A Study to Explore the Utility of Fractional Exhaled Nitric Oxide (FeNO) Compared With Methacholine Challenge (MCC) Testing in the Assessment of Patients With Suspected But Undiagnosed Asthma
Study Details
Study Description
Brief Summary
Objective: To explore the utility of Fractional Exhaled Nitric Oxide (FeNO) compared with Methacholine Challenge (MCC) testing in assessing patients with suspected but undiagnosed asthma
Number of participants: Approximately 50 subjects will be enrolled
Reference product: NIOX MINO® Instrument (09-1100)
Performance assessments: FeNO measurements will be performed according to the "Perform FeNO Measurement" guidelines on page 7 of the NIOX MINO® User Manual. MCC testing will be performed according to the ATS guidelines and the allergy and asthma specialists procedure for conducting MCC tests
Safety assessments: The Investigator is responsible for the detection, reporting, and documentation of events meeting the definition of an Adverse Event (AE) and/or Serious Injuries as provided in this clinical investigation plan from the time that informed consent has been provided and during the study period
Criteria for evaluations: This is an exploratory study and there are currently no plans for a formal statistical analysis. Information gained from this study may used to design and power subsequent studies in patients with suspected but undiagnosed asthma. Information collected will be summarized in a clinical study report
Condition or Disease | Intervention/Treatment | Phase |
---|---|---|
|
Detailed Description
INTRODUCTION
Overview:
The measurement of exhaled nitric oxide (FeNO) is the only clinical test for measuring airway inflammation that can be performed consistently and accurately in clinical practice at the point-of-care. Airway inflammation is now recognized as the central mechanism in the pathogenesis of asthma. The measurement of FeNO with the NIOX MINO® device provides a rapid, noninvasive, and inexpensive tool to assess airway inflammation in inflammatory airway diseases such as asthma. The test is easy to perform and requires minimal training for the operator to conduct the test.
Role of Exhaled Nitric Oxide (FeNO):
FeNO has evolved as a predictive and prognostic biomarker for airway inflammation. Nitric oxide (NO) gas is produced in the epithelial cells of the bronchial wall as an intrinsic part of the inflammatory process.
Measuring the amount of FeNO is useful in the initial assessment of patients with chronic cough or non-specific respiratory symptoms suggestive of asthma, and for the management of patients with established asthma who are receiving corticosteroid treatment.
Intended Use:
NIOX MINO® measures Nitric Oxide (NO) in human breath. Nitric Oxide is frequently increased in inflammatory processes such as asthma. The fractional NO concentration in expired breath (FeNO) can be measured by NIOX MINO® with assurance that such measurements are repeatable and according to guidelines for NO measurement established by the American Thoracic Society.
Rationale for Study:
The burden of asthma in the U.S. is significant and growing. In addition, it has been established that asthma is both over- and under- diagnosed. Equally concerning is the fact that nearly 75% of these misdiagnosed patients are receiving anti-asthma treatments. Data such as these point out the need for a simple, inexpensive and reliable way to establish the diagnosis of asthma.
Presently, the assessment of patients with suspected but undiagnosed asthma is to perform Methacholine Challenge (MCC) testing to evaluate airway hyperresponsiveness. However, MCC, testing is difficult, time-consuming, and costly and also involves some level of risk to the patient. Nonetheless, there is some evidence that FeNO testing may be able to identify the patients who actually have asthma among those with suspected but undiagnosed asthma who have normal pulmonary function tests, thus potentially eliminating the need for MCC testing in many patients.
CLINICAL INVESTIGATION OBJECTIVES To explore the utility of FeNO compared with MCC in assessing patients with suspected but undiagnosed asthma.
CLINICAL INVESTIGATION PLAN
This is an exploratory, multi-center, single-visit, outpatient study. Approximately 50 subjects will participate in the study during an (approximately 8-16 week study enrollment period:
-
Demographics
-
Physical Characteristics
-
Smoking history (cigarettes only)
-
FeNO
-
Spirometry
-
Asthma and Allergy Specialists MCC Test Procedure
Subject discharge from the study:
Once all information has been collected and all procedures have been performed, the subject will be discharged from the clinic and their study participation will be complete
Medical device:
The NIOX MINO® was initially cleared by the FDA on March 4, 2008 as a new hand-held device for the measurement of exhaled Nitric Oxide, a marker of eosinophilic airway inflammation. The most recent clearance by the FDA was on September 2, 2010 for Instrument (09-1100), which will be used in this study. NIOX MINO® can be used in children 7-17 years of age and in adults 18 years of age or older in the initial assessment and management of asthma.
The NIOX MINO® is a 10 second test based on exhaled breath measured at a 50 ml/second flow rate. Results are provided at the point of care within 2 minutes after the successful completion of the breath test. The test cannot be influenced by patient effort or variations in the clinician's test technique.
Definitions:
Adverse event: Any incident where the use of a medical device (including in vitro diagnostics) is suspected to have resulted in an adverse outcome in a patient.
Serious Injury means injury or illness that:
-
Is life-threatening
-
Results in permanent impairment of a body function or permanent damage to a body structure, or
-
Necessitates medical or surgical intervention to preclude permanent impairment of a body function or permanent damage to a body structure.
Malfunction: the failure of a device to meet its performance specifications or otherwise perform as intended.
Caused or contributed: the death or serous injury was or may have been attributed to a medical device, or that the medical device was or may have been a factor in a death or serious injury, including events occurring as a result of:
-
Failure
-
Malfunction
-
Improper or inadequate design
-
Manufacture
-
Labeling
-
User error
Study Design
Arms and Interventions
Arm | Intervention/Treatment |
---|---|
FeNO Participants with suspected but undiagnosed asthma will have a fractional exhaled nitric oxide (FeNO) measurement taken by NIOX MINO® Instrument (09-1100) according to the 'Perform FeNO Measurement' guidelines on page 7 of the NIOX MINO® User Manual (February 2011) during their study visit. Following that they will have a Methacholine Challenge (MCC) test performed according to the ATS guidelines and the Allergy and Asthma Specialists Procedure for conducting MCC tests |
Device: NIOX MINO® Instrument (09-1100)
The NIOX MINO® Instrument (09-1100) is a 10 second test based on exhaled breath measured at a 50 ml/second flow rate.
|
Outcome Measures
Primary Outcome Measures
- Methacholine Challenge (MCC) Results [approximately 1-hour]
Fractional Exhaled Nitric Oxide (FeNO) measurements will be performed according to the "Perform FeNO Measurement" guidelines on page 7 of the NIOX MINO® User Manual (February, 2011). Methacholine Challenge (MCC) tests will be performed according to the ATS guidelines and the Allergy and Asthma Specialists Procedure for conducting MCC tests. Values from the two tests will be compared against final asthma diagnosis.
- Mean FeNO Levels by Methacholine Challenge (MCC) Results: MCC Results [approximately 1-hour]
Fractional Exhaled Nitric Oxide (FeNO) measurements will be performed according to the "Perform FeNO Measurement" guidelines on page 7 of the NIOX MINO® User Manual (February, 2011). Methacholine Challenge (MCC) tests will be performed according to the ATS guidelines and the Allergy and Asthma Specialists Procedure for conducting MCC tests. Values from the two tests will be compared against final asthma diagnosis.
