A Study to Evaluate Accuracy and Validity of the "Chang Gung" Ventricular Systolic Dysfunction Screening Software

Sponsor
Chang Gung Memorial Hospital (Other)
Overall Status
Recruiting
CT.gov ID
NCT05861115
Collaborator
(none)
1,172
1
1
4.7
249.5

Study Details

Study Description

Brief Summary

The purpose of this research is to test a software tool called the "Chang Gung" Ventricular Systolic Dysfunction screening software, which uses a 12-lead electrocardiogram to determine if a patient has left ventricular systolic dysfunction. The goal is to determine if the software can accurately identify patients with this condition, which would help doctors diagnose and treat it more effectively.

The trial will involve using the software on patients and comparing its results to those obtained through echocardiograms, which are currently the gold standard for diagnosing left ventricular systolic dysfunction. Only patients who meet specific eligibility criteria will be able to participate in the trial, and the software will be administered by trained healthcare professionals.

The study will help determine if the software is a useful tool for diagnosing left ventricular systolic dysfunction, which could lead to earlier diagnosis and better outcomes for patients. The research team will collect and analyze data on the accuracy of the software and its usability in clinical practice.

Overall, this study will provide important information for doctors and patients about a new tool for diagnosing left ventricular systolic dysfunction.

Condition or Disease Intervention/Treatment Phase
  • Device: "Chang Gung" Ventricular Systolic Dysfunction Screening Software
N/A

Detailed Description

This is a retrospective study conducted in Taiwan that aimed to test the performance of "Chang Gung" Ventricular Systolic Dysfunction screening software using electrocardiogram (ECG) and echocardiography data. The study data was obtained from a database that includes six hospitals of Chang Gung Memorial Hospital, between January 1, 2006, and December 31, 2019.

The software was developed using a training set of 133,225 data and validated using a set of 57,134 data. For clinical validation, a total of 1,172 test data were randomly selected from the testing set, stratified by hospital classification, age group and gender. The hospital classification, age group and gender ratios were based on the proportion of the testing data. The test data were also stratified by the presence or absence of left ventricular systolic dysfunction(LVSD) for test group and control group, defined as a heart output rate of less than 40% within 14 days before and after the ECG recording.

During the clinical trial, a cardiologist with 15 years of experience in treating cardiovascular disease examined the ECG data without any exclusion criteria. The cardiologist also confirmed the accuracy of the left ventricular ejection fraction (LVEF) measurement, which was defined as LVSD in the echocardiography reports. The LVEF was extracted from legally-binding echocardiography reports, not by the cardiologist during the clinical trial. The ECG data were screened and filtered for quality before being input into the software. The primary outcome was the sensitivity of the software, which was defined as not inferior to 0.86. The study also analyzed secondary outcome measures, including the area under the receiver operating characteristic curve, accuracy, specificity, positive predictive value, negative predictive value, false-positive rate, and false-negative rate.

Study Design

Study Type:
Interventional
Anticipated Enrollment :
1172 participants
Allocation:
N/A
Intervention Model:
Single Group Assignment
Intervention Model Description:
The investigators collected retrospective data and gave to the software to interpret. Echocardiography as the gold standard, and tested whether the software could correctly interpret left ventricular systolic dysfunction.The investigators collected retrospective data and gave to the software to interpret. Echocardiography as the gold standard, and tested whether the software could correctly interpret left ventricular systolic dysfunction.
Masking:
None (Open Label)
Primary Purpose:
Diagnostic
Official Title:
A Study to Evaluate Accuracy and Validity of the "Chang Gung" Ventricular Systolic Dysfunction Screening Software
Actual Study Start Date :
Apr 15, 2023
Anticipated Primary Completion Date :
Jun 5, 2023
Anticipated Study Completion Date :
Sep 5, 2023

Arms and Interventions

Arm Intervention/Treatment
Experimental: Software diagnosis

Software diagnosis with gold standard of echocardiography.

Device: "Chang Gung" Ventricular Systolic Dysfunction Screening Software
This software is suitable for 12-lead ECG signals of adults over 20 years old, and assists doctors in screening patients for left ventricular systolic dysfunction.

Outcome Measures

Primary Outcome Measures

  1. Sensitivity [baseline]

    The rate of test results that correctly indicate the presence.

Secondary Outcome Measures

  1. Specificity [Baseline]

    The rate of test results that correctly indicate the absence.

  2. Accuracy [Baseline]

    The rate of all test results that correctly indicated.

  3. Area Under the receiver operating characteristic Curve [Baseline]

    A graphical plot that illustrates the diagnostic ability of a binary classifier system as its discrimination threshold is varied.

  4. Positive predictive value [Baseline]

    The proportions of positive results in statistics and diagnostic tests that are true positive results

  5. Negative predictive value [Baseline]

    The proportions of negative results in statistics and diagnostic tests that are true negative results

  6. False positive rate [Baseline]

    The rate of test result which wrongly indicates that a particular condition or attribute is present

  7. False negative rate [Baseline]

    The rate of test result which wrongly indicates that a particular condition or attribute is absent

Eligibility Criteria

Criteria

Ages Eligible for Study:
20 Years to 100 Years
Sexes Eligible for Study:
All
Accepts Healthy Volunteers:
No
Inclusion Criteria:
  • Equal or greater than twenty years old.

  • Static 12-lead electrocardiogram of General Electric MUSE XML format file.

  • The data comes from the static 12-lead electrocardiogram device of General Electric (model MAC5500).

  • Echocardiography data within 14 days before or after the electrocardiogram acquisition time.

  • The electrocardiogram signal is 500 Hz.

  • The Alternating current (AC) filter of the electrocardiogram signal is 60 Hz.

  • The length of the electrocardiogram signal is ten seconds (the electrocardiogram output of model MAC5500 is ten seconds, and there is no need to capture a segment).

Exclusion Criteria:
  • Cases used in the model development process.

  • Lacks any electrode.

  • Contain any electrode lacks a segment.

Contacts and Locations

Locations

Site City State Country Postal Code
1 Chang Gung memorial hospital Taoyuan City Taiwan 333

Sponsors and Collaborators

  • Chang Gung Memorial Hospital

Investigators

  • Study Chair: Chang-Fu Kuo, MD/Ph.D, Associate Professor and Director Division of Rheumatology

Study Documents (Full-Text)

None provided.

More Information

Publications

Responsible Party:
Chang Gung Memorial Hospital
ClinicalTrials.gov Identifier:
NCT05861115
Other Study ID Numbers:
  • 202200402B0C602
First Posted:
May 16, 2023
Last Update Posted:
May 16, 2023
Last Verified:
May 1, 2023
Individual Participant Data (IPD) Sharing Statement:
No
Plan to Share IPD:
No
Studies a U.S. FDA-regulated Drug Product:
No
Studies a U.S. FDA-regulated Device Product:
No
Additional relevant MeSH terms:

Study Results

No Results Posted as of May 16, 2023