Isokinetic Performance After Cruciate-substituting Ultra-congruent and Posterior Stabilized Total Knee Arthroplasties

Sponsor
Aksaray University Training and Research Hospital (Other)
Overall Status
Completed
CT.gov ID
NCT04419311
Collaborator
(none)
66
1
2
14.3
4.6

Study Details

Study Description

Brief Summary

The choice between ultra-congruent (UC) insert or posterior cruciate ligament-stabilized (PS) insert in posterior cruciate ligament (PCL) sacrificing total knee arthroplasty (TKA) remains debatable. Despite the potential advantages of the UC insert over PS insert with its different design, there are concerns about inferior clinical outcome related to its use. Therefore, isokinetic performance was used in this study to objectively evaluate knee function after TKA and the clinical scores of the patients were also evaluated. To the best of our knowledge, no prospective randomized study has compared the isokinetic performance of the knee following the use of UC and PS inserts in TKA. The hypothesis of the present study was that compared with the PS insert, the UC insert would be associated with a lower clinical outcome and isokinetic performance following TKA.

A total of 65 patients scheduled to undergo TKA on for primary knee osteoarthritis were randomly assigned to either the UC (32 patients) or the PS group (33 patients). The Knee Society Score (KSS) and isokinetic performance results of each patient were recorded preoperatively and at postoperative 3, 6 and 12 months. The physiatrist performing isokinetic tests and patients were blinded to the study.

Condition or Disease Intervention/Treatment Phase
  • Device: Vanguard® Knee System
N/A

Detailed Description

Eligibility of all patients between 55 and 80 years, scheduled to undergo unilateral TKA on for primary knee osteoarthritis will be evaluated. The exclusion criteria are rheumatological joint diseases, previous knee surgery, neuromuscular diseases, bilateral TKA, or insufficiency of collateral ligaments. Patients will be randomized in a 1:1 ratio via computer-generated randomization using Microsoft Excel 2016 (Microsoft Corporation, Seattle, WA, USA) to be allocated in the UC insert or PS insert group before the operation. Unblinded senior resident will implement the randomization. Patients and physiatrists performing isokinetic measurements will be blinded to group allocation.

During the UC and PS operations, Vanguard® Complete Knee System prosthesis (Zimmer Biomet Inc., Warsaw, IN, USA) will be implanted using the same surgical technique in all patients. The Vanguard anterior-stabilized (AS) insert is a UC deep-dish design with a 10 mm prominent anterior lip and 5mm posterior lip. This bone-conserving design prevents anterior femoral subluxation because of the prominent anterior lip. The insert design allows it to be used with the Vanguard cruciate-retaining femoral component and the highly congruent articulating surface increases rotational stability. There is more contact area between the femoral component and the weight-bearing surface to decrease the shear stress between the femur and polyethylene insert.

A tourniquet will be inflated to pressure of 300 mmHg after spinal anesthesia. All operations will be performed with the same surgical technique by a single senior surgeon. Patellar surface arthroplasty will be performed in all the cases. Both femoral and tibial prostheses will be implanted with pressured bone cement. A suction drain will be placed inside the knee capsule. On the morning of the first day after surgery patients will be mobilized under the supervision of the physical therapist. All patients will undergo the same rehabilitation procedure during the outpatient period and will be examined by the same physiatrist at the regular 3, 6, and 12-month follow-up visits.

The primary outcome is isokinetic performance, measured as peak knee extensor and flexor torque values in Newton-meters, on the operated knee. Measurements wilil be done preoperatively and at 3, 6, and 12-month follow-up examinations under the supervision of the same physiatrist. At same time points, the Knee Society Score (KSS) will be evaluated as well. Isokinetic measurements will be performed using a Biodex System III Isokinetic Dynamometer, version 3.03 (Biodex Medical Inc., Shirley, NY, USA) by the same senior physiatrist. Patients will be positioned on the dynamometer with the hip in 90° in a sitting position for the knee flexion and extension measurements. Lateral movement of the knee will be prevented during full extension and flexion of the knee by a thigh strap on the operated leg. The physical therapist will help the patients to achieve proper positioning before each test. Concentric isokinetic knee flexion-extensions will be assessed at a preset velocity of 60º/sec, over a range of motion of 0º to 110º for both parameters. A fixed number of 10 flexion-extension repetitions will be completed by each patient. Instructions will be provided, and one trial repetition will then performed by all patients before the measurements were taken.

Sample size estimation will be performed using the extension peak torque as a primary effect variable. As there is no similar study with isokinetic measurements regarding this subject, a difference in mean values of 10 Nm and standard deviation of 12 Nm were assumed for each group. The group sample sizes of 31 and 31 achieved a power of 0.90 to detect a difference of 10 Nm between the two groups with estimated group standard deviations of 12 for each group and with a significance level (alpha) of 0.05 using a two-sided, two-sample test. Considering the loss to follow-up, two patients were added to each group. Thus, recruitment will end after 33 patients were assigned to either groups. All data will be calculated as mean and standard deviation. The Student's t-test will be used for statistical analysis of the patient data. Statistical calculations will be performed with SPSS 22.0 software (IBM SPSS Statistics for Windows, Version 22.0. Armonk, NY: IBM Corp.). A value of p< 0.05 will be considered statistically significant.

Study Design

Study Type:
Interventional
Actual Enrollment :
66 participants
Allocation:
Randomized
Intervention Model:
Parallel Assignment
Intervention Model Description:
Patients were randomized in a 1:1 ratio via computer-generated randomization using Microsoft Excel 2016 (Microsoft Corporation, Seattle, WA, USA) to be allocated in the UC insert or PS insert group before the TKA operation (Figure 1). Unblinded senior resident implemented the randomization. Patients and physiatrists performing isokinetic measurements were blinded to group allocation.Patients were randomized in a 1:1 ratio via computer-generated randomization using Microsoft Excel 2016 (Microsoft Corporation, Seattle, WA, USA) to be allocated in the UC insert or PS insert group before the TKA operation (Figure 1). Unblinded senior resident implemented the randomization. Patients and physiatrists performing isokinetic measurements were blinded to group allocation.
Masking:
Double (Participant, Investigator)
Masking Description:
Patients and physiatrists performing isokinetic measurements were blinded to group allocation.
Primary Purpose:
Treatment
Official Title:
Isokinetic Performance After Cruciate-substituting Ultra-congruent and Posterior Stabilized Total Knee Arthroplasties: A Randomized Controlled Trial
Actual Study Start Date :
Jan 19, 2017
Actual Primary Completion Date :
Mar 30, 2018
Actual Study Completion Date :
Mar 30, 2018

Arms and Interventions

Arm Intervention/Treatment
Experimental: Ultra-congruent insert group

Ultra-congruent inserts were used during total knee arthroplasty in patients randomized to this group.

Device: Vanguard® Knee System
Posterior cruciate ligament retention versus sacrificing is one of the main debates in total knee arthroplasty (TKA) and retention or sacrificing depends on the individual preference of the surgeon during the surgery. Whenever the surgeon decides to sacrifice the posterior cruciate ligament (PCL), another controversial question arises regarding the tibial insert type. While the posterior cruciate ligament-stabilized (PS) insert is widely used as the tibial insert in PCL-sacrificing TKA, it has some disadvantages such as increased polyethylene wear, additional bone resection, breakage of the post and patellar clunk syndrome. The ultra-congruent (UC) insert was designed to prevent bone loss in particular, and the other mentioned disadvantages of the conventional PS insert. However, patients with postoperative hyperextension have been seen to be associated with inferior clinical outcomes and knees become gradually more extended until two years after TKA using the UC insert.

Experimental: Posterior cruciate ligament-stabilized insert

Posterior cruciate ligament-stabilized inserts were used during total knee arthroplasty in patients randomized to this group.

Device: Vanguard® Knee System
Posterior cruciate ligament retention versus sacrificing is one of the main debates in total knee arthroplasty (TKA) and retention or sacrificing depends on the individual preference of the surgeon during the surgery. Whenever the surgeon decides to sacrifice the posterior cruciate ligament (PCL), another controversial question arises regarding the tibial insert type. While the posterior cruciate ligament-stabilized (PS) insert is widely used as the tibial insert in PCL-sacrificing TKA, it has some disadvantages such as increased polyethylene wear, additional bone resection, breakage of the post and patellar clunk syndrome. The ultra-congruent (UC) insert was designed to prevent bone loss in particular, and the other mentioned disadvantages of the conventional PS insert. However, patients with postoperative hyperextension have been seen to be associated with inferior clinical outcomes and knees become gradually more extended until two years after TKA using the UC insert.

