N_ORS: Feasibility of Nurse-led Retinal Imaging for Retinopathy of Prematurity Screening Employing the Optos California

Sponsor
University of Nottingham (Other)
Overall Status
Completed
CT.gov ID
NCT06063863
Collaborator
University Hospitals of Derby and Burton NHS Foundation Trust (Other)
52
1
1
26.7
1.9

Study Details

Study Description

Brief Summary

Retinopathy of prematurity (ROP) is a preventable cause of blindness in babies who are born early i.e. premature. Internationally, there is a shortage of skilled ophthalmologists willing and able to screen for ROP. Even in the UK, not all hospitals have skilled ophthalmologists and premature babies have to travel to other hospitals, often long distances, to have their eyes examined. As a missed examination can lead to sight loss, this is a burden for families and carers of premature babies. To fill this gap, previous studies have explored the use of non-ophthalmologists healthcare workers to increase the workforce screening for ROP.

Recently, the Optos ultra-widefield retinal-imaging device (Optos PLC, Dunfermline, Scotland, UK) has been used to help document different stages of ROP in infants. This specialised retinal imaging system uses an internal ellipsoid mirror to capture fundal imaging angles of up to 200 degrees, or more than 80% of the entire retina, in a single image. A single retinal image can be acquired in a quarter of a second and is automatically captured when the infant's pupils are aligned with the Optos imaging device. No contact with the eye is necessary to capture an image of the retina. To date, there are no studies that have validated the Optos as a nurse-led screening tool for ROP.

This is a prospective study to determine and validate the feasibility of neonatal nurse-led retinal imagers for ROP screening employing the Optos imaging device. The main purpose of this study will be to test if it is possible for trained nurses to take good images of the back of babies eyes (retina) and if these images can be used by remotely placed ophthalmologists to diagnose and grade ROP. The investigators will compare how good the diagnosis and grading done using Optos images are compared to the current gold standard method (BIO). The investigators will also test how much agreement there is between ophthalmologists in interpreting Optos images by asking two ophthalmologists to grade the images.

Condition or Disease Intervention/Treatment Phase
  • Diagnostic Test: Nurse led screening with Optos ultra-widefield retinal-imaging device
N/A

Detailed Description

Retinopathy of Prematurity (ROP) is the major cause of blindness in babies who are born early i.e. premature (defined as those born at less than 37 weeks' gestation). All premature infants are at risk of developing ROP and are therefore routinely screened for the condition. With improvements, it has become one of the few preventable causes of childhood blindness, and recent reports suggest that most babies are saved from sight loss with proper screening and treatment. As neonatal care gets better, more premature babies are surviving and the need for ROP screening is increasing all over the world: about 8200 babies need screening in the UK each year. Very few babies need treatment - less than 2% in the UK. At the Royal Derby Hospital (RDH), about 4 to 5 infants are screened each week, with only 1 to 2 needing treatment per year.

The international classification of ROP describes ROP by its severity or stage (Stages 1-5), location by zone (I-III), extent by clock hours, and by the presence of pre-plus and plus disease. The aim of screening is to find out if the baby has any of these so that treatment can be done if needed. All babies born before 32 weeks or at less than 1500 g birth weight must have regular screening. For those born at <27 weeks gestation, the first ROP screening examination should be undertaken when at 30 to 31 weeks corrected gestational age while for babies who are more mature at birth, it is done between 4 to 5 weeks of actual age. Follow-up examinations should be performed every 1 to 2 weeks thereafter until the ophthalmologists (eye doctors) can see that the retina is fully matured. If there are signs of ROP, examinations should be performed more frequently, either weekly or twice a week.

All ROP screening examinations are currently done by a highly trained ophthalmologist using binocular indirect ophthalmoscopy (BIO), which is considered the gold standard method. The retinal findings are seen and drawn by hand in the patient's notes. These screening examinations are time-consuming for the ophthalmologist, and costly for the medical system as they need the ophthalmologist to be physically present and to do the screening. This makes it difficult for babies in remote hospitals or settings without trained ophthalmologists to have ROP screening. To reduce this difficulty several groups have developed cameras that can take pictures of the retina. The pictures can then be transferred to ophthalmologists who can diagnose and grade the ROP remotely.

The most common imaging system used for ROP screening is the RetCam (Clarity Medical Systems, Pleasanton, California, USA). The RetCam imaging system, a hand-held camera, is connected by a fiber optic cable to an electro-optical box housed in a movable cart. To take images with RetCam, the babies pupils have to be dilated (opened up) by using medicines and local painkillers. The eyelids are held open with a lid speculum and the camera probe is applied with a contact lens gel such that it is touching the front of the eye, over the open pupil. RetCam images are not able to include the whole of the back of the eye in one picture and therefore multiple images have to be taken. This takes time and effort from highly-skilled imagers. Also, the picture quality is not good if the pupil is not opened properly or if there is some obstruction in the way such as blood in the eye. As the camera has to touch the eye, it can hurt the eye and make the baby uncomfortable such as by stimulating eye nerves that cause the heart to slow down and lower the baby's blood oxygen levels. Too much pressure on the eye can also change the picture and make the diagnosis difficult.

