Safety and Effectiveness Study of the MEL 80 Excimer Laser Using LASIK in the Treatment of Hyperopia

Sponsor
Carl Zeiss Meditec, Inc. (Industry)
Overall Status
Completed
CT.gov ID
NCT00762281
Collaborator
(none)
189
6
1
51
31.5
0.6

Study Details

Study Description

Brief Summary

The purpose of this study is to determine whether the MEL 80 Excimer Laser is effective in the treatment of hyperopia, when used as part of the Laser In Situ Keratomileusis (LASIK) procedure.

Condition or Disease Intervention/Treatment Phase
  • Device: MEL 80 Hyperopic LASIK Treatment
N/A

Detailed Description

LASIK has become one of the most common refractive eye procedures performed in the country. In the hyperopia procedure, a steepening occurs on the corneal surface, rather than the flattening procedure (myopic treatment). The surgeon will produce a standard keratomileusis flap using a microkeratome, exposing the corneal stroma. Recontouring under the flap is then accomplished by the removal of tissue from the stroma with the laser. This recontouring results in an altering of effective lens power of the central cornea, measured in diopters (D). The MEL 80 Excimer Laser System will be evaluated for its ability to create accurate and stable hyperopic refractive correction results.

Study Design

Study Type:
Interventional
Actual Enrollment :
189 participants
Allocation:
N/A
Intervention Model:
Single Group Assignment
Masking:
None (Open Label)
Primary Purpose:
Treatment
Official Title:
A Prospective, Randomized, Multicenter Trial To Evaluate The Safety And Effectiveness Of The MEL 80 Excimer Laser Using LASIK (Laser In Situ Keratomileusis) For The Correction Of ≤ +6.0 D Of Hyperopia With Or Without Astigmatism Of +0.50 D To +3.50 D And MRSE ≤ +6.50 D
Study Start Date :
Jul 1, 2004
Actual Primary Completion Date :
Oct 1, 2006
Actual Study Completion Date :
Oct 1, 2008

Arms and Interventions

Arm Intervention/Treatment
Other: Treatment of Hyperopic LASIK

Treatment of Hyperopic corrections ≤ +6.0 D with or without Astigmatism of +0.50 to +3.50 D and MRSE ≤ +6.50 D.

Device: MEL 80 Hyperopic LASIK Treatment
Treatment of Hyperopic corrections ≤ +6.0 D with or without Astigmatism of +0.50 to +3.50 D and MRSE ≤ +6.50 D.

Outcome Measures

Primary Outcome Measures

  1. At the point of stability, a minimum of 75% of eyes should have an achieved refraction within ± 1.00 D of the intended outcome, and at least 50% of eyes should be within ± 0.50 D of the intended outcome. [Point of stability]

  2. A minimum of 85% of eyes targeted for emmetropia should have an uncorrected visual acuity of 20/40 or better at the postoperative interval at which stability has been established. [Point of stability]

  3. A minimum of 95% of eyes should have a change of < 1.00 D in manifest refraction spherical equivalent (MRSE) between 2 refractions performed at least 3 months apart, and the mean rate of MRSE change per month should be < 0.04 D. [Point of stability]

  4. 75% of eyes undergoing astigmatic treatment should be within ± 1.00 D of the attempted astigmatism correction by the point of stability. [Point of stability]

  5. Distance BSCVA of worse than 20/40 at the postoperative interval at which stability has been established should occur in less than 1.0% of eyes that had a BSCVA of 20/20 or better before surgery. [Point of stability]

  6. Loss of more than 2 lines of BSCVA should occur in less than 5.0% of eyes. [Point of stability]

  7. Less than 5% of eyes treated for sphere only should have a magnitude of postoperative manifest refractive astigmatism that varies from baseline cylinder by greater than 2.00 D at the postoperative interval at which stability has been established. [Point of stability]

  8. Incidence of Adverse Events to occur in less 1% of eyes [Postoperative visits]

Secondary Outcome Measures

  1. Subject Satisfaction: As measured by a subjective questionnaire, and will be considered as a secondary efficacy variable. [Postoperative visits 3, 6, 9, 12, 18 and 24 months]

  2. Incidence of Complications [Postoperative visits]

  3. Patient Symptoms: As measured by a subjective questionnaire, will be considered as a secondary safety variable. [Preoperative and Postoperative visits 3, 6, 9, 12, 18 and 24 months]

Eligibility Criteria

Criteria

Ages Eligible for Study:
21 Years and Older
Sexes Eligible for Study:
All
Accepts Healthy Volunteers:
Yes
Inclusion Criteria:
  • Naturally occurring hyperopia up to +6.0 D, with or without astigmatism of +0.50 D to +3.50 D at the spectacle plane, and MRSE ≤ +6.50 D;