Outcome Measures

Primary Outcome Measures

  1. Baseline Assessment of Peak Knee Extensor Torque Preoperatively [Preoperative]

    Peak extensor torque values of each patient was recorded for the operated knee preoperatively under the supervision of the same physiatrist. The Isokinetic measurements were performed using a Biodex System III Isokinetic Dynamometer, version 3.03 (Biodex Medical Inc.,Shirley, NY, USA). Concentric isokinetic knee flexion-extensions were assessed at a preset velocity of 60º/sec, over a range of motion of 0º to 110º for both parameters. A fixed number of 10 flexion-extension repetitions was completed by each patient. Torque was assessed in Newton-meters (N m).

  2. Baseline Assessment of Peak Knee Flexor Torque Preoperatively [Preoperative]

    Peak flexor torque values of each patient was recorded for the operated knee preoperatively under the supervision of the same physiatrist. The Isokinetic measurements were performed using a Biodex System III Isokinetic Dynamometer, version 3.03 (Biodex Medical Inc.,Shirley, NY, USA). Concentric isokinetic knee flexion-extensions were assessed at a preset velocity of 60º/sec, over a range of motion of 0º to 110º for both parameters. A fixed number of 10 flexion-extension repetitions was completed by each patient. Torque was assessed in Newton-meters (N m).

  3. Assessment of Peak Knee Extensor Torque at 3 Months Postoperatively [3 months postoperative]

    Peak extensor torque values of each patient was recorded for the operated knee at postoperative 3 months under the supervision of the same physiatrist. The Isokinetic measurements were performed using a Biodex System III Isokinetic Dynamometer, version 3.03 (Biodex Medical Inc.,Shirley, NY, USA). Concentric isokinetic knee flexion-extensions were assessed at a preset velocity of 60º/sec, over a range of motion of 0º to 110º for both parameters. A fixed number of 10 flexion-extension repetitions was completed by each patient. Torque was assessed in Newton-meters (N m).

  4. Assessment of Peak Knee Flexor Torque at 3 Months Postoperatively [3 months postoperative]

    Peak flexor torque values of each patient was recorded for the operated knee at postoperative 3 months under the supervision of the same physiatrist. The Isokinetic measurements were performed using a Biodex System III Isokinetic Dynamometer, version 3.03 (Biodex Medical Inc.,Shirley, NY, USA). Concentric isokinetic knee flexion-extensions were assessed at a preset velocity of 60º/sec, over a range of motion of 0º to 110º for both parameters. A fixed number of 10 flexion-extension repetitions was completed by each patient. Torque was assessed in Newton-meters (N m).

  5. Assessment of Peak Knee Extensor Torque at 6 Months Postoperatively [6 months postoperative]

    Peak extensor torque values of each patient was recorded for the operated knee at postoperative 6 months under the supervision of the same physiatrist. The Isokinetic measurements were performed using a Biodex System III Isokinetic Dynamometer, version 3.03 (Biodex Medical Inc.,Shirley, NY, USA). Concentric isokinetic knee flexion-extensions were assessed at a preset velocity of 60º/sec, over a range of motion of 0º to 110º for both parameters. A fixed number of 10 flexion-extension repetitions was completed by each patient. Torque was assessed in Newton-meters (N m).

  6. Assessment of Peak Knee Flexor Torque at 6 Months Postoperatively [6 months postoperative]

    Peak flexor torque values of each patient was recorded for the operated knee at postoperative 6 months under the supervision of the same physiatrist. The Isokinetic measurements were performed using a Biodex System III Isokinetic Dynamometer, version 3.03 (Biodex Medical Inc.,Shirley, NY, USA). Concentric isokinetic knee flexion-extensions were assessed at a preset velocity of 60º/sec, over a range of motion of 0º to 110º for both parameters. A fixed number of 10 flexion-extension repetitions was completed by each patient. Torque was assessed in Newton-meters (N m).

  7. Assessment of Peak Knee Extensor Torque at 12 Months Postoperatively [12 months postoperative]

    Peak extensor torque values of each patient was recorded for the operated knee at postoperative 12 months under the supervision of the same physiatrist. The Isokinetic measurements were performed using a Biodex System III Isokinetic Dynamometer, version 3.03 (Biodex Medical Inc.,Shirley, NY, USA). Concentric isokinetic knee flexion-extensions were assessed at a preset velocity of 60º/sec, over a range of motion of 0º to 110º for both parameters. A fixed number of 10 flexion-extension repetitions was completed by each patient. Torque was assessed in Newton-meters (N m).

  8. Assessment of Peak Knee Flexor Torque at 12 Months Postoperatively [12 months postoperative]

    Peak flexor torque values of each patient was recorded for the operated knee at postoperative 12 months under the supervision of the same physiatrist. The Isokinetic measurements were performed using a Biodex System III Isokinetic Dynamometer, version 3.03 (Biodex Medical Inc.,Shirley, NY, USA). Concentric isokinetic knee flexion-extensions were assessed at a preset velocity of 60º/sec, over a range of motion of 0º to 110º for both parameters. A fixed number of 10 flexion-extension repetitions was completed by each patient. Torque was assessed in Newton-meters (N m).

Secondary Outcome Measures

  1. Baseline Knee Society Clinical Score Evaluation [Preoperative]

    The Knee Society Score were applied preoperatively. The Knee Society Score (KSS) is comprised to two sections (each worth 100 points) for a maximum 200 points. The KSS Scale ranges from 0 to 200. One section is the Knee Society Clinical Score (KSCS) - points are given for pain, motion, and stability and points are deducted for flexion contracture, extension lag, and misalignment. The KSCS ranges from 0 to 100. The other section is the Knee Society Functional Score (KSFS) - points are assigned for walking distances and climbing stairs and points are deducted for use of walking aids. The KSFS ranges from 0 to 100. For each section, a score of 80-100 was considered excellent; 70-79, good; 60-69, fair; < 60, poor.

  2. Knee Society Clinical Score Evaluation at 3 Months Postoperatively [3 months postoperative]

    The Knee Society Clinical Score were applied at postoperative 3 months. The Knee Society Score (KSS) is comprised to two sections (each worth 100 points) for a maximum 200 points. The KSS Scale ranges from 0 to 200. One section is the Knee Society Clinical Score (KSCS) - points are given for pain, motion, and stability and points are deducted for flexion contracture, extension lag, and misalignment. The KSCS ranges from 0 to 100. The other section is the Knee Society Functional Score (KSFS) - points are assigned for walking distances and climbing stairs and points are deducted for use of walking aids. The KSFS ranges from 0 to 100. For each section, a score of 80-100 was considered excellent; 70-79, good; 60-69, fair; < 60, poor.

  3. Knee Society Clinical Score Evaluation at 6 Months Postoperatively [6 months postoperative]

    The Knee Society Clinical Score were applied at postoperative 6 months. The Knee Society Score (KSS) is comprised to two sections (each worth 100 points) for a maximum 200 points. The KSS Scale ranges from 0 to 200. One section is the Knee Society Clinical Score (KSCS) - points are given for pain, motion, and stability and points are deducted for flexion contracture, extension lag, and misalignment. The KSCS ranges from 0 to 100. The other section is the Knee Society Functional Score (KSFS) - points are assigned for walking distances and climbing stairs and points are deducted for use of walking aids. The KSFS ranges from 0 to 100. For each section, a score of 80-100 was considered excellent; 70-79, good; 60-69, fair; < 60, poor.

  4. Knee Society Clinical Score Evaluation at 12 Months Postoperatively [12 months postoperative]

    The Knee Society Clinical Score were applied at postoperative 12 months. The Knee Society Score (KSS) is comprised to two sections (each worth 100 points) for a maximum 200 points. The KSS Scale ranges from 0 to 200. One section is the Knee Society Clinical Score (KSCS) - points are given for pain, motion, and stability and points are deducted for flexion contracture, extension lag, and misalignment. The KSCS ranges from 0 to 100. The other section is the Knee Society Functional Score (KSFS) - points are assigned for walking distances and climbing stairs and points are deducted for use of walking aids. The KSFS ranges from 0 to 100. For each section, a score of 80-100 was considered excellent; 70-79, good; 60-69, fair; < 60, poor.

