Description | Subject Satisfaction - ease of locating the deflation touch pads |
Time Frame | 12 months post-surgery |
Outcome Measure Data
Analysis Population Description |
---|
[Not Specified] |
Arm/Group Title | Titan® OTR IPP |
---|---|
Arm/Group Description | Subjects implanted with Titan® One Touch Release (OTR) Inflatable Penile Prosthesis (IPP) Inflatable Penile Prosthesis : Hydraulic system designed to be surgically implanted into the penis for the management of erectile dysfunction. |
Measure Participants | 89 |
Number [% satisfactory or somewhat satisfactory] |
77.6
|
Title | Evaluate User Acceptance of Titan® OTR - Question 4 |
---|---|
Description | Subject Satisfaction - ease of inflation |
Time Frame | 3 months post-surgery |
Outcome Measure Data
Analysis Population Description |
---|
[Not Specified] |
Arm/Group Title | Titan® OTR IPP |
---|---|
Arm/Group Description | Subjects implanted with Titan® One Touch Release (OTR) Inflatable Penile Prosthesis (IPP) Inflatable Penile Prosthesis : Hydraulic system designed to be surgically implanted into the penis for the management of erectile dysfunction. |
Measure Participants | 94 |
Number [% satisfactory or somewhat satisfactory] |
68.1
|
Title | Evaluate User Acceptance of Titan® OTR - Question 4 |
---|---|
Description | Subject Satisfaction - ease of inflation |
Time Frame | 6 months post-surgery |
Outcome Measure Data
Analysis Population Description |
---|
[Not Specified] |
Arm/Group Title | Titan® OTR IPP |
---|---|
Arm/Group Description | Subjects implanted with Titan® One Touch Release (OTR) Inflatable Penile Prosthesis (IPP) Inflatable Penile Prosthesis : Hydraulic system designed to be surgically implanted into the penis for the management of erectile dysfunction. |
Measure Participants | 95 |
Number [% satisfactory or somewhat satisfactory] |
85.3
|
Title | Evaluate User Acceptance of Titan® OTR - Question 4 |
---|---|
Description | Subject Satisfaction - ease of inflation |
Time Frame | 12 months post-surgery |
Outcome Measure Data
Analysis Population Description |
---|
[Not Specified] |
Arm/Group Title | Titan® OTR IPP |
---|---|
Arm/Group Description | Subjects implanted with Titan® One Touch Release (OTR) Inflatable Penile Prosthesis (IPP) Inflatable Penile Prosthesis : Hydraulic system designed to be surgically implanted into the penis for the management of erectile dysfunction. |
Measure Participants | 88 |
Number [% satisfactory or somewhat satisfactory] |
86.4
|
Title | Evaluate User Acceptance of Titan® OTR - Question 5 |
---|---|
Description | Subject Satisfaction - ease of deflation |
Time Frame | 3 months post-surgery |
Outcome Measure Data
Analysis Population Description |
---|
[Not Specified] |
Arm/Group Title | Titan® OTR IPP |
---|---|
Arm/Group Description | Subjects implanted with Titan® One Touch Release (OTR) Inflatable Penile Prosthesis (IPP) Inflatable Penile Prosthesis : Hydraulic system designed to be surgically implanted into the penis for the management of erectile dysfunction. |
Measure Participants | 94 |
Number [% satisfactory or somewhat satisfactory] |
60.6
|
Title | Evaluate User Acceptance of Titan® OTR - Question 5 |
---|---|
Description | Subject Satisfaction - ease of deflation |
Time Frame | 6 months post-surgery |
Outcome Measure Data
Analysis Population Description |
---|
[Not Specified] |
Arm/Group Title | Titan® OTR IPP |
---|---|
Arm/Group Description | Subjects implanted with Titan® One Touch Release (OTR) Inflatable Penile Prosthesis (IPP) Inflatable Penile Prosthesis : Hydraulic system designed to be surgically implanted into the penis for the management of erectile dysfunction. |
Measure Participants | 96 |
Number [% satisfactory or somewhat satisfactory] |
70.8
|
Title | Evaluate User Acceptance of Titan® OTR - Question 5 |
---|---|