- FeNO by Methacholine Challenge (MCC) Results: FeNO <25ppb [approximately 1-hour]
Fractional Exhaled Nitric Oxide (FeNO) measurements will be performed according to the "Perform FeNO Measurement" guidelines on page 7 of the NIOX MINO® User Manual (February, 2011). Methacholine Challenge (MCC) tests will be performed according to the ATS guidelines and the Allergy and Asthma Specialists Procedure for conducting MCC tests. Values from the two tests will be compared against final asthma diagnosis. A positive response to methacholine challenge is defined as a greater than or equal to 20% decrease in FEV1 at a methacholine dose less than or equal to 16mg/ml from baseline in response to methacholine. A negative response to methacholine challenge is defined as a less than 20% decrease in FEV1 at a methacholine dose less than or equal to 16mg/ml from baseline.
- FeNO by Methacholine Challenge (MCC) Results: FeNO >=25ppb to <=50ppb [approximately 1-hour]
Fractional Exhaled Nitric Oxide (FeNO) measurements will be performed according to the "Perform FeNO Measurement" guidelines on page 7 of the NIOX MINO® User Manual (February, 2011). Methacholine Challenge (MCC) tests will be performed according to the ATS guidelines and the Allergy and Asthma Specialists Procedure for conducting MCC tests. Values from the two tests will be compared against final asthma diagnosis. A positive response to methacholine challenge is defined as a greater than or equal to 20% decrease in FEV1 at a methacholine dose less than or equal to 16mg/ml from baseline in response to methacholine. A negative response to methacholine challenge is defined as a less than 20% decrease in FEV1 at a methacholine dose less than or equal to 16mg/ml from baseline.
- FeNO by Methacholine Challenge (MCC) Results: FeNO >50ppb [approximately 1-hour]
Fractional Exhaled Nitric Oxide (FeNO) measurements will be performed according to the "Perform FeNO Measurement" guidelines on page 7 of the NIOX MINO® User Manual (February, 2011). Methacholine Challenge (MCC) tests will be performed according to the ATS guidelines and the Allergy and Asthma Specialists Procedure for conducting MCC tests. Values from the two tests will be compared against final asthma diagnosis. A positive response to methacholine challenge is defined as a greater than or equal to 20% decrease in FEV1 at a methacholine dose less than or equal to 16mg/ml from baseline in response to methacholine. A negative response to methacholine challenge is defined as a less than 20% decrease in FEV1 at a methacholine dose less than or equal to 16mg/ml from baseline.
- Asthma Diagnosis by MCC Results: Positive MCC Response [approximately 1-hour]
Fractional Exhaled Nitric Oxide (FeNO) measurements was performed according to the "Perform FeNO Measurement" guidelines on page 7 of the NIOX MINO® User Manual (February, 2011). Methacholine Challenge (MCC) tests were performed according to the ATS guidelines and the Allergy and Asthma Specialists Procedure for conducting MCC tests. Values from the two tests will be compared against final asthma diagnosis. A positive response to methacholine challenge is defined as a greater than or equal to 20% decrease in FEV1 at a methacholine dose less than or equal to 16mg/ml from baseline in response to methacholine.
- Asthma Diagnosis by MCC Results: Negative MCC Response [approximately 1-hour]
Fractional Exhaled Nitric Oxide (FeNO) measurements was performed according to the "Perform FeNO Measurement" guidelines on page 7 of the NIOX MINO® User Manual (February, 2011). Methacholine Challenge (MCC) tests were performed according to the ATS guidelines and the Allergy and Asthma Specialists Procedure for conducting MCC tests. Values from the two tests will be compared against final asthma diagnosis. A negative response to methacholine challenge is defined as a less than 20% decrease in FEV1 at a methacholine dose less than or equal to 16mg/ml from baseline.
- Asthma Diagnosis by FeNO: FeNO <25ppb [approximately 1-hour]
Fractional Exhaled Nitric Oxide (FeNO) measurements were performed according to the "Perform FeNO Measurement" guidelines on page 7 of the NIOX MINO® User Manual (February, 2011). Methacholine Challenge (MCC) tests were performed according to the ATS guidelines and the Allergy and Asthma Specialists Procedure for conducting MCC tests. Values from the two tests will be compared against final asthma diagnosis.
- Asthma Diagnosis by FeNO: FeNO >=25ppb to <= 50ppb [approximately 1-hour]
Fractional Exhaled Nitric Oxide (FeNO) measurements were performed according to the "Perform FeNO Measurement" guidelines on page 7 of the NIOX MINO® User Manual (February, 2011). Methacholine Challenge (MCC) tests were performed according to the ATS guidelines and the Allergy and Asthma Specialists Procedure for conducting MCC tests. Values from the two tests will be compared against final asthma diagnosis.
- Asthma Diagnosis by FeNO: FeNO >50ppb [approximately 1-hour]
Fractional Exhaled Nitric Oxide (FeNO) measurements were performed according to the "Perform FeNO Measurement" guidelines on page 7 of the NIOX MINO® User Manual (February, 2011). Methacholine Challenge (MCC) tests were performed according to the ATS guidelines and the Allergy and Asthma Specialists Procedure for conducting MCC tests. Values from the two tests will be compared against final asthma diagnosis.
- Asthma Diagnosis by FeNO: Mean FeNO Value [approximately 1-hour]
Fractional Exhaled Nitric Oxide (FeNO) measurements were performed according to the "Perform FeNO Measurement" guidelines on page 7 of the NIOX MINO® User Manual (February, 2011). Methacholine Challenge (MCC) tests were performed according to the ATS guidelines and the Allergy and Asthma Specialists Procedure for conducting MCC tests. Values from the two tests will be compared against final asthma diagnosis.
- Sensitivity (%) for FeNO [approximately 1-hour]
Fractional Exhaled Nitric Oxide (FeNO) measurements were performed according to the "Perform FeNO Measurement" guidelines on page 7 of the NIOX MINO® User Manual (February, 2011). Methacholine Challenge (MCC) tests were performed according to the ATS guidelines and the Allergy and Asthma Specialists Procedure for conducting MCC tests. Values from the two tests will be compared against final asthma diagnosis. Sensitivity is a way of saying how likely a test is positive if you have a disease and is expressed as (%) percent. For this study, sensitivity was the ability of FeNO to correctly identify a study participant who has asthma.
- Specificity (%) for FeNO [approximately 1-hour]
Fractional Exhaled Nitric Oxide (FeNO) measurements were performed according to the "Perform FeNO Measurement" guidelines on page 7 of the NIOX MINO® User Manual (February, 2011). Methacholine Challenge (MCC) tests were performed according to the ATS guidelines and the Allergy and Asthma Specialists Procedure for conducting MCC tests. Values from the two tests will be compared against final asthma diagnosis. Specificity is a way of saying how likely a test is negative if you do not have a disease and is expressed as (%) percent. For this study, specificity was the ability of FeNO to correctly identify a study participant who does not have asthma.
- Positive Predictive Value (%) for FeNO [approximately 1-hour]
Fractional Exhaled Nitric Oxide (FeNO) measurements were performed according to the "Perform FeNO Measurement" guidelines on page 7 of the NIOX MINO® User Manual (February, 2011). Methacholine Challenge (MCC) tests were performed according to the ATS guidelines and the Allergy and Asthma Specialists Procedure for conducting MCC tests. Values from the two tests will be compared against final asthma diagnosis. Positive predictive value is a way of saying the likelihood a positive test means you have a disease. In this study, it is defined as the percentage of study participants with a high FeNO who have asthma.