  5. Baseline Knee Society Functional Score Evaluation [Preoperative]

    The Knee Society Functional Score were applied preoperatively. The Knee Society Score (KSS) is comprised to two sections (each worth 100 points) for a maximum 200 points. The KSS Scale ranges from 0 to 200. One section is the Knee Society Clinical Score (KSCS) - points are given for pain, motion, and stability and points are deducted for flexion contracture, extension lag, and misalignment. The KSCS ranges from 0 to 100. The other section is the Knee Society Functional Score (KSFS) - points are assigned for walking distances and climbing stairs and points are deducted for use of walking aids. The KSFS ranges from 0 to 100. For each section, a score of 80-100 was considered excellent; 70-79, good; 60-69, fair; < 60, poor.

  6. Knee Society Functional Score Evaluation at 3 Months Postoperatively [3 months postoperative]

    The Knee Society Functional Score were applied at postoperative 3 months. The Knee Society Score (KSS) is comprised to two sections (each worth 100 points) for a maximum 200 points. The KSS Scale ranges from 0 to 200. One section is the Knee Society Clinical Score (KSCS) - points are given for pain, motion, and stability and points are deducted for flexion contracture, extension lag, and misalignment. The KSCS ranges from 0 to 100. The other section is the Knee Society Functional Score (KSFS) - points are assigned for walking distances and climbing stairs and points are deducted for use of walking aids. The KSFS ranges from 0 to 100. For each section, a score of 80-100 was considered excellent; 70-79, good; 60-69, fair; < 60, poor.

  7. Knee Society Functional Score Evaluation at 6 Months Postoperatively [6 months postoperative]

    The Knee Society Functional Score were applied at postoperative 6 months. The Knee Society Score (KSS) is comprised to two sections (each worth 100 points) for a maximum 200 points. The KSS Scale ranges from 0 to 200. One section is the Knee Society Clinical Score (KSCS) - points are given for pain, motion, and stability and points are deducted for flexion contracture, extension lag, and misalignment. The KSCS ranges from 0 to 100. The other section is the Knee Society Functional Score (KSFS) - points are assigned for walking distances and climbing stairs and points are deducted for use of walking aids. The KSFS ranges from 0 to 100. For each section, a score of 80-100 was considered excellent; 70-79, good; 60-69, fair; < 60, poor.

  8. Knee Society Functional Score Evaluation at 12 Months Postoperatively [12 months postoperative]

    The Knee Society Functional Score were applied at postoperative 12 months. The Knee Society Score (KSS) is comprised to two sections (each worth 100 points) for a maximum 200 points. The KSS Scale ranges from 0 to 200. One section is the Knee Society Clinical Score (KSCS) - points are given for pain, motion, and stability and points are deducted for flexion contracture, extension lag, and misalignment. The KSCS ranges from 0 to 100. The other section is the Knee Society Functional Score (KSFS) - points are assigned for walking distances and climbing stairs and points are deducted for use of walking aids. The KSFS ranges from 0 to 100. For each section, a score of 80-100 was considered excellent; 70-79, good; 60-69, fair; < 60, poor.

Eligibility Criteria

Criteria

Ages Eligible for Study:
55 Years to 80 Years
Sexes Eligible for Study:
All
Accepts Healthy Volunteers:
No
Inclusion Criteria:
  • Age between 55 and 80 years

  • Scheduled to undergo unilateral TKA on for primary knee osteoarthritis

Exclusion Criteria:
  • Rheumatological joint diseases

  • Previous knee surgery

  • Neuromuscular diseases

  • Bilateral TKA

  • Insufficiency of collateral ligaments

Contacts and Locations

Locations

Site City State Country Postal Code
1 Aksaray University Training and Research Hospital Aksaray Turkey 68200

Sponsors and Collaborators

  • Aksaray University Training and Research Hospital

Investigators

  • Principal Investigator: Deniz Çankaya, MD, Aksaray University Training and Research Hospital

Study Documents (Full-Text)

More Information

Publications

None provided.
Responsible Party:
Erdem A Sezgin, Co-Principal Investigator, Aksaray University Training and Research Hospital
ClinicalTrials.gov Identifier:
NCT04419311
Other Study ID Numbers:
  • 1585/2017
First Posted:
Jun 5, 2020
Last Update Posted:
Mar 16, 2021
Last Verified:
Feb 1, 2021
Studies a U.S. FDA-regulated Drug Product:
No
Studies a U.S. FDA-regulated Device Product:
Yes
Product Manufactured in and Exported from the U.S.:
Yes
Keywords provided by Erdem A Sezgin, Co-Principal Investigator, Aksaray University Training and Research Hospital
Additional relevant MeSH terms:

Study Results

Participant Flow

Recruitment Details
Pre-assignment Detail
Arm/Group Title Ultra-congruent Insert Group Posterior Cruciate Ligament-stabilized Insert
Arm/Group Description Ultra-congruent inserts were used during total knee arthroplasty in patients randomized to this group. Posterior cruciate ligament-stabilized inserts were used during total knee arthroplasty in patients randomized to this group.
Period Title: Overall Study
STARTED 33 33
COMPLETED 32 33
NOT COMPLETED 1 0

Baseline Characteristics

Arm/Group Title Ultra-congruent Insert Group Posterior Cruciate Ligament-stabilized Insert Total
Arm/Group Description Ultra-congruent inserts were used during total knee arthroplasty in patients randomized to this group. Posterior cruciate ligament-stabilized inserts were used during total knee arthroplasty in patients randomized to this group. Total of all reporting groups
Overall Participants 33 33 66
Age (years) [Mean (Standard Deviation) ]
Mean (Standard Deviation) [years]
69.18
(8.55)
67.67
(8.12)
68.42
(8.31)
Sex: Female, Male (Count of Participants)
Female
29
87.9%
26
78.8%
55
83.3%
Male
4
12.1%
7
21.2%
11
16.7%
Race and Ethnicity Not Collected (Count of Participants)
Count of Participants [Participants]
0
0%
Region of Enrollment (participants) [Number]
Turkey
33
100%
33
100%
66
100%

Outcome Measures

1. Primary Outcome
Title Baseline Assessment of Peak Knee Extensor Torque Preoperatively
Description Peak extensor torque values of each patient was recorded for the operated knee preoperatively under the supervision of the same physiatrist. The Isokinetic measurements were performed using a Biodex System III Isokinetic Dynamometer, version 3.03 (Biodex Medical Inc.,Shirley, NY, USA). Concentric isokinetic knee flexion-extensions were assessed at a preset velocity of 60º/sec, over a range of motion of 0º to 110º for both parameters. A fixed number of 10 flexion-extension repetitions was completed by each patient. Torque was assessed in Newton-meters (N m).
Time Frame Preoperative

Outcome Measure Data

Analysis Population Description
[Not Specified]
Arm/Group Title Ultra-congruent Insert Group Posterior Cruciate Ligament-stabilized Insert
Arm/Group Description Ultra-congruent inserts were used during total knee arthroplasty in patients randomized to this group. Posterior cruciate ligament-stabilized inserts were used during total knee arthroplasty in patients randomized to this group.
Measure Participants 32 33
Mean (Standard Deviation) [Newton meters]
55.56
(8.65)
58.33
(8.96)
2. Primary Outcome
Title Baseline Assessment of Peak Knee Flexor Torque Preoperatively
Description Peak flexor torque values of each patient was recorded for the operated knee preoperatively under the supervision of the same physiatrist. The Isokinetic measurements were performed using a Biodex System III Isokinetic Dynamometer, version 3.03 (Biodex Medical Inc.,Shirley, NY, USA). Concentric isokinetic knee flexion-extensions were assessed at a preset velocity of 60º/sec, over a range of motion of 0º to 110º for both parameters. A fixed number of 10 flexion-extension repetitions was completed by each patient. Torque was assessed in Newton-meters (N m).
Time Frame Preoperative