Description | Subject Satisfaction - ease of deflation |
Time Frame | 12 months post-surgery |
Outcome Measure Data
Analysis Population Description |
---|
[Not Specified] |
Arm/Group Title | Titan® OTR IPP |
---|---|
Arm/Group Description | Subjects implanted with Titan® One Touch Release (OTR) Inflatable Penile Prosthesis (IPP) Inflatable Penile Prosthesis : Hydraulic system designed to be surgically implanted into the penis for the management of erectile dysfunction. |
Measure Participants | 90 |
Number [% satisfactory or somewhat satisfactory] |
73.3
|
Title | Evaluate User Acceptance of Titan® OTR - Question 6 |
---|---|
Description | Subject Satisfaction - hardness of erection when inflated |
Time Frame | 3 months post-surgery |
Outcome Measure Data
Analysis Population Description |
---|
[Not Specified] |
Arm/Group Title | Titan® OTR IPP |
---|---|
Arm/Group Description | Subjects implanted with Titan® One Touch Release (OTR) Inflatable Penile Prosthesis (IPP) Inflatable Penile Prosthesis : Hydraulic system designed to be surgically implanted into the penis for the management of erectile dysfunction. |
Measure Participants | 94 |
Number [% satisfactory or somewhat satisfactory] |
82.0
|
Title | Evaluate User Acceptance of Titan® OTR - Question 6 |
---|---|
Description | Subject Satisfaction - hardness of erection when inflated |
Time Frame | 6 months post-surgery |
Outcome Measure Data
Analysis Population Description |
---|
[Not Specified] |
Arm/Group Title | Titan® OTR IPP |
---|---|
Arm/Group Description | Subjects implanted with Titan® One Touch Release (OTR) Inflatable Penile Prosthesis (IPP) Inflatable Penile Prosthesis : Hydraulic system designed to be surgically implanted into the penis for the management of erectile dysfunction. |
Measure Participants | 94 |
Number [% satisfactory or somewhat satisfactory] |
91.5
|
Title | Evaluate User Acceptance of Titan® OTR - Question 6 |
---|---|
Description | Subject Satisfaction - hardness of erection when inflated |
Time Frame | 12 months post-surgery |
Outcome Measure Data
Analysis Population Description |
---|
[Not Specified] |
Arm/Group Title | Titan® OTR IPP |
---|---|
Arm/Group Description | Subjects implanted with Titan® One Touch Release (OTR) Inflatable Penile Prosthesis (IPP) Inflatable Penile Prosthesis : Hydraulic system designed to be surgically implanted into the penis for the management of erectile dysfunction. |
Measure Participants | 89 |
Number [% satisfactory or somewhat satisfactory] |
93.2
|
Title | Evaluate User Acceptance of Titan® OTR - Question 7 |
---|---|
Description | Subject Satisfaction - width when inflated |
Time Frame | 3 months post-surgery |
Outcome Measure Data
Analysis Population Description |
---|
[Not Specified] |
Arm/Group Title | Titan® OTR IPP |
---|---|
Arm/Group Description | Subjects implanted with Titan® One Touch Release (OTR) Inflatable Penile Prosthesis (IPP) Inflatable Penile Prosthesis : Hydraulic system designed to be surgically implanted into the penis for the management of erectile dysfunction. |
Measure Participants | 94 |
Number [% satisfactory or somewhat satisfactory] |
77.6
|
Title | Evaluate User Acceptance of Titan® OTR - Question 7 |
---|---|
Description | Subject Satisfaction - width when inflated |
Time Frame | 6 months post-surgery |
Outcome Measure Data
Analysis Population Description |
---|
[Not Specified] |
Arm/Group Title | Titan® OTR IPP |
---|---|
Arm/Group Description | Subjects implanted with Titan® One Touch Release (OTR) Inflatable Penile Prosthesis (IPP) Inflatable Penile Prosthesis : Hydraulic system designed to be surgically implanted into the penis for the management of erectile dysfunction. |
Measure Participants | 94 |
Number [% satisfactory or somewhat satisfactory] |
86.1
|
Title | Evaluate User Acceptance of Titan® OTR - Question 7 |
---|---|