- Negative Predictive Value (%) for FeNO [approximately 1-hour]
Fractional Exhaled Nitric Oxide (FeNO) measurements were performed according to the "Perform FeNO Measurement" guidelines on page 7 of the NIOX MINO® User Manual (February, 2011). Methacholine Challenge (MCC) tests were performed according to the ATS guidelines and the Allergy and Asthma Specialists Procedure for conducting MCC tests. Values from the two tests will be compared against final asthma diagnosis. Negative predictive value is a way of saying the likelihood of a negative test means you do not have disease. In this study it is defined as the percentage of study participants with a low FeNO who do not have asthma.
- FeNO Values by ICS Use: FeNO <25ppb [approximately 1-hour]
Fractional Exhaled Nitric Oxide (FeNO) measurements were performed according to the "Perform FeNO Measurement" guidelines on page 7 of the NIOX MINO® User Manual (February, 2011). Methacholine Challenge (MCC) tests were performed according to the ATS guidelines and the Allergy and Asthma Specialists Procedure for conducting MCC tests. Values from the two tests were compared against final asthma diagnosis. Data are presented for participants with FeNO <25ppb who either did not use Inhaled Corticosteroids (ICS) or did use Inhaled Corticosteroids.
- FeNO Values by ICS Use: FeNO >= 50ppb [approximately 1-hour]
Fractional Exhaled Nitric Oxide (FeNO) measurements were performed according to the "Perform FeNO Measurement" guidelines on page 7 of the NIOX MINO® User Manual (February, 2011). Methacholine Challenge (MCC) tests were performed according to the ATS guidelines and the Allergy and Asthma Specialists Procedure for conducting MCC tests. Values from the two tests were compared against final asthma diagnosis. Data are presented for participants with FeNO >=50ppb who either did not use Inhaled Corticosteroids (ICS) or did use Inhaled Corticosteroids.
- FeNO Values by ICS Use: FeNO >= 25ppb to <= 50ppb [approximately 1-hour]
Fractional Exhaled Nitric Oxide (FeNO) measurements were performed according to the "Perform FeNO Measurement" guidelines on page 7 of the NIOX MINO® User Manual (February, 2011). Methacholine Challenge (MCC) tests were performed according to the ATS guidelines and the Allergy and Asthma Specialists Procedure for conducting MCC tests. Values from the two tests were compared against final asthma diagnosis. Data are presented for participants with FeNO <=25ppb to <=50ppb who either did not use Inhaled Corticosteroids (ICS) or did use Inhaled Corticosteroids.
- FeNO Values by Smoking Status: FeNO <25ppb [approximately 1-hour]
Fractional Exhaled Nitric Oxide (FeNO) measurements were performed according to the "Perform FeNO Measurement" guidelines on page 7 of the NIOX MINO® User Manual (February, 2011). Methacholine Challenge (MCC) tests were performed according to the ATS guidelines and the Allergy and Asthma Specialists Procedure for conducting MCC tests. Values from the two tests were compared against final asthma diagnosis. Data are presented for participants with FeNO <25ppb who either did not smoke previously or are current smokers.
- FeNO Values by Smoking Status: FeNO >= 25ppb to <= 50ppb [approximately 1-hour]
Fractional Exhaled Nitric Oxide (FeNO) measurements were performed according to the "Perform FeNO Measurement" guidelines on page 7 of the NIOX MINO® User Manual (February, 2011). Methacholine Challenge (MCC) tests were performed according to the ATS guidelines and the Allergy and Asthma Specialists Procedure for conducting MCC tests. Values from the two tests were compared against final asthma diagnosis. Data are presented for participants with FeNO >=25ppb to <=50ppb who either did not smoke previously or are current smokers.
- FeNO Values by Smoking Status: FeNO >= 50ppb [approximately 1-hour]
Fractional Exhaled Nitric Oxide (FeNO) measurements were performed according to the "Perform FeNO Measurement" guidelines on page 7 of the NIOX MINO® User Manual (February, 2011). Methacholine Challenge (MCC) tests were performed according to the ATS guidelines and the Allergy and Asthma Specialists Procedure for conducting MCC tests. Values from the two tests were compared against final asthma diagnosis. Data are presented for participants with FeNO >=50ppb who either did not smoke previously or are current smokers.
Eligibility Criteria
Criteria
Inclusion Criteria:
- Subjects with suspected but undiagnosed asthma that are scheduled to undergo MCC testing as part of the routine evaluation of suspected asthma.
Exclusion Criteria:
-
Concurrent Conditions or Disease:
-
current diagnosis of asthma, COPD, or other chronic respiratory disease
-
upper or lower respiratory infection (including the common cold) that has not resolved within 4 weeks prior to the MCC test
Medication Use:
-
scheduled use of asthma and allergy medication(s) should be discontinued at least 48 hours prior to MCC testing
-
rescue medications should be discontinued at least 8 hour prior to MCC testing
Smoking:
- smoking should be avoided for at least 8 hours prior to MCC testing
Exercise:
- exercise or outdoor activity should be avoided for at least 2 hours prior to testing
Vaccinations:
-vaccinations should be avoided at least 2 weeks prior to MCC testing
Study participation outside of the protocol:
-subjects currently enrolled in studies of investigational or non-investigational drugs or medical devices and/or who participated in these studies within 30 days prior to this study are excluded
Contacts and Locations
Locations
Site | City | State | Country | Postal Code | |
---|---|---|---|---|---|
1 | Allergy and Asthma Specialists, PC | Blue Bell | Pennsylvania | United States | 19422 |
2 | Allergy and Asthma Specialists, PC | Collegeville | Pennsylvania | United States | 19426 |
3 | Allergy and Asthma Specialists, PC | Doylestown | Pennsylvania | United States | 18901 |
4 | Allergy and Asthma Specialists, PC | Jenkintown | Pennsylvania | United States | 19102 |
5 | Allergy and Asthma Specialists, PC | King of Prussia | Pennsylvania | United States | 19406 |
6 | Allergy and Asthma Specialists, PC | Lansdale | Pennsylvania | United States | 19446 |
7 | Allergy and Asthma Specialists, PC | Philadelphia | Pennsylvania | United States | 19102 |
8 | Allergy and Asthma Specialists, PC | Pottsville | Pennsylvania | United States | 19464 |
Sponsors and Collaborators
- Aerocrine AB
Investigators
- Principal Investigator: Robert Anolik, MD, Allergy and Asthma Specialists, PC
Study Documents (Full-Text)
None provided.More Information
Publications
None provided.- AER-044
Study Results
Participant Flow
Recruitment Details | |
---|---|
Pre-assignment Detail |
Arm/Group Title | FeNO |
---|---|
Arm/Group Description | Participants with suspected but undiagnosed asthma had a fractional exhaled nitric oxide (FeNO) measurement taken by NIOX MINO® Instrument (09-1100) according to the 'Perform FeNO Measurement' guidelines on page 7 of the NIOX MINO® User Manual (February 2011) during their study visit. Following that they had a Methacholine Challenge (MCC) test performed according to the ATS guidelines and the Allergy and Asthma Specialists Procedure for conducting MCC tests NIOX MINO® Instrument (09-1100): The NIOX MINO® Instrument (09-1100) is a 10 second test based on exhaled breath measured at a 50 ml/second flow rate. |