Outcome Measure Data

Analysis Population Description
[Not Specified]
Arm/Group Title Ultra-congruent Insert Group Posterior Cruciate Ligament-stabilized Insert
Arm/Group Description Ultra-congruent inserts were used during total knee arthroplasty in patients randomized to this group. Posterior cruciate ligament-stabilized inserts were used during total knee arthroplasty in patients randomized to this group.
Measure Participants 32 33
Mean (Standard Deviation) [Newton meters]
37.84
(9.56)
40.3
(7.79)
3. Primary Outcome
Title Assessment of Peak Knee Extensor Torque at 3 Months Postoperatively
Description Peak extensor torque values of each patient was recorded for the operated knee at postoperative 3 months under the supervision of the same physiatrist. The Isokinetic measurements were performed using a Biodex System III Isokinetic Dynamometer, version 3.03 (Biodex Medical Inc.,Shirley, NY, USA). Concentric isokinetic knee flexion-extensions were assessed at a preset velocity of 60º/sec, over a range of motion of 0º to 110º for both parameters. A fixed number of 10 flexion-extension repetitions was completed by each patient. Torque was assessed in Newton-meters (N m).
Time Frame 3 months postoperative

Outcome Measure Data

Analysis Population Description
[Not Specified]
Arm/Group Title Ultra-congruent Insert Group Posterior Cruciate Ligament-stabilized Insert
Arm/Group Description Ultra-congruent inserts were used during total knee arthroplasty in patients randomized to this group. Posterior cruciate ligament-stabilized inserts were used during total knee arthroplasty in patients randomized to this group.
Measure Participants 32 33
Mean (Standard Deviation) [Newton meters]
44.38
(7.37)
45.85
(7.46)
4. Primary Outcome
Title Assessment of Peak Knee Flexor Torque at 3 Months Postoperatively
Description Peak flexor torque values of each patient was recorded for the operated knee at postoperative 3 months under the supervision of the same physiatrist. The Isokinetic measurements were performed using a Biodex System III Isokinetic Dynamometer, version 3.03 (Biodex Medical Inc.,Shirley, NY, USA). Concentric isokinetic knee flexion-extensions were assessed at a preset velocity of 60º/sec, over a range of motion of 0º to 110º for both parameters. A fixed number of 10 flexion-extension repetitions was completed by each patient. Torque was assessed in Newton-meters (N m).
Time Frame 3 months postoperative

Outcome Measure Data

Analysis Population Description
[Not Specified]
Arm/Group Title Ultra-congruent Insert Group Posterior Cruciate Ligament-stabilized Insert
Arm/Group Description Ultra-congruent inserts were used during total knee arthroplasty in patients randomized to this group. Vanguard® Knee System: Posterior cruciate ligament retention versus sacrificing is one of the main debates in total knee arthroplasty (TKA) and retention or sacrificing depends on the individual preference of the surgeon during the surgery. Whenever the surgeon decides to sacrifice the posterior cruciate ligament (PCL), another controversial question arises regarding the tibial insert type. While the posterior cruciate ligament-stabilized (PS) insert is widely used as the tibial insert in PCL-sacrificing TKA, it has some disadvantages such as increased polyethylene wear, additional bone resection, breakage of the post and patellar clunk syndrome. The ultra-congruent (UC) insert was designed to prevent bone loss in particular, and the other mentioned disadvantages of the conventional PS insert. However, patients with postoperative hyperextension have been seen to be associated with inferior clinical outcomes and knees become gradually more extended until two years after TKA using the UC insert. Posterior cruciate ligament-stabilized inserts were used during total knee arthroplasty in patients randomized to this group. Vanguard® Knee System: Posterior cruciate ligament retention versus sacrificing is one of the main debates in total knee arthroplasty (TKA) and retention or sacrificing depends on the individual preference of the surgeon during the surgery. Whenever the surgeon decides to sacrifice the posterior cruciate ligament (PCL), another controversial question arises regarding the tibial insert type. While the posterior cruciate ligament-stabilized (PS) insert is widely used as the tibial insert in PCL-sacrificing TKA, it has some disadvantages such as increased polyethylene wear, additional bone resection, breakage of the post and patellar clunk syndrome. The ultra-congruent (UC) insert was designed to prevent bone loss in particular, and the other mentioned disadvantages of the conventional PS insert. However, patients with postoperative hyperextension have been seen to be associated with inferior clinical outcomes and knees become gradually more extended until two years after TKA using the UC insert.
Measure Participants 32 33
Mean (Standard Deviation) [Newton meters]
31.91
(10.11)
29.39
(7.16)
5. Primary Outcome
Title Assessment of Peak Knee Extensor Torque at 6 Months Postoperatively
Description Peak extensor torque values of each patient was recorded for the operated knee at postoperative 6 months under the supervision of the same physiatrist. The Isokinetic measurements were performed using a Biodex System III Isokinetic Dynamometer, version 3.03 (Biodex Medical Inc.,Shirley, NY, USA). Concentric isokinetic knee flexion-extensions were assessed at a preset velocity of 60º/sec, over a range of motion of 0º to 110º for both parameters. A fixed number of 10 flexion-extension repetitions was completed by each patient. Torque was assessed in Newton-meters (N m).
Time Frame 6 months postoperative

Outcome Measure Data

Analysis Population Description
[Not Specified]
Arm/Group Title Ultra-congruent Insert Group Posterior Cruciate Ligament-stabilized Insert
Arm/Group Description Ultra-congruent inserts were used during total knee arthroplasty in patients randomized to this group. Vanguard® Knee System: Posterior cruciate ligament retention versus sacrificing is one of the main debates in total knee arthroplasty (TKA) and retention or sacrificing depends on the individual preference of the surgeon during the surgery. Whenever the surgeon decides to sacrifice the posterior cruciate ligament (PCL), another controversial question arises regarding the tibial insert type. While the posterior cruciate ligament-stabilized (PS) insert is widely used as the tibial insert in PCL-sacrificing TKA, it has some disadvantages such as increased polyethylene wear, additional bone resection, breakage of the post and patellar clunk syndrome. The ultra-congruent (UC) insert was designed to prevent bone loss in particular, and the other mentioned disadvantages of the conventional PS insert. However, patients with postoperative hyperextension have been seen to be associated with inferior clinical outcomes and knees become gradually more extended until two years after TKA using the UC insert. Posterior cruciate ligament-stabilized inserts were used during total knee arthroplasty in patients randomized to this group. Vanguard® Knee System: Posterior cruciate ligament retention versus sacrificing is one of the main debates in total knee arthroplasty (TKA) and retention or sacrificing depends on the individual preference of the surgeon during the surgery. Whenever the surgeon decides to sacrifice the posterior cruciate ligament (PCL), another controversial question arises regarding the tibial insert type. While the posterior cruciate ligament-stabilized (PS) insert is widely used as the tibial insert in PCL-sacrificing TKA, it has some disadvantages such as increased polyethylene wear, additional bone resection, breakage of the post and patellar clunk syndrome. The ultra-congruent (UC) insert was designed to prevent bone loss in particular, and the other mentioned disadvantages of the conventional PS insert. However, patients with postoperative hyperextension have been seen to be associated with inferior clinical outcomes and knees become gradually more extended until two years after TKA using the UC insert.
Measure Participants 32 33
Mean (Standard Deviation) [Newton meters]
53.53
(7.95)
52.64
(6.83)
6. Primary Outcome
Title Assessment of Peak Knee Flexor Torque at 6 Months Postoperatively
Description Peak flexor torque values of each patient was recorded for the operated knee at postoperative 6 months under the supervision of the same physiatrist. The Isokinetic measurements were performed using a Biodex System III Isokinetic Dynamometer, version 3.03 (Biodex Medical Inc.,Shirley, NY, USA). Concentric isokinetic knee flexion-extensions were assessed at a preset velocity of 60º/sec, over a range of motion of 0º to 110º for both parameters. A fixed number of 10 flexion-extension repetitions was completed by each patient. Torque was assessed in Newton-meters (N m).
Time Frame 6 months postoperative