Description | Subject Satisfaction - width when inflated |
Time Frame | 12 months post-surgery |
Outcome Measure Data
Analysis Population Description |
---|
[Not Specified] |
Arm/Group Title | Titan® OTR IPP |
---|---|
Arm/Group Description | Subjects implanted with Titan® One Touch Release (OTR) Inflatable Penile Prosthesis (IPP) Inflatable Penile Prosthesis : Hydraulic system designed to be surgically implanted into the penis for the management of erectile dysfunction. |
Measure Participants | 89 |
Number [% satisfactory or somewhat satisfactory] |
79.8
|
Title | Evaluate User Acceptance of Titan® OTR - Question 8 |
---|---|
Description | Subject Satisfaction - length when inflated |
Time Frame | 3 months post-surgery |
Outcome Measure Data
Analysis Population Description |
---|
[Not Specified] |
Arm/Group Title | Titan® OTR IPP |
---|---|
Arm/Group Description | Subjects implanted with Titan® One Touch Release (OTR) Inflatable Penile Prosthesis (IPP) Inflatable Penile Prosthesis : Hydraulic system designed to be surgically implanted into the penis for the management of erectile dysfunction. |
Measure Participants | 94 |
Number [% satisfactory or somewhat satisfactory] |
54.3
|
Title | Evaluate User Acceptance of Titan® OTR - Question 8 |
---|---|
Description | Subject Satisfaction - length when inflated |
Time Frame | 6 months post-surgery |
Outcome Measure Data
Analysis Population Description |
---|
[Not Specified] |
Arm/Group Title | Titan® OTR IPP |
---|---|
Arm/Group Description | Subjects implanted with Titan® One Touch Release (OTR) Inflatable Penile Prosthesis (IPP) Inflatable Penile Prosthesis : Hydraulic system designed to be surgically implanted into the penis for the management of erectile dysfunction. |
Measure Participants | 95 |
Number [% satisfactory or somewhat satisfactory] |
61.1
|
Title | Evaluate User Acceptance of Titan® OTR - Question 8 |
---|---|
Description | Subject Satisfaction - length when inflated |
Time Frame | 12 months post-surgery |
Outcome Measure Data
Analysis Population Description |
---|
[Not Specified] |
Arm/Group Title | Titan® OTR IPP |
---|---|
Arm/Group Description | Subjects implanted with Titan® One Touch Release (OTR) Inflatable Penile Prosthesis (IPP) Inflatable Penile Prosthesis : Hydraulic system designed to be surgically implanted into the penis for the management of erectile dysfunction. |
Measure Participants | 90 |
Number [% satisfactory or somewhat satisfactory] |
68.9
|
Title | Evaluate User Acceptance of Titan® OTR - Question 9 |
---|---|
Description | Subject Satisfaction - would you recommend this penile implant device to men with the same erectile difficulty that you had? |
Time Frame | 3 months post-surgery |
Outcome Measure Data
Analysis Population Description |
---|
[Not Specified] |
Arm/Group Title | Titan® OTR IPP |
---|---|
Arm/Group Description | Subjects implanted with Titan® One Touch Release (OTR) Inflatable Penile Prosthesis (IPP) Inflatable Penile Prosthesis : Hydraulic system designed to be surgically implanted into the penis for the management of erectile dysfunction. |
Measure Participants | 94 |
Number [% responding "yes" and "probably"] |
79.8
|
Title | Evaluate User Acceptance of Titan® OTR - Question 9 |
---|---|
Description | Subject Satisfaction - would you recommend this penile implant device to men with the same erectile difficulty that you had? |
Time Frame | 6 months post-surgery |
Outcome Measure Data
Analysis Population Description |
---|
[Not Specified] |
Arm/Group Title | Titan® OTR IPP |
---|---|
Arm/Group Description | Subjects implanted with Titan® One Touch Release (OTR) Inflatable Penile Prosthesis (IPP) Inflatable Penile Prosthesis : Hydraulic system designed to be surgically implanted into the penis for the management of erectile dysfunction. |
Measure Participants | 96 |