Period Title: Overall Study | |
STARTED | 50 |
COMPLETED | 50 |
NOT COMPLETED | 0 |
Baseline Characteristics
Arm/Group Title | FeNO |
---|---|
Arm/Group Description | Participants with suspected but undiagnosed asthma had a fractional exhaled nitric oxide (FeNO) measurement taken by NIOX MINO® Instrument (09-1100) according to the 'Perform FeNO Measurement' guidelines on page 7 of the NIOX MINO® User Manual (February 2011) during their study visit. Following that they had a Methacholine Challenge (MCC) test performed according to the ATS guidelines and the Allergy and Asthma Specialists Procedure for conducting MCC tests NIOX MINO® Instrument (09-1100): The NIOX MINO® Instrument (09-1100) is a 10 second test based on exhaled breath measured at a 50 ml/second flow rate. |
Overall Participants | 50 |
Age (years) [Mean (Full Range) ] | |
Mean (Full Range) [years] |
24.7
|
Sex: Female, Male (Count of Participants) | |
Female |
23
46%
|
Male |
27
54%
|
Region of Enrollment (participants) [Number] | |
United States |
50
100%
|
Outcome Measures
Title | Methacholine Challenge (MCC) Results |
---|---|
Description | Fractional Exhaled Nitric Oxide (FeNO) measurements will be performed according to the "Perform FeNO Measurement" guidelines on page 7 of the NIOX MINO® User Manual (February, 2011). Methacholine Challenge (MCC) tests will be performed according to the ATS guidelines and the Allergy and Asthma Specialists Procedure for conducting MCC tests. Values from the two tests will be compared against final asthma diagnosis. |
Time Frame | approximately 1-hour |
Outcome Measure Data
Analysis Population Description |
---|
[Not Specified] |
Arm/Group Title | FeNO |
---|---|
Arm/Group Description | Participants with suspected but undiagnosed asthma will have a fractional exhaled nitric oxide (FeNO) measurement taken by NIOX MINO® Instrument (09-1100) according to the 'Perform FeNO Measurement' guidelines on page 7 of the NIOX MINO® User Manual (February 2011) during their study visit. Following that they will have a Methacholine Challenge (MCC) test performed according to the ATS guidelines and the Allergy and Asthma Specialists Procedure for conducting MCC tests NIOX MINO® Instrument (09-1100): The NIOX MINO® Instrument (09-1100) is a 10 second test based on exhaled breath measured at a 50 ml/second flow rate. |
Measure Participants | 50 |
>= 20% decrease in FEV1 from BL MCC Dilution |
14
28%
|
<20% Decrease in FEV1 from BL MCC Dilution |
36
72%
|
Title | Mean FeNO Levels by Methacholine Challenge (MCC) Results: MCC Results |
---|---|
Description | Fractional Exhaled Nitric Oxide (FeNO) measurements will be performed according to the "Perform FeNO Measurement" guidelines on page 7 of the NIOX MINO® User Manual (February, 2011). Methacholine Challenge (MCC) tests will be performed according to the ATS guidelines and the Allergy and Asthma Specialists Procedure for conducting MCC tests. Values from the two tests will be compared against final asthma diagnosis. |
Time Frame | approximately 1-hour |
Outcome Measure Data
Analysis Population Description |
---|
[Not Specified] |
Arm/Group Title | FeNO |
---|---|
Arm/Group Description | Participants with suspected but undiagnosed asthma will have a fractional exhaled nitric oxide (FeNO) measurement taken by NIOX MINO® Instrument (09-1100) according to the 'Perform FeNO Measurement' guidelines on page 7 of the NIOX MINO® User Manual (February 2011) during their study visit. Following that they will have a Methacholine Challenge (MCC) test performed according to the ATS guidelines and the Allergy and Asthma Specialists Procedure for conducting MCC tests NIOX MINO® Instrument (09-1100): The NIOX MINO® Instrument (09-1100) is a 10 second test based on exhaled breath measured at a 50 ml/second flow rate. |
Measure Participants | 50 |
>= 20% decrease in FEV1 from BL MCC Dilution |
17.4
(12.54)
|
<20% Decrease in FEV1 from BL MCC Dilution |
17.7
(9.77)
|
Title | FeNO by Methacholine Challenge (MCC) Results: FeNO <25ppb |
---|---|
Description | Fractional Exhaled Nitric Oxide (FeNO) measurements will be performed according to the "Perform FeNO Measurement" guidelines on page 7 of the NIOX MINO® User Manual (February, 2011). Methacholine Challenge (MCC) tests will be performed according to the ATS guidelines and the Allergy and Asthma Specialists Procedure for conducting MCC tests. Values from the two tests will be compared against final asthma diagnosis. A positive response to methacholine challenge is defined as a greater than or equal to 20% decrease in FEV1 at a methacholine dose less than or equal to 16mg/ml from baseline in response to methacholine. A negative response to methacholine challenge is defined as a less than 20% decrease in FEV1 at a methacholine dose less than or equal to 16mg/ml from baseline. |
Time Frame | approximately 1-hour |
Outcome Measure Data
Analysis Population Description |
---|
Participants with a FeNO < 25 ppb |
Arm/Group Title | FeNO |
---|---|
Arm/Group Description | Participants with suspected but undiagnosed asthma will have a fractional exhaled nitric oxide (FeNO) measurement taken by NIOX MINO® Instrument (09-1100) according to the 'Perform FeNO Measurement' guidelines on page 7 of the NIOX MINO® User Manual (February 2011) during their study visit. Following that they will have a Methacholine Challenge (MCC) test performed according to the ATS guidelines and the Allergy and Asthma Specialists Procedure for conducting MCC tests NIOX MINO® Instrument (09-1100): The NIOX MINO® Instrument (09-1100) is a 10 second test based on exhaled breath measured at a 50 ml/second flow rate. |
Measure Participants | 38 |
Positive Response to MCC |
12
24%
|
Negative Response to MCC |
26
52%
|
Title | FeNO by Methacholine Challenge (MCC) Results: FeNO >=25ppb to <=50ppb |
---|---|
Description | Fractional Exhaled Nitric Oxide (FeNO) measurements will be performed according to the "Perform FeNO Measurement" guidelines on page 7 of the NIOX MINO® User Manual (February, 2011). Methacholine Challenge (MCC) tests will be performed according to the ATS guidelines and the Allergy and Asthma Specialists Procedure for conducting MCC tests. Values from the two tests will be compared against final asthma diagnosis. A positive response to methacholine challenge is defined as a greater than or equal to 20% decrease in FEV1 at a methacholine dose less than or equal to 16mg/ml from baseline in response to methacholine. A negative response to methacholine challenge is defined as a less than 20% decrease in FEV1 at a methacholine dose less than or equal to 16mg/ml from baseline. |
Time Frame | approximately 1-hour |
Outcome Measure Data
Analysis Population Description |
---|
Participants with FeNO >= 25 ppb to <= 50 ppb |
Arm/Group Title | FeNO |
---|---|
