ATD: Study to Evaluate a Prototype Non-Invasive BG Measurement System

Sponsor
Afon Technology (Industry)
Overall Status
Not yet recruiting
CT.gov ID
NCT05023798
Collaborator
(none)
50
1
11.5
4.4

Study Details

Study Description

Brief Summary

  1. Diabetes mellitus is a chronic disease currently affecting more than 425 million people, of which one-third are people older than 65 years. In the UK, the number of people currently diagnosed with diabetes surpassed 3.8 million in 2019, with someone being diagnosed with the illness every two minutes (Diabetes.org figures). The prototype device being tested is a non-invasive blood glucose measurement system worn on the wrist. This would help people with diabetes manage their condition better and help prevent complications.

  2. The main objectives of the research are:

  3. To determine how accurate and effective the Afon prototype non-invasive blood glucose measurement system is, as compared to a gold standard invasive method.

  4. To chart the Afon device's predicted blood glucose levels over time.

  5. The study will be done with 30-50 patients. Eligible patients will have been diagnosed with diabetes (type 1 or 2) at least one year prior, be between 18 and 80 years old, and with a BMI between 18-35 kg/m2. For details of the full list of inclusion and exclusion criteria, see accompanying documentation.

  6. The trial will be conducted at the Joint Clinical Research Facility (JCRF), Institute of Life Science 2, Swansea University, SA2 8PP.

  7. Participants will attend the site for a total of 5 visits, one for screening, and four study visits, no more than 7 days apart. The study will run for one year.

Condition or Disease Intervention/Treatment Phase
  • Device: ATD (device test)

Detailed Description

Diabetes is a chronic disease currently affecting more than 425 million people worldwide. In the UK, the number of people currently diagnosed with diabetes surpassed 3.8 million in 2019, with someone being diagnosed with the illness every two minutes (Diabetes.org figures).By the end of this year, 4 million deaths will happen worldwide as a result of diabetes and its complications. These complications include renal failure, blindness, and amputation. It is well established clinically that good glycaemic control minimises the risk of long-term complications and improves morbidity and mortality. Current problems with frequent blood glucose testing are mainly due to patient intolerability. Many patients find finger prick testing for blood glucose levels to be painful, dislike using sharp objects and seeing their own blood. There is also a risk of infection, and in the long term, this practice can result in damage to the finger tissue. This is because the fingers have a high concentration of sensory nerve endings. Additionally, the current invasive blood glucose monitoring systems suffer limitations in terms of the requirement for continuous calibration and susceptibility to contamination by the growth of various living organisms.

The Afon technology device would be a gamechanger in terms of improving the quality of life of millions of patients and improve their chances to manage the condition without severe complications.

Study Design

Study Type:
Observational
Anticipated Enrollment :
50 participants
Observational Model:
Cohort
Time Perspective:
Other
Official Title:
Development Study to Undertake a Comprehensive Performance Evaluation of Prototype Non-Invasive Blood Glucose Measurement Systems Compared to a Standard (Invasive) Measurement Method of Blood Glucose
Anticipated Study Start Date :
Sep 15, 2021
Anticipated Primary Completion Date :
Aug 30, 2022
Anticipated Study Completion Date :
Aug 30, 2022

Arms and Interventions

Arm Intervention/Treatment
ATD Cohort 1

The first iteration of the sensor will be tested and results compared to the gold-standard laboratory measuring equipment.

Device: ATD (device test)
Medical device test (no intervention)

ATD Cohort 2

The next iteration of the sensor will be tested and results compared to the gold-standard laboratory measuring equipment. The prototype will be tested on a cohort of 3-5 patients.

Device: ATD (device test)
Medical device test (no intervention)

ATD Cohort 3

The next iteration of the sensor will be tested and results compared to the gold-standard laboratory measuring equipment. The prototype will be tested on a cohort of 3-5 patients.

Device: ATD (device test)
Medical device test (no intervention)

ATD Cohort 4

The next iteration of the sensor will be tested and results compared to the gold-standard laboratory measuring equipment. The prototype will be tested on a cohort of 3-5 patients.

Device: ATD (device test)
Medical device test (no intervention)

ATD Cohort 5

The next iteration of the sensor will be tested and results compared to the gold-standard laboratory measuring equipment. The prototype will be tested on a cohort of 3-5 patients.

Device: ATD (device test)
Medical device test (no intervention)

ATD Cohort 6

The next iteration of the sensor will be tested and results compared to the gold-standard laboratory measuring equipment. The prototype will be tested on a cohort of 3-5 patients.

Device: ATD (device test)
Medical device test (no intervention)

ATD Cohort 7

The next iteration of the sensor will be tested and results compared to the gold-standard laboratory measuring equipment. The prototype will be tested on a cohort of 3-5 patients.