More recently, a new camera, the Optos ultra-widefield retinal-imaging device (Optos PLC, Dunfermline, Scotland, UK) has been used to take and store pictures of different stages of ROP in babies. This camera has a specialised mirror to capture pictures that contain details of nearly all of the retina in a single photo. It is quick, the complete single photo can the taken in a quarter of a second. In addition, it can take pictures without touching the baby's eyes so there is no risk of damage to the eye. Thus, the Optos camera has many advantages over the RetCam.

If this camera could be used by nurses trained to take pictures and ophthalmologists could diagnose and grade ROP from these pictures, babies could be screened without every hospital needing to have a specially trained ophthalmologist or the baby travelling distances to go to a centre that had such a doctor. As most babies only need screening, ophthalmologists could ask for only those who need treatment (about 2 in 100 babies) to be transferred to specialist centres. This could reduce the cost of care to the NHS and problems for families of premature babies.

In the Derby neonatal unit, the study team have already tested the Optos camera and found that it is safe to use in babies and did not cause them any undue discomfort. In this study, the study team want to see if it is possible to train a nurse to take pictures of the retina using the Optos camera and if ophthalmologists can diagnose and grade ROP accurately using these images. No studies have been done to test this.

Study Design

Study Type:
Interventional
Actual Enrollment :
52 participants
Allocation:
N/A
Intervention Model:
Single Group Assignment
Masking:
None (Open Label)
Primary Purpose:
Diagnostic
Official Title:
Feasibility of Nurse-led Retinal Imaging for Retinopathy of Prematurity Screening Employing the Optos California
Actual Study Start Date :
May 10, 2021
Actual Primary Completion Date :
Aug 1, 2023
Actual Study Completion Date :
Aug 1, 2023

Arms and Interventions

Arm Intervention/Treatment
Experimental: Nurse led screening for ROP with OPTOS camera

Screening completed by nurse trained in the use of an OPTOS camera to obtain images, at the same time as screening via standard procedure by a trained ophthalmologist using binocular indirect ophthalmoscopy (BIO).

Diagnostic Test: Nurse led screening with Optos ultra-widefield retinal-imaging device
The screening will be completed by nurse trained in using an OPTOS camera, at the same time as screening for ROP by a trained ophthalmologist using standard procedure for screening for ROP using binocular indirect ophthalmoscopy (BIO). Images taken with OPTOS will then be remotely assessed by a qualified independent opthalmologist, to see if they can diagnose and grade ROP accurately using these images

Outcome Measures

Primary Outcome Measures

  1. Retinopathy of prematurity diagnosis [18 months]

    To evaluate the agreement by measuring the Kappa Co-efficient for agreement, for diagnosis and grading of ROP and posterior pole (pre-plus or plus disease) and peripheral retinal vascular changes, between nurse-led OPTOS imaging and indirect Ophthalmoscopy performed by an ophthalmologist.

  2. Retinopathy of prematurity grading [18 months]

    To evaluate the agreement by measuring the Kappa Co-efficient for agreement, for diagnosis and grading of ROP and posterior pole (pre-plus or plus disease) and peripheral retinal vascular changes, between nurse-led OPTOS imaging and indirect Ophthalmoscopy performed

Eligibility Criteria

Criteria

Ages Eligible for Study:
32 Weeks to 1 Year
Sexes Eligible for Study:
All
Accepts Healthy Volunteers:
No
Inclusion Criteria:
  • Infants (upper age limit 1 years old) eligible for routine ROP screening as part UK ROP screening guidelines (any infant born at or before 32 weeks' gestation and / or weigh 1500 grams or less).

  • Infants with parents/legal guardians who can give written informed consent.

Exclusion Criteria:
  • Infants not eligible for routine ROP screening under UK ROP screening guidelines.

  • Infants who are deemed not well enough by the attending consultant neonatologist for retinal examination or retinal imaging.

  • Any premature infant with media opacities that prevents adequate visualisation of the retina.

  • Inability of the parents to give written, informed consent.

Contacts and Locations

Locations

Site City State Country Postal Code
1 University Hospitals of Derby and Burton Derby Derbyshire United Kingdom DE22 3DT

Sponsors and Collaborators

  • University of Nottingham
  • University Hospitals of Derby and Burton NHS Foundation Trust

Investigators

  • Principal Investigator: Shalini Ojha, Professor, University of Nottingham

Study Documents (Full-Text)

None provided.

More Information

Publications

Responsible Party:
University of Nottingham
ClinicalTrials.gov Identifier:
NCT06063863
Other Study ID Numbers:
  • 20046
First Posted:
Oct 3, 2023
Last Update Posted:
Oct 3, 2023
Last Verified:
Sep 1, 2023
Individual Participant Data (IPD) Sharing Statement:
No
Plan to Share IPD:
No
Studies a U.S. FDA-regulated Drug Product:
No
Studies a U.S. FDA-regulated Device Product:
No
Additional relevant MeSH terms:

Study Results

No Results Posted as of Oct 3, 2023