Outcome Measure Data

Analysis Population Description
[Not Specified]
Arm/Group Title Ultra-congruent Insert Group Posterior Cruciate Ligament-stabilized Insert
Arm/Group Description Ultra-congruent inserts were used during total knee arthroplasty in patients randomized to this group. Vanguard® Knee System: Posterior cruciate ligament retention versus sacrificing is one of the main debates in total knee arthroplasty (TKA) and retention or sacrificing depends on the individual preference of the surgeon during the surgery. Whenever the surgeon decides to sacrifice the posterior cruciate ligament (PCL), another controversial question arises regarding the tibial insert type. While the posterior cruciate ligament-stabilized (PS) insert is widely used as the tibial insert in PCL-sacrificing TKA, it has some disadvantages such as increased polyethylene wear, additional bone resection, breakage of the post and patellar clunk syndrome. The ultra-congruent (UC) insert was designed to prevent bone loss in particular, and the other mentioned disadvantages of the conventional PS insert. However, patients with postoperative hyperextension have been seen to be associated with inferior clinical outcomes and knees become gradually more extended until two years after TKA using the UC insert. Posterior cruciate ligament-stabilized inserts were used during total knee arthroplasty in patients randomized to this group. Vanguard® Knee System: Posterior cruciate ligament retention versus sacrificing is one of the main debates in total knee arthroplasty (TKA) and retention or sacrificing depends on the individual preference of the surgeon during the surgery. Whenever the surgeon decides to sacrifice the posterior cruciate ligament (PCL), another controversial question arises regarding the tibial insert type. While the posterior cruciate ligament-stabilized (PS) insert is widely used as the tibial insert in PCL-sacrificing TKA, it has some disadvantages such as increased polyethylene wear, additional bone resection, breakage of the post and patellar clunk syndrome. The ultra-congruent (UC) insert was designed to prevent bone loss in particular, and the other mentioned disadvantages of the conventional PS insert. However, patients with postoperative hyperextension have been seen to be associated with inferior clinical outcomes and knees become gradually more extended until two years after TKA using the UC insert.
Measure Participants 32 33
Mean (Standard Deviation) [Newton meters]
36.19
(9.86)
35.27
(7.35)
7. Primary Outcome
Title Assessment of Peak Knee Extensor Torque at 12 Months Postoperatively
Description Peak extensor torque values of each patient was recorded for the operated knee at postoperative 12 months under the supervision of the same physiatrist. The Isokinetic measurements were performed using a Biodex System III Isokinetic Dynamometer, version 3.03 (Biodex Medical Inc.,Shirley, NY, USA). Concentric isokinetic knee flexion-extensions were assessed at a preset velocity of 60º/sec, over a range of motion of 0º to 110º for both parameters. A fixed number of 10 flexion-extension repetitions was completed by each patient. Torque was assessed in Newton-meters (N m).
Time Frame 12 months postoperative

Outcome Measure Data

Analysis Population Description
[Not Specified]
Arm/Group Title Ultra-congruent Insert Group Posterior Cruciate Ligament-stabilized Insert
Arm/Group Description Ultra-congruent inserts were used during total knee arthroplasty in patients randomized to this group. Vanguard® Knee System: Posterior cruciate ligament retention versus sacrificing is one of the main debates in total knee arthroplasty (TKA) and retention or sacrificing depends on the individual preference of the surgeon during the surgery. Whenever the surgeon decides to sacrifice the posterior cruciate ligament (PCL), another controversial question arises regarding the tibial insert type. While the posterior cruciate ligament-stabilized (PS) insert is widely used as the tibial insert in PCL-sacrificing TKA, it has some disadvantages such as increased polyethylene wear, additional bone resection, breakage of the post and patellar clunk syndrome. The ultra-congruent (UC) insert was designed to prevent bone loss in particular, and the other mentioned disadvantages of the conventional PS insert. However, patients with postoperative hyperextension have been seen to be associated with inferior clinical outcomes and knees become gradually more extended until two years after TKA using the UC insert. Posterior cruciate ligament-stabilized inserts were used during total knee arthroplasty in patients randomized to this group. Vanguard® Knee System: Posterior cruciate ligament retention versus sacrificing is one of the main debates in total knee arthroplasty (TKA) and retention or sacrificing depends on the individual preference of the surgeon during the surgery. Whenever the surgeon decides to sacrifice the posterior cruciate ligament (PCL), another controversial question arises regarding the tibial insert type. While the posterior cruciate ligament-stabilized (PS) insert is widely used as the tibial insert in PCL-sacrificing TKA, it has some disadvantages such as increased polyethylene wear, additional bone resection, breakage of the post and patellar clunk syndrome. The ultra-congruent (UC) insert was designed to prevent bone loss in particular, and the other mentioned disadvantages of the conventional PS insert. However, patients with postoperative hyperextension have been seen to be associated with inferior clinical outcomes and knees become gradually more extended until two years after TKA using the UC insert.
Measure Participants 32 33
Mean (Standard Deviation) [Newton meters]
59.34
(8.97)
61.52
(9.88)
8. Primary Outcome
Title Assessment of Peak Knee Flexor Torque at 12 Months Postoperatively
Description Peak flexor torque values of each patient was recorded for the operated knee at postoperative 12 months under the supervision of the same physiatrist. The Isokinetic measurements were performed using a Biodex System III Isokinetic Dynamometer, version 3.03 (Biodex Medical Inc.,Shirley, NY, USA). Concentric isokinetic knee flexion-extensions were assessed at a preset velocity of 60º/sec, over a range of motion of 0º to 110º for both parameters. A fixed number of 10 flexion-extension repetitions was completed by each patient. Torque was assessed in Newton-meters (N m).
Time Frame 12 months postoperative

Outcome Measure Data

Analysis Population Description
[Not Specified]
Arm/Group Title Ultra-congruent Insert Group Posterior Cruciate Ligament-stabilized Insert
Arm/Group Description Ultra-congruent inserts were used during total knee arthroplasty in patients randomized to this group. Vanguard® Knee System: Posterior cruciate ligament retention versus sacrificing is one of the main debates in total knee arthroplasty (TKA) and retention or sacrificing depends on the individual preference of the surgeon during the surgery. Whenever the surgeon decides to sacrifice the posterior cruciate ligament (PCL), another controversial question arises regarding the tibial insert type. While the posterior cruciate ligament-stabilized (PS) insert is widely used as the tibial insert in PCL-sacrificing TKA, it has some disadvantages such as increased polyethylene wear, additional bone resection, breakage of the post and patellar clunk syndrome. The ultra-congruent (UC) insert was designed to prevent bone loss in particular, and the other mentioned disadvantages of the conventional PS insert. However, patients with postoperative hyperextension have been seen to be associated with inferior clinical outcomes and knees become gradually more extended until two years after TKA using the UC insert. Posterior cruciate ligament-stabilized inserts were used during total knee arthroplasty in patients randomized to this group. Vanguard® Knee System: Posterior cruciate ligament retention versus sacrificing is one of the main debates in total knee arthroplasty (TKA) and retention or sacrificing depends on the individual preference of the surgeon during the surgery. Whenever the surgeon decides to sacrifice the posterior cruciate ligament (PCL), another controversial question arises regarding the tibial insert type. While the posterior cruciate ligament-stabilized (PS) insert is widely used as the tibial insert in PCL-sacrificing TKA, it has some disadvantages such as increased polyethylene wear, additional bone resection, breakage of the post and patellar clunk syndrome. The ultra-congruent (UC) insert was designed to prevent bone loss in particular, and the other mentioned disadvantages of the conventional PS insert. However, patients with postoperative hyperextension have been seen to be associated with inferior clinical outcomes and knees become gradually more extended until two years after TKA using the UC insert.
Measure Participants 32 33
Mean (Standard Deviation) [Newton meters]
39.94
(9.96)
41.18
(7.63)
9. Secondary Outcome
Title Baseline Knee Society Clinical Score Evaluation
Description The Knee Society Score were applied preoperatively. The Knee Society Score (KSS) is comprised to two sections (each worth 100 points) for a maximum 200 points. The KSS Scale ranges from 0 to 200. One section is the Knee Society Clinical Score (KSCS) - points are given for pain, motion, and stability and points are deducted for flexion contracture, extension lag, and misalignment. The KSCS ranges from 0 to 100. The other section is the Knee Society Functional Score (KSFS) - points are assigned for walking distances and climbing stairs and points are deducted for use of walking aids. The KSFS ranges from 0 to 100. For each section, a score of 80-100 was considered excellent; 70-79, good; 60-69, fair; < 60, poor.
Time Frame Preoperative