Arm/Group Description | Participants with suspected but undiagnosed asthma will have a fractional exhaled nitric oxide (FeNO) measurement taken by NIOX MINO® Instrument (09-1100) according to the 'Perform FeNO Measurement' guidelines on page 7 of the NIOX MINO® User Manual (February 2011) during their study visit. Following that they will have a Methacholine Challenge (MCC) test performed according to the ATS guidelines and the Allergy and Asthma Specialists Procedure for conducting MCC tests NIOX MINO® Instrument (09-1100): The NIOX MINO® Instrument (09-1100) is a 10 second test based on exhaled breath measured at a 50 ml/second flow rate. |
Measure Participants | 11 |
Positive Response to MCC |
1
2%
|
Negative Response to MCC |
10
20%
|
Title | FeNO by Methacholine Challenge (MCC) Results: FeNO >50ppb |
---|---|
Description | Fractional Exhaled Nitric Oxide (FeNO) measurements will be performed according to the "Perform FeNO Measurement" guidelines on page 7 of the NIOX MINO® User Manual (February, 2011). Methacholine Challenge (MCC) tests will be performed according to the ATS guidelines and the Allergy and Asthma Specialists Procedure for conducting MCC tests. Values from the two tests will be compared against final asthma diagnosis. A positive response to methacholine challenge is defined as a greater than or equal to 20% decrease in FEV1 at a methacholine dose less than or equal to 16mg/ml from baseline in response to methacholine. A negative response to methacholine challenge is defined as a less than 20% decrease in FEV1 at a methacholine dose less than or equal to 16mg/ml from baseline. |
Time Frame | approximately 1-hour |
Outcome Measure Data
Analysis Population Description |
---|
Participants with FeNO >50ppb |
Arm/Group Title | FeNO |
---|---|
Arm/Group Description | Participants with suspected but undiagnosed asthma will have a fractional exhaled nitric oxide (FeNO) measurement taken by NIOX MINO® Instrument (09-1100) according to the 'Perform FeNO Measurement' guidelines on page 7 of the NIOX MINO® User Manual (February 2011) during their study visit. Following that they will have a Methacholine Challenge (MCC) test performed according to the ATS guidelines and the Allergy and Asthma Specialists Procedure for conducting MCC tests NIOX MINO® Instrument (09-1100): The NIOX MINO® Instrument (09-1100) is a 10 second test based on exhaled breath measured at a 50 ml/second flow rate. |
Measure Participants | 1 |
Positive Response to MCC |
1
2%
|
Negative Response to MCC |
0
0%
|
Title | Asthma Diagnosis by MCC Results: Positive MCC Response |
---|---|
Description | Fractional Exhaled Nitric Oxide (FeNO) measurements was performed according to the "Perform FeNO Measurement" guidelines on page 7 of the NIOX MINO® User Manual (February, 2011). Methacholine Challenge (MCC) tests were performed according to the ATS guidelines and the Allergy and Asthma Specialists Procedure for conducting MCC tests. Values from the two tests will be compared against final asthma diagnosis. A positive response to methacholine challenge is defined as a greater than or equal to 20% decrease in FEV1 at a methacholine dose less than or equal to 16mg/ml from baseline in response to methacholine. |
Time Frame | approximately 1-hour |
Outcome Measure Data
Analysis Population Description |
---|
[Not Specified] |
Arm/Group Title | FeNO |
---|---|
Arm/Group Description | Participants with suspected but undiagnosed asthma will have a fractional exhaled nitric oxide (FeNO) measurement taken by NIOX MINO® Instrument (09-1100) according to the 'Perform FeNO Measurement' guidelines on page 7 of the NIOX MINO® User Manual (February 2011) during their study visit. Following that they will have a Methacholine Challenge (MCC) test performed according to the ATS guidelines and the Allergy and Asthma Specialists Procedure for conducting MCC tests NIOX MINO® Instrument (09-1100): The NIOX MINO® Instrument (09-1100) is a 10 second test based on exhaled breath measured at a 50 ml/second flow rate. |
Measure Participants | 14 |
Asthma |
11
22%
|
Non-Asthma |
3
6%
|
Title | Asthma Diagnosis by MCC Results: Negative MCC Response |
---|---|
Description | Fractional Exhaled Nitric Oxide (FeNO) measurements was performed according to the "Perform FeNO Measurement" guidelines on page 7 of the NIOX MINO® User Manual (February, 2011). Methacholine Challenge (MCC) tests were performed according to the ATS guidelines and the Allergy and Asthma Specialists Procedure for conducting MCC tests. Values from the two tests will be compared against final asthma diagnosis. A negative response to methacholine challenge is defined as a less than 20% decrease in FEV1 at a methacholine dose less than or equal to 16mg/ml from baseline. |
Time Frame | approximately 1-hour |
Outcome Measure Data
Analysis Population Description |
---|
[Not Specified] |
Arm/Group Title | FeNO |
---|---|
Arm/Group Description | Participants with suspected but undiagnosed asthma will have a fractional exhaled nitric oxide (FeNO) measurement taken by NIOX MINO® Instrument (09-1100) according to the 'Perform FeNO Measurement' guidelines on page 7 of the NIOX MINO® User Manual (February 2011) during their study visit. Following that they will have a Methacholine Challenge (MCC) test performed according to the ATS guidelines and the Allergy and Asthma Specialists Procedure for conducting MCC tests NIOX MINO® Instrument (09-1100): The NIOX MINO® Instrument (09-1100) is a 10 second test based on exhaled breath measured at a 50 ml/second flow rate. |
Measure Participants | 36 |
Asthma |
0
0%
|
Non-Asthma |
36
72%
|
Title | Asthma Diagnosis by FeNO: FeNO <25ppb |
---|---|
Description | Fractional Exhaled Nitric Oxide (FeNO) measurements were performed according to the "Perform FeNO Measurement" guidelines on page 7 of the NIOX MINO® User Manual (February, 2011). Methacholine Challenge (MCC) tests were performed according to the ATS guidelines and the Allergy and Asthma Specialists Procedure for conducting MCC tests. Values from the two tests will be compared against final asthma diagnosis. |
Time Frame | approximately 1-hour |
Outcome Measure Data
Analysis Population Description |
---|
[Not Specified] |
Arm/Group Title | FeNO |
---|---|
Arm/Group Description | Participants with suspected but undiagnosed asthma will have a fractional exhaled nitric oxide (FeNO) measurement taken by NIOX MINO® Instrument (09-1100) according to the 'Perform FeNO Measurement' guidelines on page 7 of the NIOX MINO® User Manual (February 2011) during their study visit. Following that they will have a Methacholine Challenge (MCC) test performed according to the ATS guidelines and the Allergy and Asthma Specialists Procedure for conducting MCC tests NIOX MINO® Instrument (09-1100): The NIOX MINO® Instrument (09-1100) is a 10 second test based on exhaled breath measured at a 50 ml/second flow rate. |
Measure Participants | 38 |
Asthma |
10
20%
|
Non-Asthma |
28
56%
|
Title | Asthma Diagnosis by FeNO: FeNO >=25ppb to <= 50ppb |
---|---|
Description | Fractional Exhaled Nitric Oxide (FeNO) measurements were performed according to the "Perform FeNO Measurement" guidelines on page 7 of the NIOX MINO® User Manual (February, 2011). Methacholine Challenge (MCC) tests were performed according to the ATS guidelines and the Allergy and Asthma Specialists Procedure for conducting MCC tests. Values from the two tests will be compared against final asthma diagnosis. |