Number [% responding "yes" and "probably"] |
86.5
|
Title | Evaluate User Acceptance of Titan® OTR - Question 9 |
---|---|
Description | Subject Satisfaction - would you recommend this penile implant device to men with the same erectile difficulty that you had? |
Time Frame | 12 months post-surgery |
Outcome Measure Data
Analysis Population Description |
---|
[Not Specified] |
Arm/Group Title | Titan® OTR IPP |
---|---|
Arm/Group Description | Subjects implanted with Titan® One Touch Release (OTR) Inflatable Penile Prosthesis (IPP) Inflatable Penile Prosthesis : Hydraulic system designed to be surgically implanted into the penis for the management of erectile dysfunction. |
Measure Participants | 90 |
Number [% responding "yes" and "probably"] |
87.8
|
Title | Evaluate User Acceptance of Titan® OTR - Question 10 |
---|---|
Description | Subject Satisfaction - if you had the decision to make again, would you undergo this penile implant procedure again? |
Time Frame | 3 months post-surgery |
Outcome Measure Data
Analysis Population Description |
---|
[Not Specified] |
Arm/Group Title | Titan® OTR IPP |
---|---|
Arm/Group Description | Subjects implanted with Titan® One Touch Release (OTR) Inflatable Penile Prosthesis (IPP) Inflatable Penile Prosthesis : Hydraulic system designed to be surgically implanted into the penis for the management of erectile dysfunction. |
Measure Participants | 94 |
Number [% responding "yes" and "probably"] |
72.4
|
Title | Evaluate User Acceptance of Titan® OTR - Question 10 |
---|---|
Description | Subject Satisfaction - if you had the decision to make again, would you undergo this penile implant procedure again? |
Time Frame | 6 months post-surgery |
Outcome Measure Data
Analysis Population Description |
---|
[Not Specified] |
Arm/Group Title | Titan® OTR IPP |
---|---|
Arm/Group Description | Subjects implanted with Titan® One Touch Release (OTR) Inflatable Penile Prosthesis (IPP) Inflatable Penile Prosthesis : Hydraulic system designed to be surgically implanted into the penis for the management of erectile dysfunction. |
Measure Participants | 96 |
Number [% responding "yes" and "probably"] |
83.3
|
Title | Evaluate User Acceptance of Titan® OTR - Question 10 |
---|---|
Description | Subject Satisfaction - if you had the decision to make again, would you undergo this penile implant procedure again? |
Time Frame | 12 months post-surgery |
Outcome Measure Data
Analysis Population Description |
---|
[Not Specified] |
Arm/Group Title | Titan® OTR IPP |
---|---|
Arm/Group Description | Subjects implanted with Titan® One Touch Release (OTR) Inflatable Penile Prosthesis (IPP) Inflatable Penile Prosthesis : Hydraulic system designed to be surgically implanted into the penis for the management of erectile dysfunction. |
Measure Participants | 90 |
Number [% responding "yes" and "probably"] |
86.7
|
Title | Assess the Ease of Implant of the Titan® OTR Question 1 |
---|---|
Description | Titan OTR pre-implant product preparation was straightforward/simple? |
Time Frame | At implant |
Outcome Measure Data
Analysis Population Description |
---|
[Not Specified] |
Arm/Group Title | Titan® OTR IPP |
---|---|
Arm/Group Description | Subjects implanted with Titan® One Touch Release (OTR) Inflatable Penile Prosthesis (IPP) Inflatable Penile Prosthesis : Hydraulic system designed to be surgically implanted into the penis for the management of erectile dysfunction. |
Measure Participants | 113 |
Completely disagree |
0.0
|
Somewhat disagree |
0.0
|
Neutral |
2.7
|
Somewhat agree |
22.1
|
Completely agree |
75.2
|
Title | Assess the Ease of Implant of the Titan® OTR - Question 2 |
---|---|
Description | Titan OTR pre-implant product preparation was easier than your usual pump of choice? |
Time Frame | At implant |
Outcome Measure Data
Analysis Population Description |
