Minimally Invasive Nasal Trabeculostomy - Performance and Safety Evaluation Study
Study Details
Study Description
Brief Summary
Minimally Invasive Nasal Trabeculostomy System is intended for the reduction of elevated Intraocular Pressure (IOP).
The system is a powered surgical device, designed to create trabeculostomy openings of ~100 microns diameter each, via an ab-interno approach, directly from the anterior chamber (AC) into Schlemm's canal (SC) by removing segments of the trabecular meshwork (TM) (i.e., an ab-interno nasal trabeculostomy). The ab-interno approach aims to decrease IOP by increasing aqueous outflow through a direct communication from the AC to the outer wall of SC and the collector channels.
The opening in the nasal TM to the SC is created by a sterile surgical instrument (the Surgical Device) made of stainless steel and plastic, which is coupled to a rotating system (the Activation Device). The Surgical Device of the Minimally Invasive Nasal Trabeculostomy System consists of a stainless steel micro trephine, covered by a small needle-shaped sleeve with a rotating mechanism behind it. While being rotated, the Surgical Device enters and moves along the TM, removing TM tissue with minimal collateral damage.
Mechanical rotation of the Surgical Device is achieved by the rotating system (i.e., Activation Device), which is comprised of: (1) Controller which dictates the activation pulse duration, the rotations per minute (RPM) and the forwards and backwards movement of the micro trephine; (2) Motor; and (3) Footswitch.
Condition or Disease | Intervention/Treatment | Phase |
---|---|---|
|
N/A |
Study Design
Arms and Interventions
Arm | Intervention/Treatment |
---|---|
Experimental: single-arm study with intra-subject comparisons Study eyes of subjects with uncontrolled (IOP > 20 mmHg) open-angle glaucoma (OAG), where IOP is not controlled when using maximum tolerated glaucoma medications. |
Device: MINT
Mechanical rotation of the Surgical Device is achieved by the rotating system (i.e., Activation Device), which is comprised of: (1) Controller which dictates the activation pulse duration, the rotations per minute (RPM) and the forwards and backwards movement of the micro trephine; (2) Motor; and (3) Footswitch.
|
Outcome Measures
Primary Outcome Measures
- Proportion of subjects achieved a 20% or greater reduction in IOP from Baseline on the same or fewer number of medications [time frame: 24 weeks [24 months]
• Proportion of subjects achieving a 20% or greater reduction in IOP from baseline on the same or fewer number of medications. Subjects who will undergo a glaucoma-related secondary surgical intervention prior to the 24 weeks visit will be considered failures in the analysis.
Eligibility Criteria
Criteria
Inclusion Criteria:
-
- Male or female ≥ 18 years to ≤ 85 years old 2. Unsatisfactory IOP level, i.e., IOP ≥21 mmHg, at the screening visit in the study eye 3. Manifest open angle glaucoma (OAG) with typical disc or visual field changes and open iridocorneal angle in the study eye (in accordance with European Glaucoma Society (EGS) criteria 1) 4. Optic nerve appearance characteristic of glaucoma in the study eye 5. Shaffer grade ≥ III in all four angle quadrants in the study eye 6. Subject is treated with 0 to 5 hypotensive medications in the study eye 7. Subject is able and willing to attend all scheduled follow-up exams 8. Subject understands and signs the informed consent
Exclusion Criteria:
-
- Ocular conditions with a poorer prognosis in the fellow eye than in the study eye
-
Closed angle forms of glaucoma in either eye unless scheduled for cataract surgery immediately prior to the Minimally Invasive Nasal Trabeculostomy procedure or the study eye is pseudophakic with posterior chamber intraocular lens (PCIOL) 3. Congenital or developmental glaucoma in either eye 4. Fixation threatening visual-field defects or IOP ≥40 mmHg in the study eye
-
Other secondary glaucoma (such as neovascular, uveitic, lens-induced, trauma-induced, or glaucoma associated with increased episcleral venous pressure) in the study eye
-
Peripheral anterior synechiae (PAS), rubeosis or other angle abnormalities in the study eye
-
Subject has history of penetrating keratoplasty (PKP)
-
Any previous surgery in the study eye (at the access site of the Minimally Invasive Nasal Trabeculostomy System application), where the conjunctiva is not intact and elastic (except for clear corneal cataract surgery)
-
Any ocular disease or history in study eye such as severe dry eye, active proliferative retinopathy, ICE syndrome, epithelial or fibrous down growth, and ocular pathology that may interfere with accurate IOP measurements
-
Prior surgery for an ab-interno or ab-externo device implanted in or through the Schlemm's canal in the study eye.
-
Best-corrected visual acuity worse than 20/40 (Snellen equivalent) in the fellow eye
-
History of idiopathic or autoimmune uveitis in either eye
-
Severe trauma in study eye
-
Active iris neovascularization, previous cyclodestructive procedure, prior scleral buckling procedure, presence of silicone oil, need for glaucoma surgery combined with other ocular procedures or anticipated need for additional ocular surgery in study eye within 12 month period
-
Vitreous present in anterior chamber, prior vitrectomy or vitreous hemorrhage in study eye
-
Aphakia
-
Prior vitreoretinal surgery in study eye
-
Clinically significant ocular inflammation or infection within 90 days prior to screening
-
Unable to discontinue use of blood thinners in accordance with surgeon's standard preoperative instructions
-
Uncontrolled systemic disease that in the opinion of the investigator would put the subject's health at risk and/or prevent the subject from completing all study visits
-
Current participation or participation in another investigational drug or device clinical trial within the last 30 days before screening visit
-
Pregnant or lactating women
Contacts and Locations
Locations
Site | City | State | Country | Postal Code | |
---|---|---|---|---|---|
1 | S.V.Malayan'S Eye Center | Yerevan | Armenia | 0048 |
Sponsors and Collaborators
- Sanoculis Ltd
Investigators
None specified.Study Documents (Full-Text)
None provided.More Information
Publications