Time Frame | approximately 1-hour |
Outcome Measure Data
Analysis Population Description |
---|
[Not Specified] |
Arm/Group Title | FeNO |
---|---|
Arm/Group Description | Participants with suspected but undiagnosed asthma will have a fractional exhaled nitric oxide (FeNO) measurement taken by NIOX MINO® Instrument (09-1100) according to the 'Perform FeNO Measurement' guidelines on page 7 of the NIOX MINO® User Manual (February 2011) during their study visit. Following that they will have a Methacholine Challenge (MCC) test performed according to the ATS guidelines and the Allergy and Asthma Specialists Procedure for conducting MCC tests NIOX MINO® Instrument (09-1100): The NIOX MINO® Instrument (09-1100) is a 10 second test based on exhaled breath measured at a 50 ml/second flow rate. |
Measure Participants | 11 |
Asthma |
0
0%
|
Non-Asthma |
11
22%
|
Title | Asthma Diagnosis by FeNO: FeNO >50ppb |
---|---|
Description | Fractional Exhaled Nitric Oxide (FeNO) measurements were performed according to the "Perform FeNO Measurement" guidelines on page 7 of the NIOX MINO® User Manual (February, 2011). Methacholine Challenge (MCC) tests were performed according to the ATS guidelines and the Allergy and Asthma Specialists Procedure for conducting MCC tests. Values from the two tests will be compared against final asthma diagnosis. |
Time Frame | approximately 1-hour |
Outcome Measure Data
Analysis Population Description |
---|
[Not Specified] |
Arm/Group Title | FeNO |
---|---|
Arm/Group Description | Participants with suspected but undiagnosed asthma will have a fractional exhaled nitric oxide (FeNO) measurement taken by NIOX MINO® Instrument (09-1100) according to the 'Perform FeNO Measurement' guidelines on page 7 of the NIOX MINO® User Manual (February 2011) during their study visit. Following that they will have a Methacholine Challenge (MCC) test performed according to the ATS guidelines and the Allergy and Asthma Specialists Procedure for conducting MCC tests NIOX MINO® Instrument (09-1100): The NIOX MINO® Instrument (09-1100) is a 10 second test based on exhaled breath measured at a 50 ml/second flow rate. |
Measure Participants | 1 |
Asthma |
1
2%
|
Non-Asthma |
0
0%
|
Title | Asthma Diagnosis by FeNO: Mean FeNO Value |
---|---|
Description | Fractional Exhaled Nitric Oxide (FeNO) measurements were performed according to the "Perform FeNO Measurement" guidelines on page 7 of the NIOX MINO® User Manual (February, 2011). Methacholine Challenge (MCC) tests were performed according to the ATS guidelines and the Allergy and Asthma Specialists Procedure for conducting MCC tests. Values from the two tests will be compared against final asthma diagnosis. |
Time Frame | approximately 1-hour |
Outcome Measure Data
Analysis Population Description |
---|
[Not Specified] |
Arm/Group Title | FeNO |
---|---|
Arm/Group Description | Participants with suspected but undiagnosed asthma will have a fractional exhaled nitric oxide (FeNO) measurement taken by NIOX MINO® Instrument (09-1100) according to the 'Perform FeNO Measurement' guidelines on page 7 of the NIOX MINO® User Manual (February 2011) during their study visit. Following that they will have a Methacholine Challenge (MCC) test performed according to the ATS guidelines and the Allergy and Asthma Specialists Procedure for conducting MCC tests NIOX MINO® Instrument (09-1100): The NIOX MINO® Instrument (09-1100) is a 10 second test based on exhaled breath measured at a 50 ml/second flow rate. |
Measure Participants | 50 |
Asthma |
16.9
(12.21)
|
Non-Asthma |
17.8
(10.11)
|
Title | Sensitivity (%) for FeNO |
---|---|
Description | Fractional Exhaled Nitric Oxide (FeNO) measurements were performed according to the "Perform FeNO Measurement" guidelines on page 7 of the NIOX MINO® User Manual (February, 2011). Methacholine Challenge (MCC) tests were performed according to the ATS guidelines and the Allergy and Asthma Specialists Procedure for conducting MCC tests. Values from the two tests will be compared against final asthma diagnosis. Sensitivity is a way of saying how likely a test is positive if you have a disease and is expressed as (%) percent. For this study, sensitivity was the ability of FeNO to correctly identify a study participant who has asthma. |
Time Frame | approximately 1-hour |
Outcome Measure Data
Analysis Population Description |
---|
[Not Specified] |
Arm/Group Title | FeNO |
---|---|
Arm/Group Description | Participants with suspected but undiagnosed asthma will have a fractional exhaled nitric oxide (FeNO) measurement taken by NIOX MINO® Instrument (09-1100) according to the 'Perform FeNO Measurement' guidelines on page 7 of the NIOX MINO® User Manual (February 2011) during their study visit. Following that they will have a Methacholine Challenge (MCC) test performed according to the ATS guidelines and the Allergy and Asthma Specialists Procedure for conducting MCC tests NIOX MINO® Instrument (09-1100): The NIOX MINO® Instrument (09-1100) is a 10 second test based on exhaled breath measured at a 50 ml/second flow rate. |
Measure Participants | 50 |
FeNO <= 10 ppb |
35.7
|
FeNO <= 15 ppb |
57.1
|
FeNO <=20 ppb |
85.7
|
FeNO <=25 ppb |
85.7
|
FeNO <=30 ppb |
85.7
|
FeNO <= 35 ppb |
85.7
|
FeNO <= 40 ppb |
92.9
|
FeNO <= 45 ppb |
92.9
|
FeNO <= 50 ppb |
92.9
|
FeNO <= 55 ppb |
100.0
|
Title | Specificity (%) for FeNO |
---|---|
Description | Fractional Exhaled Nitric Oxide (FeNO) measurements were performed according to the "Perform FeNO Measurement" guidelines on page 7 of the NIOX MINO® User Manual (February, 2011). Methacholine Challenge (MCC) tests were performed according to the ATS guidelines and the Allergy and Asthma Specialists Procedure for conducting MCC tests. Values from the two tests will be compared against final asthma diagnosis. Specificity is a way of saying how likely a test is negative if you do not have a disease and is expressed as (%) percent. For this study, specificity was the ability of FeNO to correctly identify a study participant who does not have asthma. |
Time Frame | approximately 1-hour |
Outcome Measure Data
Analysis Population Description |
---|
[Not Specified] |
Arm/Group Title | FeNO |
---|---|
Arm/Group Description | Participants with suspected but undiagnosed asthma will have a fractional exhaled nitric oxide (FeNO) measurement taken by NIOX MINO® Instrument (09-1100) according to the 'Perform FeNO Measurement' guidelines on page 7 of the NIOX MINO® User Manual (February 2011) during their study visit. Following that they will have a Methacholine Challenge (MCC) test performed according to the ATS guidelines and the Allergy and Asthma Specialists Procedure for conducting MCC tests NIOX MINO® Instrument (09-1100): The NIOX MINO® Instrument (09-1100) is a 10 second test based on exhaled breath measured at a 50 ml/second flow rate. |
Measure Participants | 50 |
FeNO <= 10 ppb |
75.0
|
FeNO <= 15 ppb |
55.6
|
FeNO <=20 ppb |
33.3
|
FeNO <=25 ppb |
25.0
|
FeNO <=30 ppb |
11.1
|
FeNO <= 35 ppb |