---|
[Not Specified] |
Arm/Group Title | Titan® OTR IPP |
---|---|
Arm/Group Description | Subjects implanted with Titan® One Touch Release (OTR) Inflatable Penile Prosthesis (IPP) Inflatable Penile Prosthesis : Hydraulic system designed to be surgically implanted into the penis for the management of erectile dysfunction. |
Measure Participants | 113 |
Completely disagree |
0.0
|
Somewhat disagree |
0.0
|
Neutral |
10.6
|
Somewhat agree |
40.7
|
Completely agree |
48.7
|
Title | Assess the Ease of Implant of the Titan® OTR - Question 3 |
---|---|
Description | The subject was easily able to accommodate the OTR pump? |
Time Frame | At implant |
Outcome Measure Data
Analysis Population Description |
---|
[Not Specified] |
Arm/Group Title | Titan® OTR IPP |
---|---|
Arm/Group Description | Subjects implanted with Titan® One Touch Release (OTR) Inflatable Penile Prosthesis (IPP) Inflatable Penile Prosthesis : Hydraulic system designed to be surgically implanted into the penis for the management of erectile dysfunction. |
Measure Participants | 113 |
Completely disagree |
0.0
|
Somewhat disagree |
0.0
|
Neutral |
2.7
|
Somewhat agree |
20.4
|
Completely agree |
77.0
|
Title | Assess the Ease of Training for Titan® OTR - Question 1 |
---|---|
Description | It was easy for the subject to find the inflation bulb? |
Time Frame | 6 weeks |
Outcome Measure Data
Analysis Population Description |
---|
[Not Specified] |
Arm/Group Title | Titan® OTR IPP |
---|---|
Arm/Group Description | Subjects implanted with Titan® One Touch Release (OTR) Inflatable Penile Prosthesis (IPP) Inflatable Penile Prosthesis : Hydraulic system designed to be surgically implanted into the penis for the management of erectile dysfunction. |
Measure Participants | 108 |
Not at all |
0.0
0%
|
A little |
0.9
0.7%
|
Somewhat |
4.6
3.7%
|
Very |
13.9
11.2%
|
Extremely |
80.6
65%
|
Title | Assess the Ease of Training for Titan® OTR - Question 2 |
---|---|
Description | It was easy for the subject to find the deflation touch pads? |
Time Frame | 6 weeks |
Outcome Measure Data
Analysis Population Description |
---|
[Not Specified] |
Arm/Group Title | Titan® OTR IPP |
---|---|
Arm/Group Description | Subjects implanted with Titan® One Touch Release (OTR) Inflatable Penile Prosthesis (IPP) Inflatable Penile Prosthesis : Hydraulic system designed to be surgically implanted into the penis for the management of erectile dysfunction. |
Measure Participants | 108 |
Not at all |
1.9
1.5%
|
A little |
3.7
3%
|
Somewhat |
9.3
7.5%
|
Very |
26.9
21.7%
|
Extremely |
58.3
47%
|
Title | Assess the Ease of Training for Titan® OTR - Question 3 |
---|---|
Description | It was easy for the subject to inflate the device? |
Time Frame | 6 weeks |
Outcome Measure Data
Analysis Population Description |
---|
[Not Specified] |
Arm/Group Title | Titan® OTR IPP |
---|---|
Arm/Group Description | Subjects implanted with Titan® One Touch Release (OTR) Inflatable Penile Prosthesis (IPP) Inflatable Penile Prosthesis : Hydraulic system designed to be surgically implanted into the penis for the management of erectile dysfunction. |
Measure Participants | 108 |
Not at all |
1.9
1.5%
|
A little |
0.9
0.7%
|
Somewhat |
4.6
3.7%
|
Very |
17.6
14.2%
|
Extremely |
75.0
60.5%
|
Title | Assess the Ease of Training for Titan® OTR - Question 4 |
---|---|
Description | It was easy for the subject to compress the deflation touch pads? |
Time Frame | 6 weeks |
Outcome Measure Data
Analysis Population Description |
---|
[Not Specified] |
Arm/Group Title | Titan® OTR IPP |
---|---|
Arm/Group Description | Subjects implanted with Titan® One Touch Release (OTR) Inflatable Penile Prosthesis (IPP) Inflatable Penile Prosthesis : Hydraulic system designed to be surgically implanted into the penis for the management of erectile dysfunction. |