  • Have ≤ 0.75 D of latent hyperopia as determined by the difference between the preoperative MRSE and CRSE;

  • A stable refraction for at least the last 12 months as documented by previous clinical records, i.e., the spherical and cylindrical portions of the manifest refraction have not progressed at a rate of more than 0.50 D during the year prior to the baseline examination in the eye to be treated;

  • Hard contact lens wearers must have two central keratometry readings and two manifest refractions taken at least one week apart that do not differ by more than 0.50 D;

  • Visual acuity correctable to at least 20/40 in both eyes;

  • Operative eye must be targeted for emmetropia;

  • At least 21 years of age;

  • Willing and able to return for scheduled follow up examinations for 24 months after surgery;

  • Sign and be given a copy of the written Informed Consent form.

Exclusion Criteria:
  • A history of anterior segment pathology, including cataracts (in the operative eye);

  • Severe dry eye syndrome unresolved by treatment;

  • Residual, recurrent, active ocular or uncontrolled eyelid disease, corneal scars within the ablation zone or other corneal abnormality such as recurrent corneal erosion or severe basement membrane disease;

  • Ophthalmoscopic signs of keratoconus (or keratoconus suspect);

  • An ablation deeper than 250 microns from the corneal endothelium;

  • Irregular or unstable (distorted/not clear) corneal mires on central keratometry readings;

  • Blind in the fellow eye;

  • Undergone previous intraocular or corneal surgery of any kind in the operative eye(s), including any type of excimer laser surgery for either refractive or therapeutic purposes;

  • A history of ocular Herpes zoster or Herpes simplex keratitis;

  • A history of steroid-responsive rise in intraocular pressure, glaucoma, or preoperative IOP >21 mm Hg;

  • Diabetes, diagnosed autoimmune disease, connective tissue disease, or clinically significant atopic syndrome;

  • Immunocompromised or use chronic systemic corticosteroid or other immunosuppressive therapy;

  • Pregnant, lactating, or be of childbearing potential and not practicing a medically approved method of birth control;

  • A known sensitivity to planned study medications;

  • Participating in any other ophthalmic drug or device clinical trial during the time of this clinical investigation;

  • At risk for angle closure or for developing strabismus postoperatively.

  • For Fellow (Second) Eyes in Simultaneous Bilateral Treatment Procedures

    1. Flap complications during the first eye's surgery such as a free cap, partial flap, thin flap, or irregular flap.
    1. Epithelial defect exceeding 2 mm x 2 mm in dimension, or clinically significant debris in the interface between the flap and underlying stroma for the first eye.
    1. Severe blepharospasm in the first eye that may have prevented or impeded the completion of the keratectomy and/or the laser ablation procedure.
    1. Poor subject cooperation with instructions for the first eye's surgery and/or poor subject fixation on the laser fixation target.
    1. Aborted LASIK procedure in the first eye or PRK was performed in the first eye because LASIK was not possible.

Contacts and Locations

Locations

Site City State Country Postal Code
1 US Navy Refractive Surgery Center San Diego California United States 92106
2 Dishler Laser Institute Greenwood Village Colorado United States 80111
3 Discover Vision Centers Kansas City Missouri United States 64055
4 Fine, Hoffman, and Packer Eugene Oregon United States 97401
5 Texan Eye Care Austin Texas United States 78746
6 Davis Duehr Dean Eye Clinic Madison Wisconsin United States 53717

Sponsors and Collaborators

  • Carl Zeiss Meditec, Inc.

Investigators

  • Principal Investigator: Steven Dell, MD, Texan Eye Care
  • Principal Investigator: John Doane, MD, Discover Vision Centers
  • Principal Investigator: Richard Hoffman, MD, Fine, Hoffman, and Packer LLC
  • Principal Investigator: Howard Fine, MD, Fine, Hoffman, and Packer LLC
  • Principal Investigator: Mark Packer, MD, FIne, Hoffman, and Packer LLC
  • Principal Investigator: David Tanzer, MD, US Navy Refractive Surgery Center, San Diego, CA
  • Principal Investigator: Steve Schallhorn, MD, US Navy Refractive Surgery Center, San Diego, CA
  • Principal Investigator: John Vukich, MD, Davis Duehr Dean Eye Clinic
  • Principal Investigator: Jon Dishler, MD, Dishler Laser Institute

Study Documents (Full-Text)

None provided.

More Information

Publications

None provided.
Responsible Party:
Carl Zeiss Meditec, Inc.
ClinicalTrials.gov Identifier:
NCT00762281
Other Study ID Numbers:
  • MEL 80-2004-2
First Posted:
Sep 30, 2008
Last Update Posted:
Aug 13, 2012
Last Verified:
Aug 1, 2012

Study Results

No Results Posted as of Aug 13, 2012