Testing Automated Echocardiography

Sponsor
University of Pennsylvania (Other)
Overall Status
Enrolling by invitation
CT.gov ID
NCT04943965
Collaborator
Philips Medical Systems (Industry)
120
1
24.6
4.9

Study Details

Study Description

Brief Summary

The purpose of this research is to determine if software can accurately predict certain heart parameters from chest wall ultrasound (non-radiation, non-invasive, imaging modality). An approved ultrasound device (Lumify) will be used to obtain images prior to cardiac surgery.

Condition or Disease Intervention/Treatment Phase
  • Diagnostic Test: Ultrasound imaging

Detailed Description

This is a prospective, non-randomized, single visit clinical study. No additional interventions or follow-ups to the subject will be conducted as part of the study. All investigational scanning will be conducted using the as low as reasonably achievable (ALARA) principle to ensure the safety of the subjects. As ultrasound imaging is based on non-ionizing radiation, it is generally considered safe when used by appropriately trained health care providers. The study population will be adults. There are no specific enrollment targets, however it is expected that 120 subjects or fewer will be scanned to obtain the required 40 echocardiographic exams to complete this study. Exceeding the enrollment target of 120 subjects will require an amendment to the protocol. It is estimated this study will last for 12 months, at that time it will be evaluated and may potentially be amended.

This study is considered non-significant risk. There is no specific benefit to individuals participating in the study, however the feedback received is used to improve Philips systems and that can lead to increased diagnostic ability in future products.

Study Design

Study Type:
Observational
Anticipated Enrollment :
120 participants
Observational Model:
Cohort
Time Perspective:
Prospective
Official Title:
Repeatability, Reproducibility, and Efficacy Testing of Automated Hemodynamic Output by Transthoracic Echocardiography
Actual Study Start Date :
Aug 11, 2021
Anticipated Primary Completion Date :
Aug 31, 2022
Anticipated Study Completion Date :
Aug 31, 2023

Arms and Interventions

Arm Intervention/Treatment
Patient undergoing a cardiac surgery

Patients undergoing cardiac surgeries listed in inclusion criteria will be in this group.

Diagnostic Test: Ultrasound imaging
The intervention is non-invasive, chest wall, echocardiographic image acquisition in the cardiac operating room prior to a scheduled cardiac surgery. Echocardiographic images will be taken at three, key, timepoints in the operating room prior to cardiac surgery: Pre-induction. Post-induction / Pre-PAC Post-induction / Post-PAC
Other Names:
  • Philips Ultrasound
  • Lumify Ultrasound
  • Hemodynamic Automation device
  • Clinicians

    Clinicians defined as physicians (e.g. attendings, fellows, residents) or advanced practice providers (e.g. nurse practitioners and physician assistants) will be in this group.

    Outcome Measures

    Primary Outcome Measures

    1. Correlation Testing Lumify Ultrasound with Hemodynamic Automation [08/01/2021 - 08/31/2022]

      Correlation between the hemodynamic parameters (e.g. stroke volume, cardiac output, cardiac index) obtained from the Lumify Ultrasound with Hemodynamic Automation with same hemodynamic parameters obtained from the reference standard of transesophageal echocardiography and pulmonary arterial catheterization by a single echocardiographer.

    2. Repeatability (intraobserver variability) of hemodynamic parameters obtained from the Lumify Ultrasound with Hemodynamic Automation [08/01/2021 - 08/31/2022]

      Repeatability of the hemodynamic parameters (e.g. stroke volume, cardiac output, cardiac index) obtained from the Lumify Ultrasound with Hemodynamic Automation at set time intervals of one minute by a single echocardiographer.

    3. Reproducibility (interobserver variability) of hemodynamic parameters obtained from the Lumify Ultrasound with Hemodynamic Automation [08/01/2021 - 08/31/2022]

      Reproducibility of the hemodynamic parameters (e.g. stroke volume, cardiac output, cardiac index) obtained from the Lumify Ultrasound with Hemodynamic Automation at set time intervals of one minute by a multiple echocardiographers blinded to each other's data.

    Secondary Outcome Measures

    1. Clinical Efficacy of hemodynamic parameters obtained from the Lumify Ultrasound with Hemodynamic Automation [08/01/2022 - 08/31/2023]

      Clinician providers will evaluate the clinical efficacy of hemodynamic parameters obtained from the Lumify Ultrasound with Hemodynamic Automation by survey questions directly comparing the Lumify echocardiographic imaging to the reference standard

    Eligibility Criteria

    Criteria

    Ages Eligible for Study:
    18 Years and Older
    Sexes Eligible for Study:
    All
    Accepts Healthy Volunteers:
    No
    Inclusion Criteria:
    • Patient willing and capable of providing informed consent and physically capable of performing study related activities

    • Patient undergoing any of the following types of cardiac surgeries:

    Aortic valve repair Aortic valve replacement Mitral valve repair Mitral valve replacement Tricuspid valve repair Tricuspid valve replacement Pulmonic valve repair Pulmonic valve replacement Ascending aorta replacement Coronary Artery Bypass Graft (CABG) surgery

    Exclusion Criteria:
    • Pregnant women

    • Patients known to be under the jurisdiction of the Department of Corrections

    • Patients with persistent arrhythmias as confirmed by the principal investigator (MacKay) on a caseby-case basis (e.g. atrial fibrillation, atrial flutter, frequent premature ventricular contractions, frequent premature atrial contractions, etc.)

    • Patients with a contraindication to Pulmonary Arterial Catheter (PAC) monitoring as confirmed by the principal investigator (MacKay) on a case-by-case basis (e.g. tricuspid or pulmonic valve endocarditis, venous thrombosis present in the superior vena cava, right atrium, or main pulmonary artery)

    • Patients with a contraindication to TEE monitoring as confirmed by the principal investigator (MacKay) on a case-by-case basis (e.g. previous esophageal surgery, prior gastric bypass surgery, esophageal varices).

    Contacts and Locations

    Locations

    Site City State Country Postal Code
    1 Hospital of the University of Pennsylvania Philadelphia Pennsylvania United States 19104

    Sponsors and Collaborators

    • University of Pennsylvania
    • Philips Medical Systems

    Investigators

    • Study Chair: Roderic G Eckenhoff, MD, University of Pennsylvania, Department of Anesthesiology and Critical Care

    Study Documents (Full-Text)

    None provided.

    More Information

    Publications

    None provided.
    Responsible Party:
    Emily MacKay, DO, MSHP, Assistant Professor, University of Pennsylvania
    ClinicalTrials.gov Identifier:
    NCT04943965
    Other Study ID Numbers:
    • 848896
    First Posted:
    Jun 29, 2021
    Last Update Posted:
    Aug 30, 2021
    Last Verified:
    Aug 1, 2021
    Individual Participant Data (IPD) Sharing Statement:
    Undecided
    Plan to Share IPD:
    Undecided
    Studies a U.S. FDA-regulated Drug Product:
    No
    Studies a U.S. FDA-regulated Device Product:
    Yes
    Keywords provided by Emily MacKay, DO, MSHP, Assistant Professor, University of Pennsylvania

    Study Results

    No Results Posted as of Aug 30, 2021