Measure Participants | 108 |
Not at all |
2.8
2.3%
|
A little |
1.9
1.5%
|
Somewhat |
10.2
8.2%
|
Very |
25.9
20.9%
|
Extremely |
59.3
47.8%
|
Title | Assess the Ease of Training for Titan® OTR - Question 5 |
---|---|
Description | Subject training with OTR pump was easier than with previous pump? |
Time Frame | 6 weeks |
Outcome Measure Data
Analysis Population Description |
---|
[Not Specified] |
Arm/Group Title | Titan® OTR IPP |
---|---|
Arm/Group Description | Subjects implanted with Titan® One Touch Release (OTR) Inflatable Penile Prosthesis (IPP) Inflatable Penile Prosthesis : Hydraulic system designed to be surgically implanted into the penis for the management of erectile dysfunction. |
Measure Participants | 108 |
Not at all |
1.9
1.5%
|
A little |
1.9
1.5%
|
Somewhat |
16.7
13.5%
|
Very |
16.7
13.5%
|
Extremely |
63.0
50.8%
|
Title | Assess the Ease of Training for Titan® OTR - Question 6 |
---|---|
Description | The OTR pump was easy to use at 1st cycling? |
Time Frame | 6 weeks |
Outcome Measure Data
Analysis Population Description |
---|
[Not Specified] |
Arm/Group Title | Titan® OTR IPP |
---|---|
Arm/Group Description | Subjects implanted with Titan® One Touch Release (OTR) Inflatable Penile Prosthesis (IPP) Inflatable Penile Prosthesis : Hydraulic system designed to be surgically implanted into the penis for the management of erectile dysfunction. |
Measure Participants | 108 |
Not at all |
1.9
1.5%
|
A little |
5.6
4.5%
|
Somewhat |
8.3
6.7%
|
Very |
21.3
17.2%
|
Extremely |
63.0
50.8%
|
Title | Assess the Ease of Training for Titan® OTR - Question 7 |
---|---|
Description | The subject likes the OTR pump? |
Time Frame | 6 weeks |
Outcome Measure Data
Analysis Population Description |
---|
[Not Specified] |
Arm/Group Title | Titan® OTR IPP |
---|---|
Arm/Group Description | Subjects implanted with Titan® One Touch Release (OTR) Inflatable Penile Prosthesis (IPP) Inflatable Penile Prosthesis : Hydraulic system designed to be surgically implanted into the penis for the management of erectile dysfunction. |
Measure Participants | 108 |
Not at all |
0.9
0.7%
|
A little |
3.7
3%
|
Somewhat |
13.0
10.5%
|
Very |
20.4
16.5%
|
Extremely |
62.0
50%
|
Title | Assess the Ease of Training for Titan® OTR - Question 8 |
---|---|
Description | How easy was it for the subject to learn? |
Time Frame | 6 weeks |
Outcome Measure Data
Analysis Population Description |
---|
[Not Specified] |
Arm/Group Title | Titan® OTR IPP |
---|---|
Arm/Group Description | Subjects implanted with Titan® One Touch Release (OTR) Inflatable Penile Prosthesis (IPP) Inflatable Penile Prosthesis : Hydraulic system designed to be surgically implanted into the penis for the management of erectile dysfunction. |
Measure Participants | 107 |
Not at all |
2.8
2.3%
|
A little |
8.4
6.8%
|
Somewhat |
7.5
6%
|
Very |
29.9
24.1%
|
Extremely |
51.4
41.5%
|
Title | Assess the Ease of Training for Titan® OTR - Question 9 |
---|---|
Description | How likely the subject will need continued training or retraining? |
Time Frame | 6 weeks |
Outcome Measure Data
Analysis Population Description |
---|
[Not Specified] |
Arm/Group Title | Titan® OTR IPP |
---|---|
Arm/Group Description | Subjects implanted with Titan® One Touch Release (OTR) Inflatable Penile Prosthesis (IPP) Inflatable Penile Prosthesis : Hydraulic system designed to be surgically implanted into the penis for the management of erectile dysfunction. |
Measure Participants | 107 |
Not at all |
5.6
4.5%
|
A little |
10.3
8.3%
|
Somewhat |
11.2
9%
|
Very |
25.2
20.3%
|
Extremely |
47.7
38.5%
|
Title | Assess Partner Satisfaction (Where Applicable) |
---|---|