Outcome Measure Data

Analysis Population Description
[Not Specified]
Arm/Group Title Ultra-congruent Insert Group Posterior Cruciate Ligament-stabilized Insert
Arm/Group Description Ultra-congruent inserts were used during total knee arthroplasty in patients randomized to this group. Vanguard® Knee System: Posterior cruciate ligament retention versus sacrificing is one of the main debates in total knee arthroplasty (TKA) and retention or sacrificing depends on the individual preference of the surgeon during the surgery. Whenever the surgeon decides to sacrifice the posterior cruciate ligament (PCL), another controversial question arises regarding the tibial insert type. While the posterior cruciate ligament-stabilized (PS) insert is widely used as the tibial insert in PCL-sacrificing TKA, it has some disadvantages such as increased polyethylene wear, additional bone resection, breakage of the post and patellar clunk syndrome. The ultra-congruent (UC) insert was designed to prevent bone loss in particular, and the other mentioned disadvantages of the conventional PS insert. However, patients with postoperative hyperextension have been seen to be associated with inferior clinical outcomes and knees become gradually more extended until two years after TKA using the UC insert. Posterior cruciate ligament-stabilized inserts were used during total knee arthroplasty in patients randomized to this group. Vanguard® Knee System: Posterior cruciate ligament retention versus sacrificing is one of the main debates in total knee arthroplasty (TKA) and retention or sacrificing depends on the individual preference of the surgeon during the surgery. Whenever the surgeon decides to sacrifice the posterior cruciate ligament (PCL), another controversial question arises regarding the tibial insert type. While the posterior cruciate ligament-stabilized (PS) insert is widely used as the tibial insert in PCL-sacrificing TKA, it has some disadvantages such as increased polyethylene wear, additional bone resection, breakage of the post and patellar clunk syndrome. The ultra-congruent (UC) insert was designed to prevent bone loss in particular, and the other mentioned disadvantages of the conventional PS insert. However, patients with postoperative hyperextension have been seen to be associated with inferior clinical outcomes and knees become gradually more extended until two years after TKA using the UC insert.
Measure Participants 32 33
Mean (Standard Deviation) [score on a scale]
47.69
(12.6)
49.15
(11.68)
10. Secondary Outcome
Title Knee Society Clinical Score Evaluation at 3 Months Postoperatively
Description The Knee Society Clinical Score were applied at postoperative 3 months. The Knee Society Score (KSS) is comprised to two sections (each worth 100 points) for a maximum 200 points. The KSS Scale ranges from 0 to 200. One section is the Knee Society Clinical Score (KSCS) - points are given for pain, motion, and stability and points are deducted for flexion contracture, extension lag, and misalignment. The KSCS ranges from 0 to 100. The other section is the Knee Society Functional Score (KSFS) - points are assigned for walking distances and climbing stairs and points are deducted for use of walking aids. The KSFS ranges from 0 to 100. For each section, a score of 80-100 was considered excellent; 70-79, good; 60-69, fair; < 60, poor.
Time Frame 3 months postoperative

Outcome Measure Data

Analysis Population Description
[Not Specified]
Arm/Group Title Ultra-congruent Insert Group Posterior Cruciate Ligament-stabilized Insert
Arm/Group Description Ultra-congruent inserts were used during total knee arthroplasty in patients randomized to this group. Vanguard® Knee System: Posterior cruciate ligament retention versus sacrificing is one of the main debates in total knee arthroplasty (TKA) and retention or sacrificing depends on the individual preference of the surgeon during the surgery. Whenever the surgeon decides to sacrifice the posterior cruciate ligament (PCL), another controversial question arises regarding the tibial insert type. While the posterior cruciate ligament-stabilized (PS) insert is widely used as the tibial insert in PCL-sacrificing TKA, it has some disadvantages such as increased polyethylene wear, additional bone resection, breakage of the post and patellar clunk syndrome. The ultra-congruent (UC) insert was designed to prevent bone loss in particular, and the other mentioned disadvantages of the conventional PS insert. However, patients with postoperative hyperextension have been seen to be associated with inferior clinical outcomes and knees become gradually more extended until two years after TKA using the UC insert. Posterior cruciate ligament-stabilized inserts were used during total knee arthroplasty in patients randomized to this group. Vanguard® Knee System: Posterior cruciate ligament retention versus sacrificing is one of the main debates in total knee arthroplasty (TKA) and retention or sacrificing depends on the individual preference of the surgeon during the surgery. Whenever the surgeon decides to sacrifice the posterior cruciate ligament (PCL), another controversial question arises regarding the tibial insert type. While the posterior cruciate ligament-stabilized (PS) insert is widely used as the tibial insert in PCL-sacrificing TKA, it has some disadvantages such as increased polyethylene wear, additional bone resection, breakage of the post and patellar clunk syndrome. The ultra-congruent (UC) insert was designed to prevent bone loss in particular, and the other mentioned disadvantages of the conventional PS insert. However, patients with postoperative hyperextension have been seen to be associated with inferior clinical outcomes and knees become gradually more extended until two years after TKA using the UC insert.
Measure Participants 32 33
Mean (Standard Deviation) [score on a scale]
78.81
(11.73)
79.48
(10.80)
11. Secondary Outcome
Title Knee Society Clinical Score Evaluation at 6 Months Postoperatively
Description The Knee Society Clinical Score were applied at postoperative 6 months. The Knee Society Score (KSS) is comprised to two sections (each worth 100 points) for a maximum 200 points. The KSS Scale ranges from 0 to 200. One section is the Knee Society Clinical Score (KSCS) - points are given for pain, motion, and stability and points are deducted for flexion contracture, extension lag, and misalignment. The KSCS ranges from 0 to 100. The other section is the Knee Society Functional Score (KSFS) - points are assigned for walking distances and climbing stairs and points are deducted for use of walking aids. The KSFS ranges from 0 to 100. For each section, a score of 80-100 was considered excellent; 70-79, good; 60-69, fair; < 60, poor.
Time Frame 6 months postoperative

Outcome Measure Data

Analysis Population Description
[Not Specified]
Arm/Group Title Ultra-congruent Insert Group Posterior Cruciate Ligament-stabilized Insert
Arm/Group Description Ultra-congruent inserts were used during total knee arthroplasty in patients randomized to this group. Vanguard® Knee System: Posterior cruciate ligament retention versus sacrificing is one of the main debates in total knee arthroplasty (TKA) and retention or sacrificing depends on the individual preference of the surgeon during the surgery. Whenever the surgeon decides to sacrifice the posterior cruciate ligament (PCL), another controversial question arises regarding the tibial insert type. While the posterior cruciate ligament-stabilized (PS) insert is widely used as the tibial insert in PCL-sacrificing TKA, it has some disadvantages such as increased polyethylene wear, additional bone resection, breakage of the post and patellar clunk syndrome. The ultra-congruent (UC) insert was designed to prevent bone loss in particular, and the other mentioned disadvantages of the conventional PS insert. However, patients with postoperative hyperextension have been seen to be associated with inferior clinical outcomes and knees become gradually more extended until two years after TKA using the UC insert. Posterior cruciate ligament-stabilized inserts were used during total knee arthroplasty in patients randomized to this group. Vanguard® Knee System: Posterior cruciate ligament retention versus sacrificing is one of the main debates in total knee arthroplasty (TKA) and retention or sacrificing depends on the individual preference of the surgeon during the surgery. Whenever the surgeon decides to sacrifice the posterior cruciate ligament (PCL), another controversial question arises regarding the tibial insert type. While the posterior cruciate ligament-stabilized (PS) insert is widely used as the tibial insert in PCL-sacrificing TKA, it has some disadvantages such as increased polyethylene wear, additional bone resection, breakage of the post and patellar clunk syndrome. The ultra-congruent (UC) insert was designed to prevent bone loss in particular, and the other mentioned disadvantages of the conventional PS insert. However, patients with postoperative hyperextension have been seen to be associated with inferior clinical outcomes and knees become gradually more extended until two years after TKA using the UC insert.
Measure Participants 32 33
Mean (Standard Deviation) [score on a scale]
86.81
(5.64)
88.39
(6.29)
12. Secondary Outcome
Title Knee Society Clinical Score Evaluation at 12 Months Postoperatively
Description The Knee Society Clinical Score were applied at postoperative 12 months. The Knee Society Score (KSS) is comprised to two sections (each worth 100 points) for a maximum 200 points. The KSS Scale ranges from 0 to 200. One section is the Knee Society Clinical Score (KSCS) - points are given for pain, motion, and stability and points are deducted for flexion contracture, extension lag, and misalignment. The KSCS ranges from 0 to 100. The other section is the Knee Society Functional Score (KSFS) - points are assigned for walking distances and climbing stairs and points are deducted for use of walking aids. The KSFS ranges from 0 to 100. For each section, a score of 80-100 was considered excellent; 70-79, good; 60-69, fair; < 60, poor.
Time Frame 12 months postoperative