2.8
|
FeNO <= 40 ppb |
2.8
|
FeNO <= 45 ppb |
0.0
|
FeNO <= 50 ppb |
0.0
|
FeNO <= 55 ppb |
0.0
|
Title | Positive Predictive Value (%) for FeNO |
---|---|
Description | Fractional Exhaled Nitric Oxide (FeNO) measurements were performed according to the "Perform FeNO Measurement" guidelines on page 7 of the NIOX MINO® User Manual (February, 2011). Methacholine Challenge (MCC) tests were performed according to the ATS guidelines and the Allergy and Asthma Specialists Procedure for conducting MCC tests. Values from the two tests will be compared against final asthma diagnosis. Positive predictive value is a way of saying the likelihood a positive test means you have a disease. In this study, it is defined as the percentage of study participants with a high FeNO who have asthma. |
Time Frame | approximately 1-hour |
Outcome Measure Data
Analysis Population Description |
---|
[Not Specified] |
Arm/Group Title | FeNO |
---|---|
Arm/Group Description | Participants with suspected but undiagnosed asthma will have a fractional exhaled nitric oxide (FeNO) measurement taken by NIOX MINO® Instrument (09-1100) according to the 'Perform FeNO Measurement' guidelines on page 7 of the NIOX MINO® User Manual (February 2011) during their study visit. Following that they will have a Methacholine Challenge (MCC) test performed according to the ATS guidelines and the Allergy and Asthma Specialists Procedure for conducting MCC tests NIOX MINO® Instrument (09-1100): The NIOX MINO® Instrument (09-1100) is a 10 second test based on exhaled breath measured at a 50 ml/second flow rate. |
Measure Participants | 50 |
FeNO <= 10 ppb |
35.7
|
FeNO <= 15 ppb |
33.3
|
FeNO <=20 ppb |
33.3
|
FeNO <=25 ppb |
30.8
|
FeNO <=30 ppb |
27.3
|
FeNO <= 35 ppb |
25.5
|
FeNO <= 40 ppb |
27.1
|
FeNO <= 45 ppb |
26.5
|
FeNO <= 50 ppb |
26.5
|
FeNO <= 55 ppb |
28.0
|
Title | Negative Predictive Value (%) for FeNO |
---|---|
Description | Fractional Exhaled Nitric Oxide (FeNO) measurements were performed according to the "Perform FeNO Measurement" guidelines on page 7 of the NIOX MINO® User Manual (February, 2011). Methacholine Challenge (MCC) tests were performed according to the ATS guidelines and the Allergy and Asthma Specialists Procedure for conducting MCC tests. Values from the two tests will be compared against final asthma diagnosis. Negative predictive value is a way of saying the likelihood of a negative test means you do not have disease. In this study it is defined as the percentage of study participants with a low FeNO who do not have asthma. |
Time Frame | approximately 1-hour |
Outcome Measure Data
Analysis Population Description |
---|
[Not Specified] |
Arm/Group Title | FeNO |
---|---|
Arm/Group Description | Participants with suspected but undiagnosed asthma will have a fractional exhaled nitric oxide (FeNO) measurement taken by NIOX MINO® Instrument (09-1100) according to the 'Perform FeNO Measurement' guidelines on page 7 of the NIOX MINO® User Manual (February 2011) during their study visit. Following that they will have a Methacholine Challenge (MCC) test performed according to the ATS guidelines and the Allergy and Asthma Specialists Procedure for conducting MCC tests NIOX MINO® Instrument (09-1100): The NIOX MINO® Instrument (09-1100) is a 10 second test based on exhaled breath measured at a 50 ml/second flow rate. |
Measure Participants | 50 |
FeNO <= 10 ppb |
75.0
|
FeNO <= 15 ppb |
76.9
|
FeNO <=20 ppb |
85.7
|
FeNO <=25 ppb |
81.8
|
FeNO <=30 ppb |
66.7
|
FeNO <= 35 ppb |
33.3
|
FeNO <= 40 ppb |
50.0
|
Title | FeNO Values by ICS Use: FeNO <25ppb |
---|---|
Description | Fractional Exhaled Nitric Oxide (FeNO) measurements were performed according to the "Perform FeNO Measurement" guidelines on page 7 of the NIOX MINO® User Manual (February, 2011). Methacholine Challenge (MCC) tests were performed according to the ATS guidelines and the Allergy and Asthma Specialists Procedure for conducting MCC tests. Values from the two tests were compared against final asthma diagnosis. Data are presented for participants with FeNO <25ppb who either did not use Inhaled Corticosteroids (ICS) or did use Inhaled Corticosteroids. |
Time Frame | approximately 1-hour |
Outcome Measure Data
Analysis Population Description |
---|
[Not Specified] |
Arm/Group Title | FeNO |
---|---|
Arm/Group Description | Participants with suspected but undiagnosed asthma will have a fractional exhaled nitric oxide (FeNO) measurement taken by NIOX MINO® Instrument (09-1100) according to the 'Perform FeNO Measurement' guidelines on page 7 of the NIOX MINO® User Manual (February 2011) during their study visit. Following that they will have a Methacholine Challenge (MCC) test performed according to the ATS guidelines and the Allergy and Asthma Specialists Procedure for conducting MCC tests NIOX MINO® Instrument (09-1100): The NIOX MINO® Instrument (09-1100) is a 10 second test based on exhaled breath measured at a 50 ml/second flow rate. |
Measure Participants | 38 |
No ICS Use |
34
68%
|
Yes ICS Use |
4
8%
|
Title | FeNO Values by ICS Use: FeNO >= 50ppb |
---|---|
Description | Fractional Exhaled Nitric Oxide (FeNO) measurements were performed according to the "Perform FeNO Measurement" guidelines on page 7 of the NIOX MINO® User Manual (February, 2011). Methacholine Challenge (MCC) tests were performed according to the ATS guidelines and the Allergy and Asthma Specialists Procedure for conducting MCC tests. Values from the two tests were compared against final asthma diagnosis. Data are presented for participants with FeNO >=50ppb who either did not use Inhaled Corticosteroids (ICS) or did use Inhaled Corticosteroids. |
Time Frame | approximately 1-hour |
Outcome Measure Data
Analysis Population Description |
---|
[Not Specified] |
Arm/Group Title | FeNO |
---|---|
Arm/Group Description | Participants with suspected but undiagnosed asthma will have a fractional exhaled nitric oxide (FeNO) measurement taken by NIOX MINO® Instrument (09-1100) according to the 'Perform FeNO Measurement' guidelines on page 7 of the NIOX MINO® User Manual (February 2011) during their study visit. Following that they will have a Methacholine Challenge (MCC) test performed according to the ATS guidelines and the Allergy and Asthma Specialists Procedure for conducting MCC tests NIOX MINO® Instrument (09-1100): The NIOX MINO® Instrument (09-1100) is a 10 second test based on exhaled breath measured at a 50 ml/second flow rate. |
Measure Participants | 1 |
No ICS Use |
1
2%
|
Yes ICS Use |
0
0%
|
Title | FeNO Values by ICS Use: FeNO >= 25ppb to <= 50ppb |
---|---|
Description | Fractional Exhaled Nitric Oxide (FeNO) measurements were performed according to the "Perform FeNO Measurement" guidelines on page 7 of the NIOX MINO® User Manual (February, 2011). Methacholine Challenge (MCC) tests were performed according to the ATS guidelines and the Allergy and Asthma Specialists Procedure for conducting MCC tests. Values from the two tests were compared against final asthma diagnosis. Data are presented for participants with FeNO <=25ppb to <=50ppb who either did not use Inhaled Corticosteroids (ICS) or did use Inhaled Corticosteroids. |