Description | The Partner Treatment Satisfaction Scale (TSS) was used. The TSS provides scores ranging from 0 to 100 in 5 domains of Ease of erection, Erectile function, Pleasure from sexual activity, Satisfaction with orgasm and Confidence to complete sexual activity with higher scores indicative of worse symptoms. Available data is summarized for each domain at each visit along with change from baseline. |
Time Frame | Baseline |
Outcome Measure Data
Analysis Population Description |
---|
[Not Specified] |
Arm/Group Title | Titan® OTR IPP |
---|---|
Arm/Group Description | Subjects implanted with Titan® One Touch Release (OTR) Inflatable Penile Prosthesis (IPP) Inflatable Penile Prosthesis : Hydraulic system designed to be surgically implanted into the penis for the management of erectile dysfunction. |
Measure Participants | 59 |
Ease of erection |
22.9
(28.3)
|
Erectile function |
18.1
(23.9)
|
Pleasure from sexual activity |
35.6
(29.8)
|
Satisfaction with orgasm |
35.2
(33.2)
|
Confidence to complete sexual activity |
23.2
(27.0)
|
Title | Assess Partner Satisfaction (Where Applicable) |
---|---|
Description | The Partner Treatment Satisfaction Scale (TSS) was used. The TSS provides scores ranging from 0 to 100 in 5 domains of Ease of erection, Erectile function, Pleasure from sexual activity, Satisfaction with orgasm and Confidence to complete sexual activity with higher scores indicative of worse symptoms. Available data is summarized for each domain at each visit along with change from baseline. |
Time Frame | 6 months |
Outcome Measure Data
Analysis Population Description |
---|
[Not Specified] |
Arm/Group Title | Titan® OTR IPP |
---|---|
Arm/Group Description | Subjects implanted with Titan® One Touch Release (OTR) Inflatable Penile Prosthesis (IPP) Inflatable Penile Prosthesis : Hydraulic system designed to be surgically implanted into the penis for the management of erectile dysfunction. |
Measure Participants | 39 |
Ease of erection |
67.8
(31.3)
|
Erectile function |
74.8
(20.4)
|
Pleasure from sexual activity |
64.7
(24.8)
|
Satisfaction with orgasm |
62.2
(29.2)
|
Confidence to complete sexual activity |
71.2
(28.4)
|
Adverse Events
Time Frame | 12 months | |
---|---|---|
Adverse Event Reporting Description | The analysis is based on the per protocol population (n=113). Of 124 consented subjects, nine (9) subjects did not meet inclusion/exclusion criteria and were not implanted with the study device. Two (2) subjects did not meet enrollment criteria and these two subjects were not followed for safety or efficacy. | |
Arm/Group Title | Titan® OTR IPP | |
Arm/Group Description | Subjects implanted with Titan® One Touch Release (OTR) Inflatable Penile Prosthesis (IPP) Inflatable Penile Prosthesis : Hydraulic system designed to be surgically implanted into the penis for the management of erectile dysfunction. | |
All Cause Mortality |
||
Titan® OTR IPP | ||
Affected / at Risk (%) | # Events | |
Total | / (NaN) | |
Serious Adverse Events |
||
Titan® OTR IPP | ||
Affected / at Risk (%) | # Events | |
Total | 10/113 (8.8%) | |
Surgical and medical procedures | ||
Infection | 4/113 (3.5%) | 4 |
Hematoma | 2/113 (1.8%) | 2 |
Device malfunction | 2/113 (1.8%) | 2 |
Chronic pain | 1/113 (0.9%) | 1 |
Cylinder extrusion | 1/113 (0.9%) | 1 |
Scrotal abscess | 1/113 (0.9%) | 1 |
Other (Not Including Serious) Adverse Events |
||
Titan® OTR IPP | ||
Affected / at Risk (%) | # Events | |
Total | 14/113 (12.4%) | |
Surgical and medical procedures | ||
Auto inflation | 14/113 (12.4%) | 14 |
Limitations/Caveats
More Information
Certain Agreements
Principal Investigators are NOT employed by the organization sponsoring the study.
There is NOT an agreement between Principal Investigators and the Sponsor (or its agents) that restricts the PI's rights to discuss or publish trial results after the trial is completed.
Results Point of Contact
Name/Title | Clinical Trial Manager |
---|---|
Organization | Coloplast Corp |
Phone | 612-302-4990 |
usdibe@coloplast.com |
- DK057CC