Outcome Measure Data

Analysis Population Description
[Not Specified]
Arm/Group Title Ultra-congruent Insert Group Posterior Cruciate Ligament-stabilized Insert
Arm/Group Description Ultra-congruent inserts were used during total knee arthroplasty in patients randomized to this group. Vanguard® Knee System: Posterior cruciate ligament retention versus sacrificing is one of the main debates in total knee arthroplasty (TKA) and retention or sacrificing depends on the individual preference of the surgeon during the surgery. Whenever the surgeon decides to sacrifice the posterior cruciate ligament (PCL), another controversial question arises regarding the tibial insert type. While the posterior cruciate ligament-stabilized (PS) insert is widely used as the tibial insert in PCL-sacrificing TKA, it has some disadvantages such as increased polyethylene wear, additional bone resection, breakage of the post and patellar clunk syndrome. The ultra-congruent (UC) insert was designed to prevent bone loss in particular, and the other mentioned disadvantages of the conventional PS insert. However, patients with postoperative hyperextension have been seen to be associated with inferior clinical outcomes and knees become gradually more extended until two years after TKA using the UC insert. Posterior cruciate ligament-stabilized inserts were used during total knee arthroplasty in patients randomized to this group. Vanguard® Knee System: Posterior cruciate ligament retention versus sacrificing is one of the main debates in total knee arthroplasty (TKA) and retention or sacrificing depends on the individual preference of the surgeon during the surgery. Whenever the surgeon decides to sacrifice the posterior cruciate ligament (PCL), another controversial question arises regarding the tibial insert type. While the posterior cruciate ligament-stabilized (PS) insert is widely used as the tibial insert in PCL-sacrificing TKA, it has some disadvantages such as increased polyethylene wear, additional bone resection, breakage of the post and patellar clunk syndrome. The ultra-congruent (UC) insert was designed to prevent bone loss in particular, and the other mentioned disadvantages of the conventional PS insert. However, patients with postoperative hyperextension have been seen to be associated with inferior clinical outcomes and knees become gradually more extended until two years after TKA using the UC insert.
Measure Participants 32 33
Mean (Standard Deviation) [score on a scale]
90.69
(4.41)
92.24
(4.44)
13. Secondary Outcome
Title Baseline Knee Society Functional Score Evaluation
Description The Knee Society Functional Score were applied preoperatively. The Knee Society Score (KSS) is comprised to two sections (each worth 100 points) for a maximum 200 points. The KSS Scale ranges from 0 to 200. One section is the Knee Society Clinical Score (KSCS) - points are given for pain, motion, and stability and points are deducted for flexion contracture, extension lag, and misalignment. The KSCS ranges from 0 to 100. The other section is the Knee Society Functional Score (KSFS) - points are assigned for walking distances and climbing stairs and points are deducted for use of walking aids. The KSFS ranges from 0 to 100. For each section, a score of 80-100 was considered excellent; 70-79, good; 60-69, fair; < 60, poor.
Time Frame Preoperative

Outcome Measure Data

Analysis Population Description
[Not Specified]
Arm/Group Title Ultra-congruent Insert Group Posterior Cruciate Ligament-stabilized Insert
Arm/Group Description Ultra-congruent inserts were used during total knee arthroplasty in patients randomized to this group. Vanguard® Knee System: Posterior cruciate ligament retention versus sacrificing is one of the main debates in total knee arthroplasty (TKA) and retention or sacrificing depends on the individual preference of the surgeon during the surgery. Whenever the surgeon decides to sacrifice the posterior cruciate ligament (PCL), another controversial question arises regarding the tibial insert type. While the posterior cruciate ligament-stabilized (PS) insert is widely used as the tibial insert in PCL-sacrificing TKA, it has some disadvantages such as increased polyethylene wear, additional bone resection, breakage of the post and patellar clunk syndrome. The ultra-congruent (UC) insert was designed to prevent bone loss in particular, and the other mentioned disadvantages of the conventional PS insert. However, patients with postoperative hyperextension have been seen to be associated with inferior clinical outcomes and knees become gradually more extended until two years after TKA using the UC insert. Posterior cruciate ligament-stabilized inserts were used during total knee arthroplasty in patients randomized to this group. Vanguard® Knee System: Posterior cruciate ligament retention versus sacrificing is one of the main debates in total knee arthroplasty (TKA) and retention or sacrificing depends on the individual preference of the surgeon during the surgery. Whenever the surgeon decides to sacrifice the posterior cruciate ligament (PCL), another controversial question arises regarding the tibial insert type. While the posterior cruciate ligament-stabilized (PS) insert is widely used as the tibial insert in PCL-sacrificing TKA, it has some disadvantages such as increased polyethylene wear, additional bone resection, breakage of the post and patellar clunk syndrome. The ultra-congruent (UC) insert was designed to prevent bone loss in particular, and the other mentioned disadvantages of the conventional PS insert. However, patients with postoperative hyperextension have been seen to be associated with inferior clinical outcomes and knees become gradually more extended until two years after TKA using the UC insert.
Measure Participants 32 33
Mean (Standard Deviation) [units on a scale]
49.53
(14.11)
51.27
(11.60)
14. Secondary Outcome
Title Knee Society Functional Score Evaluation at 3 Months Postoperatively
Description The Knee Society Functional Score were applied at postoperative 3 months. The Knee Society Score (KSS) is comprised to two sections (each worth 100 points) for a maximum 200 points. The KSS Scale ranges from 0 to 200. One section is the Knee Society Clinical Score (KSCS) - points are given for pain, motion, and stability and points are deducted for flexion contracture, extension lag, and misalignment. The KSCS ranges from 0 to 100. The other section is the Knee Society Functional Score (KSFS) - points are assigned for walking distances and climbing stairs and points are deducted for use of walking aids. The KSFS ranges from 0 to 100. For each section, a score of 80-100 was considered excellent; 70-79, good; 60-69, fair; < 60, poor.
Time Frame 3 months postoperative

Outcome Measure Data

Analysis Population Description
[Not Specified]
Arm/Group Title Ultra-congruent Insert Group Posterior Cruciate Ligament-stabilized Insert
Arm/Group Description Ultra-congruent inserts were used during total knee arthroplasty in patients randomized to this group. Vanguard® Knee System: Posterior cruciate ligament retention versus sacrificing is one of the main debates in total knee arthroplasty (TKA) and retention or sacrificing depends on the individual preference of the surgeon during the surgery. Whenever the surgeon decides to sacrifice the posterior cruciate ligament (PCL), another controversial question arises regarding the tibial insert type. While the posterior cruciate ligament-stabilized (PS) insert is widely used as the tibial insert in PCL-sacrificing TKA, it has some disadvantages such as increased polyethylene wear, additional bone resection, breakage of the post and patellar clunk syndrome. The ultra-congruent (UC) insert was designed to prevent bone loss in particular, and the other mentioned disadvantages of the conventional PS insert. However, patients with postoperative hyperextension have been seen to be associated with inferior clinical outcomes and knees become gradually more extended until two years after TKA using the UC insert. Posterior cruciate ligament-stabilized inserts were used during total knee arthroplasty in patients randomized to this group. Vanguard® Knee System: Posterior cruciate ligament retention versus sacrificing is one of the main debates in total knee arthroplasty (TKA) and retention or sacrificing depends on the individual preference of the surgeon during the surgery. Whenever the surgeon decides to sacrifice the posterior cruciate ligament (PCL), another controversial question arises regarding the tibial insert type. While the posterior cruciate ligament-stabilized (PS) insert is widely used as the tibial insert in PCL-sacrificing TKA, it has some disadvantages such as increased polyethylene wear, additional bone resection, breakage of the post and patellar clunk syndrome. The ultra-congruent (UC) insert was designed to prevent bone loss in particular, and the other mentioned disadvantages of the conventional PS insert. However, patients with postoperative hyperextension have been seen to be associated with inferior clinical outcomes and knees become gradually more extended until two years after TKA using the UC insert.
Measure Participants 32 33
Mean (Standard Deviation) [units on a scale]
64.47
(14.19)
65.73
(11.90)
15. Secondary Outcome
Title Knee Society Functional Score Evaluation at 6 Months Postoperatively
Description The Knee Society Functional Score were applied at postoperative 6 months. The Knee Society Score (KSS) is comprised to two sections (each worth 100 points) for a maximum 200 points. The KSS Scale ranges from 0 to 200. One section is the Knee Society Clinical Score (KSCS) - points are given for pain, motion, and stability and points are deducted for flexion contracture, extension lag, and misalignment. The KSCS ranges from 0 to 100. The other section is the Knee Society Functional Score (KSFS) - points are assigned for walking distances and climbing stairs and points are deducted for use of walking aids. The KSFS ranges from 0 to 100. For each section, a score of 80-100 was considered excellent; 70-79, good; 60-69, fair; < 60, poor.
Time Frame 6 months postoperative