Time Frame | approximately 1-hour |
Outcome Measure Data
Analysis Population Description |
---|
[Not Specified] |
Arm/Group Title | FeNO |
---|---|
Arm/Group Description | Participants with suspected but undiagnosed asthma will have a fractional exhaled nitric oxide (FeNO) measurement taken by NIOX MINO® Instrument (09-1100) according to the 'Perform FeNO Measurement' guidelines on page 7 of the NIOX MINO® User Manual (February 2011) during their study visit. Following that they will have a Methacholine Challenge (MCC) test performed according to the ATS guidelines and the Allergy and Asthma Specialists Procedure for conducting MCC tests NIOX MINO® Instrument (09-1100): The NIOX MINO® Instrument (09-1100) is a 10 second test based on exhaled breath measured at a 50 ml/second flow rate. |
Measure Participants | 11 |
No ICS Use |
10
20%
|
Yes ICS Use |
1
2%
|
Title | FeNO Values by Smoking Status: FeNO <25ppb |
---|---|
Description | Fractional Exhaled Nitric Oxide (FeNO) measurements were performed according to the "Perform FeNO Measurement" guidelines on page 7 of the NIOX MINO® User Manual (February, 2011). Methacholine Challenge (MCC) tests were performed according to the ATS guidelines and the Allergy and Asthma Specialists Procedure for conducting MCC tests. Values from the two tests were compared against final asthma diagnosis. Data are presented for participants with FeNO <25ppb who either did not smoke previously or are current smokers. |
Time Frame | approximately 1-hour |
Outcome Measure Data
Analysis Population Description |
---|
[Not Specified] |
Arm/Group Title | FeNO |
---|---|
Arm/Group Description | Participants with suspected but undiagnosed asthma will have a fractional exhaled nitric oxide (FeNO) measurement taken by NIOX MINO® Instrument (09-1100) according to the 'Perform FeNO Measurement' guidelines on page 7 of the NIOX MINO® User Manual (February 2011) during their study visit. Following that they will have a Methacholine Challenge (MCC) test performed according to the ATS guidelines and the Allergy and Asthma Specialists Procedure for conducting MCC tests NIOX MINO® Instrument (09-1100): The NIOX MINO® Instrument (09-1100) is a 10 second test based on exhaled breath measured at a 50 ml/second flow rate. |
Measure Participants | 38 |
No Previous Smoker |
34
68%
|
Current Smoker |
4
8%
|
Title | FeNO Values by Smoking Status: FeNO >= 25ppb to <= 50ppb |
---|---|
Description | Fractional Exhaled Nitric Oxide (FeNO) measurements were performed according to the "Perform FeNO Measurement" guidelines on page 7 of the NIOX MINO® User Manual (February, 2011). Methacholine Challenge (MCC) tests were performed according to the ATS guidelines and the Allergy and Asthma Specialists Procedure for conducting MCC tests. Values from the two tests were compared against final asthma diagnosis. Data are presented for participants with FeNO >=25ppb to <=50ppb who either did not smoke previously or are current smokers. |
Time Frame | approximately 1-hour |
Outcome Measure Data
Analysis Population Description |
---|
[Not Specified] |
Arm/Group Title | FeNO |
---|---|
Arm/Group Description | Participants with suspected but undiagnosed asthma will have a fractional exhaled nitric oxide (FeNO) measurement taken by NIOX MINO® Instrument (09-1100) according to the 'Perform FeNO Measurement' guidelines on page 7 of the NIOX MINO® User Manual (February 2011) during their study visit. Following that they will have a Methacholine Challenge (MCC) test performed according to the ATS guidelines and the Allergy and Asthma Specialists Procedure for conducting MCC tests NIOX MINO® Instrument (09-1100): The NIOX MINO® Instrument (09-1100) is a 10 second test based on exhaled breath measured at a 50 ml/second flow rate. |
Measure Participants | 11 |
No Previous Smoker |
11
22%
|
Current Smoker |
0
0%
|
Title | FeNO Values by Smoking Status: FeNO >= 50ppb |
---|---|
Description | Fractional Exhaled Nitric Oxide (FeNO) measurements were performed according to the "Perform FeNO Measurement" guidelines on page 7 of the NIOX MINO® User Manual (February, 2011). Methacholine Challenge (MCC) tests were performed according to the ATS guidelines and the Allergy and Asthma Specialists Procedure for conducting MCC tests. Values from the two tests were compared against final asthma diagnosis. Data are presented for participants with FeNO >=50ppb who either did not smoke previously or are current smokers. |
Time Frame | approximately 1-hour |
Outcome Measure Data
Analysis Population Description |
---|
[Not Specified] |
Arm/Group Title | FeNO |
---|---|
Arm/Group Description | Participants with suspected but undiagnosed asthma will have a fractional exhaled nitric oxide (FeNO) measurement taken by NIOX MINO® Instrument (09-1100) according to the 'Perform FeNO Measurement' guidelines on page 7 of the NIOX MINO® User Manual (February 2011) during their study visit. Following that they will have a Methacholine Challenge (MCC) test performed according to the ATS guidelines and the Allergy and Asthma Specialists Procedure for conducting MCC tests NIOX MINO® Instrument (09-1100): The NIOX MINO® Instrument (09-1100) is a 10 second test based on exhaled breath measured at a 50 ml/second flow rate. |
Measure Participants | 1 |
No Previous Smoker |
0
0%
|
Current Smoker |
1
2%
|
Adverse Events
Time Frame | 6 months | |
---|---|---|
Adverse Event Reporting Description | Each subject from the time of consent until study completion. | |
Arm/Group Title | FeNO | |
Arm/Group Description | Participants with suspected but undiagnosed asthma had a fractional exhaled nitric oxide (FeNO) measurement taken by NIOX MINO® Instrument (09-1100) according to the 'Perform FeNO Measurement' guidelines on page 7 of the NIOX MINO® User Manual (February 2011) during their study visit. Following that they had a Methacholine Challenge (MCC) test performed according to the ATS guidelines and the Allergy and Asthma Specialists Procedure for conducting MCC tests NIOX MINO® Instrument (09-1100): The NIOX MINO® Instrument (09-1100) is a 10 second test based on exhaled breath measured at a 50 ml/second flow rate. | |
All Cause Mortality |
||
FeNO | ||
Affected / at Risk (%) | # Events | |
Total | / (NaN) | |
Serious Adverse Events |
||
FeNO | ||
Affected / at Risk (%) | # Events | |
Total | 0/50 (0%) | |
Other (Not Including Serious) Adverse Events |
||
FeNO | ||
Affected / at Risk (%) | # Events | |
Total | 0/50 (0%) |
Limitations/Caveats
More Information
Certain Agreements
Principal Investigators are NOT employed by the organization sponsoring the study.
The only disclosure restriction on the PI is that the sponsor can review results communications prior to public release and can embargo communications regarding trial results for a period that is less than or equal to 60 days. The sponsor cannot require changes to the communication and cannot extend the embargo.
Results Point of Contact
Name/Title | Nancy Herje, BSN, RN, MBA. Director, Clinical Development and Medical Affairs |
---|---|
Organization | Aerocrine, Inc. |
Phone | 919-449-8873 |
nancy.herje@aerocrine.com |
- AER-044