Device: ATD (device test)
Medical device test (no intervention)

ATD Cohort 8

The next iteration of the sensor will be tested and results compared to the gold-standard laboratory measuring equipment. The prototype will be tested on a cohort of 3-5 patients.

Device: ATD (device test)
Medical device test (no intervention)

ATD Cohort 9

The next iteration of the sensor will be tested and results compared to the gold-standard laboratory measuring equipment. The prototype will be tested on a cohort of 3-5 patients.

Device: ATD (device test)
Medical device test (no intervention)

ATD Cohort 10

The next iteration of the sensor will be tested and results compared to the gold-standard laboratory measuring equipment. The prototype will be tested on a cohort of 3-5 patients.

Device: ATD (device test)
Medical device test (no intervention)

Outcome Measures

Primary Outcome Measures

  1. Exploratory Efficacy Objective 1 [Up to 4 hours per test session.]

    To estimate real-time serial blood glucose levels from the device measurements using a gold standard methodology for blood glucose measurement as control. A comparison addressing the potential differences of MARDs and PARDs respectively will be done for the Afon system tested and assessed using mixed effect linear models with the BG as the fixed effect and subjects as random effects. Within the model the least square means, the differences between least square means of MARD of the Afon system and corresponding two-sided 95% parametric confidence interval will be calculated.

  2. Exploratory Efficacy Objective 2 [Up to 4 hours per test session.]

    To chart patient-specific predicted vs actual blood glucose levels over time, including before and after glucose challenge. The type I error is set at 5% two-sided.

  3. Safety Objective 1 [Up to 4 hours per test session.]

    To examine the safety and tolerability of the prototype sensors. Number of participants with no device-related SAE = 0.

  4. Exploratory Usability Objective [25 minutes - 1 hour during a test session.]

    To obtain patient input for wearable device design by means of a usability questionnaire1 answered questionnaire per participant.

Eligibility Criteria

Criteria

Ages Eligible for Study:
18 Years to 64 Years
Sexes Eligible for Study:
All
Accepts Healthy Volunteers:
No
Inclusion Criteria:

Patient who has a documented Type 1 diagnosis (diagnosed under age 29), or Type 2 diabetes for more than one year (with negative GAD antibody test results) and who are willing and able to give informed consent for participation in the study Male or female patient Patient aged 18 to 80 years Patients with a BMI 18-35 kg/m2.

Exclusion Criteria:
The patient may not enter the study if ANY of the following apply:

Type 1 patients diagnosed after age 29 Type 2 patients with positive GAD antibody test results Patient with other active implantable medical devices, such as pacemakers Patient who is currently participating in another clinical trial for a pharmaceutical product Patient with a history of allergy to any materials used in the prototype sensors or materials used to stabilise the devices or cabling including Rocktape, double-sided tape and prosthetic adhesive Female patient who is pregnant or lactating

  • Patient with any injury or infection of the wrist

  • Patient who is unwilling or unable to attend each visit fasting from midnight (water is permitted freely) or fasting for a minimum of 3 hours (for an afternoon session)

  • Patient who takes paracetamol or aspirin in the 24 hours preceding each visit

  • Patient with clinically significant abnormal values in clinical chemistry, as judged by the PI

  • Concurrent illness or a condition that may interfere with blood glucose levels (e. g. carcinoma, haematological disease, vasculitis, connective tissue disease, alcohol neuropathy)

  • Patient who is incapable of giving informed legal consent

  • Recent (1 month) episode of DKA, HONK or severe hypoglycaemia

  • Patient on pramlintide (Symlin/Tripro-Amylin)

  • Patient who attends the visit under the influence of alcohol or recreational drugs

  • Patient with severe macrovascular disease - stroke, CVD, CKD stage IV -V Patient who fails screening for alcohol and recreational drugs use.

Contacts and Locations

Locations

Site City State Country Postal Code
1 JCRF Swansea United Kingdom SA2 8PP

Sponsors and Collaborators

  • Afon Technology

Investigators

None specified.

Study Documents (Full-Text)

None provided.

More Information

Publications

None provided.
Responsible Party:
Afon Technology
ClinicalTrials.gov Identifier:
NCT05023798
Other Study ID Numbers:
  • 04-001
First Posted:
Aug 27, 2021
Last Update Posted:
Aug 27, 2021
Last Verified:
Aug 1, 2021
Individual Participant Data (IPD) Sharing Statement:
No
Plan to Share IPD:
No
Studies a U.S. FDA-regulated Drug Product:
No
Studies a U.S. FDA-regulated Device Product:
No