Outcome Measure Data

Analysis Population Description
[Not Specified]
Arm/Group Title Ultra-congruent Insert Group Posterior Cruciate Ligament-stabilized Insert
Arm/Group Description Ultra-congruent inserts were used during total knee arthroplasty in patients randomized to this group. Vanguard® Knee System: Posterior cruciate ligament retention versus sacrificing is one of the main debates in total knee arthroplasty (TKA) and retention or sacrificing depends on the individual preference of the surgeon during the surgery. Whenever the surgeon decides to sacrifice the posterior cruciate ligament (PCL), another controversial question arises regarding the tibial insert type. While the posterior cruciate ligament-stabilized (PS) insert is widely used as the tibial insert in PCL-sacrificing TKA, it has some disadvantages such as increased polyethylene wear, additional bone resection, breakage of the post and patellar clunk syndrome. The ultra-congruent (UC) insert was designed to prevent bone loss in particular, and the other mentioned disadvantages of the conventional PS insert. However, patients with postoperative hyperextension have been seen to be associated with inferior clinical outcomes and knees become gradually more extended until two years after TKA using the UC insert. Posterior cruciate ligament-stabilized inserts were used during total knee arthroplasty in patients randomized to this group. Vanguard® Knee System: Posterior cruciate ligament retention versus sacrificing is one of the main debates in total knee arthroplasty (TKA) and retention or sacrificing depends on the individual preference of the surgeon during the surgery. Whenever the surgeon decides to sacrifice the posterior cruciate ligament (PCL), another controversial question arises regarding the tibial insert type. While the posterior cruciate ligament-stabilized (PS) insert is widely used as the tibial insert in PCL-sacrificing TKA, it has some disadvantages such as increased polyethylene wear, additional bone resection, breakage of the post and patellar clunk syndrome. The ultra-congruent (UC) insert was designed to prevent bone loss in particular, and the other mentioned disadvantages of the conventional PS insert. However, patients with postoperative hyperextension have been seen to be associated with inferior clinical outcomes and knees become gradually more extended until two years after TKA using the UC insert.
Measure Participants 32 33
Mean (Standard Deviation) [units on a scale]
70.22
(12.98)
72.21
(11.73)
16. Secondary Outcome
Title Knee Society Functional Score Evaluation at 12 Months Postoperatively
Description The Knee Society Functional Score were applied at postoperative 12 months. The Knee Society Score (KSS) is comprised to two sections (each worth 100 points) for a maximum 200 points. The KSS Scale ranges from 0 to 200. One section is the Knee Society Clinical Score (KSCS) - points are given for pain, motion, and stability and points are deducted for flexion contracture, extension lag, and misalignment. The KSCS ranges from 0 to 100. The other section is the Knee Society Functional Score (KSFS) - points are assigned for walking distances and climbing stairs and points are deducted for use of walking aids. The KSFS ranges from 0 to 100. For each section, a score of 80-100 was considered excellent; 70-79, good; 60-69, fair; < 60, poor.
Time Frame 12 months postoperative

Outcome Measure Data

Analysis Population Description
[Not Specified]
Arm/Group Title Ultra-congruent Insert Group Posterior Cruciate Ligament-stabilized Insert
Arm/Group Description Ultra-congruent inserts were used during total knee arthroplasty in patients randomized to this group. Vanguard® Knee System: Posterior cruciate ligament retention versus sacrificing is one of the main debates in total knee arthroplasty (TKA) and retention or sacrificing depends on the individual preference of the surgeon during the surgery. Whenever the surgeon decides to sacrifice the posterior cruciate ligament (PCL), another controversial question arises regarding the tibial insert type. While the posterior cruciate ligament-stabilized (PS) insert is widely used as the tibial insert in PCL-sacrificing TKA, it has some disadvantages such as increased polyethylene wear, additional bone resection, breakage of the post and patellar clunk syndrome. The ultra-congruent (UC) insert was designed to prevent bone loss in particular, and the other mentioned disadvantages of the conventional PS insert. However, patients with postoperative hyperextension have been seen to be associated with inferior clinical outcomes and knees become gradually more extended until two years after TKA using the UC insert. Posterior cruciate ligament-stabilized inserts were used during total knee arthroplasty in patients randomized to this group. Vanguard® Knee System: Posterior cruciate ligament retention versus sacrificing is one of the main debates in total knee arthroplasty (TKA) and retention or sacrificing depends on the individual preference of the surgeon during the surgery. Whenever the surgeon decides to sacrifice the posterior cruciate ligament (PCL), another controversial question arises regarding the tibial insert type. While the posterior cruciate ligament-stabilized (PS) insert is widely used as the tibial insert in PCL-sacrificing TKA, it has some disadvantages such as increased polyethylene wear, additional bone resection, breakage of the post and patellar clunk syndrome. The ultra-congruent (UC) insert was designed to prevent bone loss in particular, and the other mentioned disadvantages of the conventional PS insert. However, patients with postoperative hyperextension have been seen to be associated with inferior clinical outcomes and knees become gradually more extended until two years after TKA using the UC insert.
Measure Participants 32 33
Mean (Standard Deviation) [units on a scale]
81.47
(10.45)
82.79
(10.06)

Adverse Events

Time Frame 1 year
Adverse Event Reporting Description Definition of adverse event does not differ from the clinicaltrials.gov definitions.
Arm/Group Title Ultra-congruent Insert Group Posterior Cruciate Ligament-stabilized Insert
Arm/Group Description Ultra-congruent inserts were used during total knee arthroplasty in patients randomized to this group. Posterior cruciate ligament-stabilized inserts were used during total knee arthroplasty in patients randomized to this group.
All Cause Mortality
Ultra-congruent Insert Group Posterior Cruciate Ligament-stabilized Insert
Affected / at Risk (%) # Events Affected / at Risk (%) # Events
Total 0/32 (0%) 0/33 (0%)
Serious Adverse Events
Ultra-congruent Insert Group Posterior Cruciate Ligament-stabilized Insert
Affected / at Risk (%) # Events Affected / at Risk (%) # Events
Total 1/32 (3.1%) 2/33 (6.1%)
Infections and infestations
Superficial infection 0/32 (0%) 0 1/33 (3%) 1
Vascular disorders
Deep vein thrombosis 1/32 (3.1%) 1 1/33 (3%) 1
Other (Not Including Serious) Adverse Events
Ultra-congruent Insert Group Posterior Cruciate Ligament-stabilized Insert
Affected / at Risk (%) # Events Affected / at Risk (%) # Events
Total 0/32 (0%) 0/33 (0%)

Limitations/Caveats

[Not Specified]

More Information

Certain Agreements

Principal Investigators are NOT employed by the organization sponsoring the study.

There is NOT an agreement between Principal Investigators and the Sponsor (or its agents) that restricts the PI's rights to discuss or publish trial results after the trial is completed.

Results Point of Contact

Name/Title Dr. Erdem Aras Sezgin
Organization Aksaray University Training and Research Hospital
Phone +905057777697
Email sezginmd@hotmail.com
Responsible Party:
Erdem A Sezgin, Co-Principal Investigator, Aksaray University Training and Research Hospital
ClinicalTrials.gov Identifier:
NCT04419311
Other Study ID Numbers:
  • 1585/2017
First Posted:
Jun 5, 2020
Last Update Posted:
Mar 16, 2021
Last Verified:
Feb 1, 2021